CRITICAL ANATOMIC DIMENSIONS FOR INTRATHORACIC CIRCULATORY ASSIST DEVICES

被引:16
|
作者
MUSSIVAND, T [1 ]
MASTERS, RG [1 ]
HENDRY, PJ [1 ]
RAJAGOPALAN, K [1 ]
WALLEY, VM [1 ]
NAHON, D [1 ]
HICKS, A [1 ]
KEON, WJ [1 ]
机构
[1] UNIV OTTAWA,OTTAWA CIVIL HOSP,INST HEART,OTTAWA K1N 6N5,ONTARIO,CANADA
关键词
MECHANICAL CIRCULATORY SUPPORT DEVICES; CHEST CAVITY DIMENSIONS; THORAX DIMENSIONS;
D O I
10.1111/j.1525-1594.1992.tb00310.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The design of totally implantable circulatory support devices must account for the anatomical constraints of the body cavity. To aid development of an implantable electrohydraulic ventricular assist device (EVAD), thoracic anatomical measurements and fit trials were performed. Various dimensions of the thoracic cavity were measured in 19 preserved and fresh cadavers. The mean left ventricle apex to chest wall distance was 2.7 +/- 0.2 cm. The mean sagittal radius of curvature at the fifth rib was 11.1 +/- 0.4 cm, and the transverse radius at the fifth rib was 9.4 +/- 0.3 cm. The mean sternal length was 18.9 +/- 1.2 cm. Using these measurements a model measuring 17.5 x 10.5 x 3.8 cm was designed. This model was placed in the chest of 11 patients undergoing open heart surgery. In one patient there was significant lung compression and in another hypotension suggesting great vessel compression. Finally, 1 of 11 patients had a short sternum and a small thorax, and the device did not fit. In conclusion, thoracic measurements may be used in the design of circulatory support devices. The model fit adequately in 73% of patient trials.
引用
收藏
页码:281 / 285
页数:5
相关论文
共 50 条
  • [1] Mechanical circulatory assist devices in critical care management
    Morales, DLS
    Oz, MC
    NEW HORIZONS-THE SCIENCE AND PRACTICE OF ACUTE MEDICINE, 1999, 7 (04): : 489 - 503
  • [2] Mechanical circulatory assist devices: a primer for critical care and emergency physicians
    Sen, Ayan
    Larson, Joel S.
    Kashani, Kianoush B.
    Libricz, Stacy L.
    Patel, Bhavesh M.
    Guru, Pramod K.
    Alwardt, Cory M.
    Pajaro, Octavio
    Farmer, J. Christopher
    CRITICAL CARE, 2016, 20
  • [3] Circulatory assist devices in cardiology
    Ferrari, M
    Figulla, HR
    DEUTSCHE MEDIZINISCHE WOCHENSCHRIFT, 2005, 130 (12) : 652 - 656
  • [4] MECHANICAL CIRCULATORY ASSIST DEVICES
    MAGOVERN, GJ
    PARK, SB
    MAGOVERN, GJ
    CHRISTLIEB, IY
    KAO, RL
    TEXAS HEART INSTITUTE JOURNAL, 1987, 14 (03) : 276 - 283
  • [5] Mechanical circulatory assist devices: a primer for critical care and emergency physicians
    Ayan Sen
    Joel S. Larson
    Kianoush B. Kashani
    Stacy L. Libricz
    Bhavesh M. Patel
    Pramod K. Guru
    Cory M. Alwardt
    Octavio Pajaro
    J. Christopher Farmer
    Critical Care, 20
  • [6] CURRENT STATUS OF CIRCULATORY ASSIST DEVICES
    COHEN, LS
    AMERICAN JOURNAL OF CARDIOLOGY, 1974, 33 (02): : 316 - 318
  • [7] Mechanical control of circulatory assist devices
    Pinder, SD
    Russell, DL
    Mussivand, T
    CANADIAN MEDICAL AND BIOLOGICAL ENGINEERING SOCIETY, PROCEEDINGS 1998, 1997, : 46 - 47
  • [8] Intrathoracic and intraabdominal wall implantation of a centrifugal blood pump for circulatory assist
    Wakisaka, Y
    Taenaka, Y
    Chikanari, K
    Okuzono, Y
    Nishimura, T
    Endo, S
    Nakatani, T
    Takano, H
    ARTIFICIAL ORGANS, 1998, 22 (06) : 493 - 497
  • [9] EFFECT OF CIRCULATORY ASSIST DEVICES ON STUNNED MYOCARDIUM
    BAVARIA, JE
    FURUKAWA, S
    KREINER, G
    GUPTA, KB
    STREICHER, J
    EDMUNDS, LH
    ANNALS OF THORACIC SURGERY, 1990, 49 (01): : 123 - 128
  • [10] CHARACTERIZATION OF ELASTOMERS FOR USE IN CIRCULATORY ASSIST DEVICES
    MCMILLIN, CR
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 1981, 182 (AUG): : 168 - POLY