CORRELATIONS BETWEEN ACOUSTIC AND TEXTURE PARAMETERS FROM RF AND B-MODE LIVER ECHOGRAMS

被引:43
|
作者
THIJSSEN, JM
OOSTERVELD, BJ
HARTMAN, PC
ROSENBUSCH, GJE
机构
[1] UNIV HOSP NIJMEGEN,INST RADIOTHERAPY,6500 HB NIJMEGEN,NETHERLANDS
[2] UNIV HOSP NIJMEGEN,DEPT DIAGNOST RADIOL,6500 HB NIJMEGEN,NETHERLANDS
来源
ULTRASOUND IN MEDICINE AND BIOLOGY | 1993年 / 19卷 / 01期
关键词
ACOUSTOSPECTROGRAPHY; ECHOGRAM; LIVER; TEXTURE ANALYSIS;
D O I
10.1016/0301-5629(93)90013-E
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Radio frequency (RF) echograms were acquired from human subjects without liver pathology (n = 126), who were in the range of 20 to 84 years of age. After appropriate correction for the equipment settings and performance characteristics, acoustospectrographic parameters were estimated. The data were corrected for the frequency-dependent attenuation and then software demodulated. Image texture parameters were calculated based on the assumption of a particular tissue model, and purely statistical parameters were also estimated. The data thus obtained from each subject were corrected for the age trend that was assessed in an earlier publication by the authors. A total of 18 parameters was considered and the correlations were analyzed. It is concluded that nine parameters could be identified which did not strongly correlate with each other. This conclusion is supported by analysis of correlation data from patients. In a companion paper, a discriminant analysis is reported, based on the selected parameters (nine) and applied to the differentiation between normals and various classes of diffuse liver disease.
引用
收藏
页码:13 / 20
页数:8
相关论文
共 50 条
  • [41] New Reconstruction Method for Needle Contrast Optimization in B-Mode Ultrasound Image by Extracting RF Signal Parameters in Frequency Domain
    Susanti, Hesty
    Suprijanto
    Kurniadi, Deddy
    JOURNAL OF ENGINEERING AND TECHNOLOGICAL SCIENCES, 2020, 52 (04): : 514 - 533
  • [42] Discordance diagnosis between B-mode ultrasonography and MRI proton density fat fraction for fatty liver
    Chul-min Lee
    Mimi Kim
    Bo-Kyeong Kang
    Dae Won Jun
    Eileen L. Yoon
    Scientific Reports, 13 (1)
  • [43] Discordance diagnosis between B-mode ultrasonography and MRI proton density fat fraction for fatty liver
    Lee, Chul-min
    Kim, Mimi
    Kang, Bo-Kyeong
    Jun, Dae Won
    Yoon, Eileen L.
    SCIENTIFIC REPORTS, 2023, 13 (01):
  • [44] A Reappraisal of the Diagnostic Performance of B-Mode Ultrasonography for Mild Liver Steatosis
    Lee, Chul-min
    Yoon, Eileen L.
    Nakajima, Atsushi
    Yoneda, Masato
    Toyoda, Hidenori
    Yasuda, Satoshi
    Lee, Jonghyun
    Kim, Mimi
    Kang, Bo-Kyeong
    Nguyen, Mindie H.
    Jun, Dae Won
    Sumida, Yoshio
    AMERICAN JOURNAL OF GASTROENTEROLOGY, 2023, 118 (05): : 840 - 847
  • [45] Agreement between A-mode and B-mode ultrasonography in the measurement of ocular distances
    Hamidzada, WA
    Osuobeni, EP
    VETERINARY RADIOLOGY & ULTRASOUND, 1999, 40 (05) : 502 - 507
  • [46] Myocardial motion analysis from B-mode echocardiograms
    Sühling, M
    Arigovindan, M
    Jansen, C
    Hunziker, P
    Unser, M
    IEEE TRANSACTIONS ON IMAGE PROCESSING, 2005, 14 (04) : 525 - 536
  • [47] ONLINE ULTRASONIC TISSUE CHARACTERIZATION USING B-MODE AND RF DATA-ANALYSIS
    SCHLAPS, D
    RATH, U
    ZUNA, I
    VOLK, JF
    LORENZ, A
    LORENZ, D
    LEHMANN, U
    VANKAICK, G
    LORENZ, WJ
    ULTRASONIC IMAGING, 1985, 7 (01) : 80 - 81
  • [48] Quantitative breast elastography from B-mode images
    Rabin, Carolina
    Benech, Nicolas
    MEDICAL PHYSICS, 2019, 46 (07) : 3001 - 3012
  • [49] A comparison of acute changes in muscle thickness between A-mode and B-mode ultrasound
    Kuehne, Tayla E.
    Yitzchaki, Noam
    Jessee, Matthew B.
    Graves, B. Sue
    Buckner, Samuel L.
    PHYSIOLOGICAL MEASUREMENT, 2019, 40 (11)
  • [50] Tendinopathy discrimination by use of spatial frequency parameters in ultrasound B-mode images
    Bashford, Gregory R.
    Tornsen, Nicholas
    Arya, Shruti
    Burnfield, Judith M.
    Kulig, Kornelia
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2008, 27 (05) : 608 - 615