AN ANALYTICAL APPROACH FOR COMPENSATION OF NONUNIFORM ATTENUATION IN CARDIAC SPECT IMAGING

被引:8
|
作者
GLICK, SJ
KING, MA
PAN, TS
SOARES, EJ
机构
[1] Massachusetts Univ., Med. Center, Worcester, MA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 1995年 / 40卷 / 10期
关键词
D O I
10.1088/0031-9155/40/10/009
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Photon attenuation can reduce the diagnostic accuracy of cardiac SPECT imaging. Bellini et nl have previously derived a mathematically exact method to compensate for attenuation in a uniform attenuator. Since the human thorax contains structures with differing attenuation properties, non-uniform attenuation compensation is required in cardiac SPECT. Given an estimate of the patient attenuation map, we show that the Bellini attenuation compensation method can be used in cardiac SPECT to provide a quantitatively accurate reconstruction of a central region in the image which includes the heart and surrounding soft tissue. Simulations using a mathematical cardiac-torso phantom were conducted to evaluate the Bellini method and to compare its performance to the ML-EM iterative algorithm, and to 180 degrees and 360 degrees filtered backprojection (FBP) with no attenuation compensation. 'Bulls-eye' polar maps and circumferential profiles showed that both the Bellini method and the ML-EM algorithm provided quantitatively accurate reconstructions of the myocardium, with a substantial reduction in attenuation-induced artifacts that were observed in the FBP images. The computational load required to implement the Bellini method is approximately equivalent to that required for one iteration of the ML-EM algorithm, thus it is suitable for routine clinical use.
引用
收藏
页码:1677 / 1693
页数:17
相关论文
共 50 条
  • [21] A SIMULATION OF EMISSION AND TRANSMISSION NOISE-PROPAGATION IN CARDIAC SPECT IMAGING WITH NONUNIFORM ATTENUATION CORRECTION
    TUNG, CH
    GULLBERG, GT
    MEDICAL PHYSICS, 1994, 21 (10) : 1565 - 1576
  • [22] Compensation for attenuation and 2D detector response in cardiac SPECT imaging.
    Glick, SJ
    King, MA
    JOURNAL OF NUCLEAR MEDICINE, 1996, 37 (05) : 61 - 61
  • [23] INITIAL EXPERIENCE WITH ATTENUATION COMPENSATION IN CLINICAL CARDIAC SPECT
    BUJENOVIC, LS
    TSUI, BMW
    LALUSH, DS
    FREY, EC
    JOHNSTON, RE
    MCCARTNEY, WH
    JOURNAL OF NUCLEAR MEDICINE, 1994, 35 (05) : P91 - P91
  • [24] Significance of nonuniform attenuation correction in quantitative brain SPECT imaging
    Rajeevan, N
    Zubal, IG
    Ramsby, SQ
    Zoghbi, SS
    Seibyl, J
    Innis, RB
    JOURNAL OF NUCLEAR MEDICINE, 1998, 39 (10) : 1719 - 1726
  • [25] Compensation for nonuniform attenuation in GA-67 SPECT without a measured attenuation map.
    Moore, SC
    Müller, SP
    Kijewski, MF
    JOURNAL OF NUCLEAR MEDICINE, 1998, 39 (05) : 36P - 36P
  • [26] The influence of attenuation, scatter, and spatial resolution compensation on diagnostic performance of cardiac SPECT/CT imaging
    Naum, A.
    Bach-Gansmo, T.
    Kleven-Madsen, N.
    Biermann, M.
    Johnsen, B.
    Husby, J. Aase
    Rotevatn, S.
    Nordrehaug, J. E.
    EUROPEAN HEART JOURNAL SUPPLEMENTS, 2011, 13 (0A) : A8 - A8
  • [27] Attenuation compensation of cardiac SPECT: A critical look at a confusing world
    Wackers, FJT
    JOURNAL OF NUCLEAR CARDIOLOGY, 2002, 9 (04) : 438 - 440
  • [28] Attenuation compensation of cardiac SPECT: A critical look at a confusing world
    Frans J. Th. Wackers
    Journal of Nuclear Cardiology, 2002, 9 : 438 - 440
  • [29] Evaluation of nonuniform attenuation correction in quantitative brain SPECT imaging.
    Rajeevan, N
    Innis, RB
    Ramsby, Q
    Zoghbi, SS
    Seibyl, JP
    Zubal, IG
    JOURNAL OF NUCLEAR MEDICINE, 1997, 38 (05) : 909 - 909
  • [30] AN ANALYTICAL APPROACH TO QUANTITATIVE RECONSTRUCTION OF NONUNIFORM ATTENUATED BRAIN SPECT
    LIANG, ZR
    YE, JH
    HARRINGTON, DP
    PHYSICS IN MEDICINE AND BIOLOGY, 1994, 39 (11): : 2023 - 2041