CT-GUIDED BONE-BIOPSY PERFORMED BY MEANS OF A COAXIAL BIOPSY SYSTEM WITH AN ECCENTRIC DRILL

被引:29
|
作者
AHLSTROM, KH
ASTROM, KGO
机构
[1] Department of Diagnostic Radiology, Akademiska Sjukhuset, University Hospital
关键词
BONE MARROW; BIOPSY; BONE NEOPLASMS; DIAGNOSIS; BONES; INTERVENTIONAL PROCEDURES; TECHNOLOGY;
D O I
10.1148/radiology.188.2.8327713
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Thirty-seven consecutive bone biopsies guided with computed tomography were performed in 32 patients by use of three different techniques to penetrate cortical bone and gain access to the lesion. The following instruments were used: a thin bone biopsy needle (12 biopsies), a conventional drill with an outer cannula (six biopsies), and a coaxial biopsy system that consists of a drill with an eccentric tip and an outer cannula (19 biopsies). This eccentric drill makes a hole in the bone larger than the diameter of the cannula and thereby makes it easy to advance the cannula over the drill until the cannula is anchored in the bone. One can then obtain multiple samples through the cannula. The thin bone biopsy needle could not penetrate thick (8 mm thick) cortical bone. The outer cannula was not anchored in the bone when the conventional drill was used. In 16 biopsies, the new coaxial biopsy system penetrated cortical bone with a thickness of 1-8 mm and was anchored there, and lesion samples were obtained through the anchored cannula.
引用
收藏
页码:549 / 552
页数:4
相关论文
共 50 条
  • [1] CT-GUIDED BONE-BIOPSY WITH A NEW COAXIAL BONE-BIOPSY SYSTEM
    ASTROM, GKO
    AHLSTROM, HK
    RADIOLOGY, 1992, 185 : 116 - 117
  • [2] AUTOMATIC BIOPSY INSTRUMENTS USED THROUGH A COAXIAL BONE-BIOPSY SYSTEM WITH AN ECCENTRIC DRILL TIP
    ASTROM, KGO
    SUNDSTROM, JC
    LINDGREN, PG
    AHLSTROM, KH
    ACTA RADIOLOGICA, 1995, 36 (03) : 237 - 242
  • [3] MR-guided bone biopsy performed with a new coaxial drill system.
    Neuerburg, J
    Adam, G
    Bucker, A
    Zilkens, KW
    Schmitz-Rode, T
    Katterbach, FJ
    Klosterhalfen, B
    Rasmussen, E
    van Vaals, JJ
    Gunther, RW
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 1998, 169 (05): : 515 - 520
  • [4] USE OF CT-GUIDED PERCUTANEOUS BONE-BIOPSY IN STAGING OF GENITOURINARY TUMORS
    GOLIMBU, C
    FIROOZNIA, H
    RAFII, M
    UROLOGY, 1983, 22 (03) : 322 - 325
  • [5] A NEW BONE-BIOPSY DRILL
    LALOR, BC
    FREEMONT, AJ
    ADAMS, PH
    CLINICAL SCIENCE, 1986, 71 : P32 - P32
  • [6] CT-Guided Bone Biopsy With a Battery-Powered Drill System: Preliminary Results
    Lee, Ryan K. L.
    Ng, Alex W. H.
    Griffith, James F.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2013, 201 (05) : 1093 - 1095
  • [7] BIOPSY OF ALL TYPES OF BONE-LESIONS WITH A COAXIAL ECCENTRIC DRILL SYSTEM
    ASTROM, GK
    AHLSTROM, HK
    RADIOLOGY, 1995, 197 : 466 - 466
  • [8] CT-GUIDED RENAL BIOPSY WITH A COAXIAL TECHNIQUE AND AUTOMATED BIOPSY GUN
    KUDRYK, BT
    MARTINEZ, CR
    RADIOLOGY, 1992, 185 : 230 - 230
  • [9] CT-guided needle biopsy performed with modified coaxial technique in the diagnosis of malignant lymphomas
    Pedote, P.
    Gaudio, F.
    Moschetta, M.
    Cimmino, A.
    Specchia, G.
    Angelalli, G.
    RADIOLOGIA MEDICA, 2010, 115 (08): : 1292 - 1303
  • [10] Importance of MRI and CT-guided bone-biopsy in the management of patients with SAPHO-syndrome
    Kirchhoff, TD
    Merkesdal, SA
    Rosenthal, H
    Prokop, M
    Wagner, A
    Galanski, MW
    RADIOLOGY, 1998, 209P : 501 - 501