SAMPLE TIP COUPLING EFFICIENCIES OF THE PHOTON-SCANNING TUNNELING MICROSCOPE

被引:45
|
作者
SALOMON, L
DEFORNEL, F
GOUDONNET, JP
机构
[1] Laboratoire Physique du Solide Section, Photoêlectricité, Université de Bourgogne, Dijon, 21004
关键词
D O I
10.1364/JOSAA.8.002009
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
The photon-scanning tunneling microscope is the photon analog to the electron-scanning tunneling microscope. It uses the evanescent field due to the total internal reflection of a light beam in a prism, modulated by a sample attached to the prism. The exponential decay of the evanescent field is characterized by the penetration depth d(p) and depends on the angle of incidence theta, the wavelength, and the polarization of the incident beam. The 1/e decay lengths range from 150 to 265 nm as deduced from the expression of the electric-field intensity in the rarer medium for theta = pi/2. If we place another optically transparent medium near the surface, frustrated total reflection occurs. It is shown theoretically and experimentally that, if we choose an appropriate angle of incidence theta(theta not-equal pi/2) and change the index of refraction of one of the media, the decay length of the electric field can be extremely small, so that images with an improved resolution can be produced.
引用
收藏
页码:2009 / 2015
页数:7
相关论文
共 50 条
  • [1] INSTRUMENTATION - PHOTON-SCANNING TUNNELING MICROSCOPE ACHIEVES NANOMETER RESOLUTION
    MORTENSEN, P
    LASER FOCUS WORLD, 1993, 29 (04): : 44 - +
  • [2] DIRECT MEASUREMENT OF STANDING EVANESCENT WAVES WITH A PHOTON-SCANNING TUNNELING MICROSCOPE
    MEIXNER, AJ
    BOPP, MA
    TARRACH, G
    APPLIED OPTICS, 1994, 33 (34): : 7995 - 8000
  • [3] Numerical study of the tip-sample interaction in the photon scanning tunneling microscope
    Weeber, JC
    deFornel, F
    Guodonnet, JP
    OPTICS COMMUNICATIONS, 1996, 126 (4-6) : 285 - 292
  • [4] ELECTROSTATIC SAMPLE-TIP INTERACTIONS IN THE SCANNING TUNNELING MICROSCOPE
    MCELLISTREM, M
    HAASE, G
    CHEN, D
    HAMERS, RJ
    PHYSICAL REVIEW LETTERS, 1993, 70 (16) : 2471 - 2474
  • [5] Calibration of tip and sample temperature of a scanning tunneling microscope using a superconductive sample
    Stocker, Matthias
    Pfeifer, Holger
    Koslowski, Berndt
    JOURNAL OF VACUUM SCIENCE & TECHNOLOGY A, 2014, 32 (03):
  • [6] High-resolution photon-scanning tunneling microscope measurements of the whispering gallery modes in a cylindrical microresonator
    Klunder, DJW
    Balistreri, MLM
    Blom, FC
    Hoekstra, HJWM
    Driessen, A
    Kuipers, L
    van Hulst, NF
    IEEE PHOTONICS TECHNOLOGY LETTERS, 2000, 12 (11) : 1531 - 1533
  • [7] An ultrahigh vacuum cryogenic scanning tunneling microscope with tip and sample exchange
    Harrell, LE
    First, PN
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1999, 70 (01): : 125 - 132
  • [9] THE PHOTON SCANNING TUNNELING MICROSCOPE
    FERRELL, TL
    GOUNDONNET, JP
    REDDICK, RC
    SHARP, SL
    WARMACK, RJ
    JOURNAL OF VACUUM SCIENCE & TECHNOLOGY B, 1991, 9 (02): : 525 - 530
  • [10] SCANNING TUNNELING MICROSCOPE TIP STRUCTURES
    NICOLAIDES, R
    LIANG, Y
    PACKARD, WE
    FU, ZW
    BLACKSTEAD, HA
    CHIN, KK
    DOW, JD
    FURDYNA, JK
    HU, WM
    JAKLEVIC, RC
    KAISER, WJ
    PELTON, AR
    ZELLER, MV
    BELLINA, J
    JOURNAL OF VACUUM SCIENCE & TECHNOLOGY A-VACUUM SURFACES AND FILMS, 1988, 6 (02): : 445 - 447