Size-dependent thresholding as an optimal method for tumor volume delineation on positron emission tomography computed tomography: A Phantom study

被引:7
|
作者
Gupta, Arun [1 ]
Sharma, Punit [1 ]
Patel, Chetan D. [1 ]
Maharjan, Sagar [1 ]
Pandey, Anil [2 ]
Kumar, Rakesh [1 ]
Malhotra, Arun [1 ]
机构
[1] All India Inst Med Sci, Inst Rotary Canc Hosp, Dept Nucl Med, New Delhi, India
[2] All India Inst Med Sci, Inst Rotary Canc Hosp, Dept Hlth Phys, New Delhi, India
来源
INDIAN JOURNAL OF NUCLEAR MEDICINE | 2011年 / 26卷 / 01期
关键词
Positron emission tomography-computed tomography; phantom thresholding; tumor volume;
D O I
10.4103/0972-3919.84598
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: Use of a fixed threshold value for tumor volume delineation in positron emission tomography (PET) images will ignore the effect of size of the lesion and source to background ratio (SBR). The purpose of this Phantom study was to evaluate the effect of the size of the lesion and SBR on the threshold to be used for PET tumor volume delineation. Materials and Methods: Phantom used in the study comprised a sphere-cylinder assembly containing six spheres of different inner diameters (1.10, 1.35, 1.44, 1.50, 1.83 and 1.93 cm) with inner volumes of 0.70, 1.30, 1.50, 1.77, 3.22 and 3.82 cm', respectively. The scans were acquired with SBR of 6:01, 7:01, 8:01 and 10:01. These SBRs were calculated from 42 patients with lymphoma to simulate clinical images. PET tumor volume was calculated using RT_Image software at different threshold values (40, 45, 50, 55, 60, 65, 70 and 75% of SUV..) for each sphere at different SBRs. The threshold intensity value at which the calculated volume was nearly equal to actual volume of spheres was considered as the standardized threshold intensity (STI) value. Results: STI values depended on the diameter of the sphere and not on the SBR. It is found that 40% threshold is suitable for calculating the volume of any lesion with diameter greater than 1.83 cm, 60% for diameter greater than 1.35 cm but less than 1.83 cm, and 75% for diameter less than 1.35 cm. Conclusion: Size-dependent thresholding is an accurate and reproducible method of tumor volume delineation on PET-computed tomography (CT).
引用
收藏
页码:22 / 26
页数:5
相关论文
共 50 条
  • [31] Tumor lesion detection - When is integrated positron emission tomography/computed tomography more accurate than side-by-side interpretation of positron emission tomography and computed tomography?
    Metser, U
    Golan, O
    Levine, CD
    Even-Sapir, E
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2005, 29 (04) : 554 - 559
  • [32] Optimal definition of biological tumor volume using positron emission tomography in an animal model
    Wu, Ingrid
    Wang, Hao
    Huso, David
    Wahl, Richard L.
    EJNMMI RESEARCH, 2015, 5
  • [33] Optimal definition of biological tumor volume using positron emission tomography in an animal model
    Ingrid Wu
    Hao Wang
    David Huso
    Richard L. Wahl
    EJNMMI Research, 5
  • [34] Positron emission tomography/computed tomography as a method for detection and monitoring of laryngeal chondritis
    Markus Gugatschka
    Thomas Schwarz
    Gerhard Friedrich
    European Archives of Oto-Rhino-Laryngology, 2008, 265 : 1373 - 1377
  • [35] Positron emission tomography/computed tomography as a method for detection and monitoring of laryngeal chondritis
    Gugatschka, Markus
    Schwarz, Thomas
    Friedrich, Gerhard
    EUROPEAN ARCHIVES OF OTO-RHINO-LARYNGOLOGY, 2008, 265 (11) : 1373 - 1377
  • [36] Segmentation of lung lesion volume by adaptive positron emission tomography image thresholding
    Erdi, YE
    Mawlawi, O
    Larson, SM
    Imbriaco, M
    Yeung, H
    Finn, R
    Humm, JL
    CANCER, 1997, 80 (12) : 2505 - 2509
  • [37] The potential of positron emission tomography for intratreatment dynamic lung tumor tracking: A phantom study
    Yang, Jaewon
    Yamamoto, Tokihiro
    Mazin, Samuel R.
    Graves, Edward E.
    Keall, Paul J.
    MEDICAL PHYSICS, 2014, 41 (02)
  • [38] Humanoid Training Phantom for Positron Emission Tomography-Computed Tomography-Guided Percutaneous Biopsy
    Singh, Harmandeep
    Kumar, Rajender
    Watts, Ankit
    Rana, Nivedita
    Vatsa, Rakhee
    Mittal, Bhagwant Rai
    INDIAN JOURNAL OF NUCLEAR MEDICINE, 2021, 36 (02): : 201 - 202
  • [39] Gastrointestinal stromal tumor (GIST) of the esophagus detected by positron emission tomography/computed tomography
    Chang, WC
    Tzao, C
    Shen, DHY
    Cheng, CY
    Yu, CP
    Hsu, HH
    DIGESTIVE DISEASES AND SCIENCES, 2005, 50 (07) : 1315 - 1318
  • [40] Gastrointestinal Stromal Tumor (GIST) of the Esophagus Detected by Positron Emission Tomography/Computed Tomography
    Wei-chou Chang
    Ching Tzao
    Daniel Hueng-yuan Shen
    Cheng-yi Cheng
    Cheng-ping Yu
    Hsian-he Hsu
    Digestive Diseases and Sciences, 2005, 50 : 1315 - 1318