IMPROVED TRANSJUGULAR INTRAHEPATIC PORTOSYSTEMIC SHUNT PATENCY WITH PTFE-COVERED STENT-GRAFTS - EXPERIMENTAL RESULTS IN SWINE

被引:114
|
作者
NISHIMINE, K
SAXON, RR
KICHIKAWA, K
MENDELHARTVIG, J
TIMMERMANS, HA
SHIM, HJ
UCHIDA, BT
BARTON, RE
KELLER, FS
ROSCH, J
机构
[1] OREGON HLTH SCI UNIV,SOTTER INTERVENT INST,PORTLAND,OR 97201
[2] NARA MED UNIV,DEPT RADIOL,NARA,JAPAN
[3] NARA MED UNIV,DEPT ONCORADIOL,NARA,JAPAN
[4] CHUNG ANG UNIV,DEPT RADIOL,SEOUL 156756,SOUTH KOREA
关键词
LIVER; INTERVENTIONAL PROCEDURE; SHUNTS; PORTOSYSTEMIC; STENTS AND PROSTHESES;
D O I
10.1148/radiology.196.2.7617843
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To evaluate the ability of stent-grafts made with polytetrafluoroethylene (PTFE) graft material to improve transjugular intrahepatic portosystemic shunt (TIPS) patency. MATERIALS AND METHODS: TIPS were created in 13 swine by using PTFE-covered stent-grafts. Uncovered TIPS were placed in 13 other swine. Twenty-one of 26 animals were followed up with portal venography for 3 months or until the shunt became severely stenotic. Five animals without severe stenosis were sacrificed before 3 months because of illness. RESULTS: At 4 weeks after TIPS placement, nine of 13 stent-graft TIPS were patent (<50% diameter narrowing) compared with only one patent stent in 13 uncovered TIPS. Six of 13 stent-graft TIPS remained patent until the animals were sacrificed. In 11 of 12 uncovered TIPS, stenosis was most prominent in the parenchymal tract. In five of seven stent-graft TIPS, stenosis was most prominent in the hepatic vein above the end of the graft material. Bile leaks were discovered in six occluded uncovered TIPS and in two of the stent-graft TIPS. CONCLUSION: PTFE-covered stent-grafts significantly improved TIPS patency in swine (P < .01). However, stenosis in the hepatic vein led to late shunt malfunction in selected cases.
引用
收藏
页码:341 / 347
页数:7
相关论文
共 50 条
  • [1] Comparison of transjugular intrahepatic portosystemic shunt dysfunction in PTFE-covered stent-grafts versus bare stents
    Barrio, J
    Ripoll, C
    Bañares, R
    Echenagusia, A
    Catalina, MV
    Camúñez, F
    Simó, G
    Santos, L
    EUROPEAN JOURNAL OF RADIOLOGY, 2005, 55 (01) : 120 - 124
  • [2] Expanded PTFE-covered stent-grafts in the treatment of transjugular intrahepatic portosystemic shunt (TIPS) stenoses and occlusions
    Echenagusia, M
    Rodriguez-Rosales, G
    Simo, G
    Camuñez, F
    Bañares, R
    Echenagusia, A
    ABDOMINAL IMAGING, 2005, 30 (06): : 750 - 754
  • [3] Expanded PTFE-covered stent-grafts in the treatment of transjugular intrahepatic portosystemic shunt (TIPS) stenoses and occlusions
    M. Echenagusia
    G. Rodriguez-Rosales
    G. Simo
    F. Camuñez
    R. Bañares
    A. Echenagusia
    Abdominal Imaging, 2005, 30 : 750 - 754
  • [4] Hepatic encephalopathy is more frequent and severe after transjugular intrahepatic portosystemic shunt (TIPS) with PTFE-covered stent-grafts
    Riggio, Oliviero
    Angeloni, Stefania
    Nicolao, Francesca
    Ridola, Lorenzo
    Salvatori, Filippo M.
    Attili, Adolfo F.
    Merli, Manuela
    HEPATOLOGY, 2006, 44 (04) : 232A - 232A
  • [5] Hepatic encephalopathy after transjugular intrahepatic portosystemic shunt (TIPS): PTFE-covered stent-grafts versus bare stents
    Angeloni, S.
    Nicolini, G.
    Nicolao, F.
    Ridola, L.
    Salvatori, F. M.
    Fanelli, F.
    Attili, A. F.
    Merli, M.
    Riggio, O.
    JOURNAL OF HEPATOLOGY, 2006, 44 : S84 - S84
  • [6] Hepatic encephalopathy after transjugular intrahepatic portosystemic shunt (TIPS): PTFE-covered stent-grafts versus bare stents
    Riggio, O
    Angeloni, S
    Nicolao, F
    Salvatori, F
    Fanelli, F
    Pepino, D
    Attili, AF
    Rossi, P
    Merli, M
    GASTROENTEROLOGY, 2004, 126 (04) : A679 - A680
  • [7] Improved patency of transjugular intrahepatic portosystemic shunts in humans: Creation and revision with PTFE stent-grafts
    Haskal, ZJ
    RADIOLOGY, 1999, 213 (03) : 759 - 766
  • [8] Transjugular intrahepatic portosystemic stent-shunt utilizing PTFE covered stents: Improved survival and shunt patency.
    Tripathi, D
    Ferguson, JW
    Barkell, H
    Macbeth, K
    Ireland, H
    Redhead, DN
    Hayes, PC
    HEPATOLOGY, 2004, 40 (04) : 628A - 628A
  • [9] ePTFE-covered stent-grafts for revision of obstructed transjugular intrahepatic portosystemic shunt
    Cejna, M
    Peck-Radosavljevic, M
    Thurnher, S
    Schoder, M
    Rand, T
    Angermayr, B
    Lammer, J
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2002, 25 (05) : 365 - 372
  • [10] ePTFE-Covered Stent-Grafts for Revision of Obstructed Transjugular Intrahepatic Portosystemic Shunt
    Manfred Cejna
    Markus Peck-Radosavljevic
    Siegfried Thurnher
    Maria Schoder
    Thomas Rand
    Bernhard Angermayr
    Johannes Lammer
    CardioVascular and Interventional Radiology, 2002, 25 : 365 - 372