DETERMINATION OF BENZIDINE IN URINE USING CAPILLARY GAS-CHROMATOGRAPHY AND NEGATIVE-ION CHEMICAL-IONIZATION MASS-SPECTROMETRY

被引:3
|
作者
JEDRZEJCZAK, K
GAIND, VS
机构
[1] Occupational Health Laboratory, Ontario Ministry of Labour, Weston, M9P 3T1, Ontario
来源
关键词
D O I
10.1007/BF00322845
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A highly sensitive and specific gas chromatographic-mass spectrometric (GC-MS) assay for the determination of benzidine in urine is reported. It is based on the solvent extraction of the hydrolysed benzidine conjugates, together with the deuterium-labelled benzidine-d8 added as an internal standard, and a two-phase derivatisation procedure using pentafluoropropionic anhydride (PFPA) in the presence of pyridine as the phase-transfer catalyst. The reaction is complete within 5 min at room temperature. The pentafluoropropyl derivatives are quantified through capillary column GC-MS using selected ion monitoring (SIM) in the negative-ion chemical ionisation mode (NICI). The lower limit of detection for benzidine was 0.5 mug/l and the calibration plot showed linearity between 2 mug/l and 200 mug/l. The recovery of the analyte added to pooled urine was above 82%. Analysis of 20 urine samples from un-exposed persons and 20 urine specimens of workers employed in a polyurethane-making plant using this procedure showed no substances likely to manifest false positive results in the range of interest.
引用
收藏
页码:136 / 140
页数:5
相关论文
共 50 条