ONLINE DETERMINATION OF RESIDUAL ALUMINUM IN POTABLE AND TREATED WATERS BY FLOW-INJECTION ANALYSIS

被引:36
|
作者
BENSON, RL [1 ]
WORSFOLD, PJ [1 ]
SWEETING, FW [1 ]
机构
[1] WESSEX WATER,SALTFORD LAB,BRISTOL BS18 3ER,ENGLAND
关键词
Aluminium; Waters;
D O I
10.1016/S0003-2670(00)80535-8
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A portable, automated, flow-injection-based solid-state spectrophotometric field monitor for the determination of residual dissolved aluminium in potable and treated waters is described. Aluminium is determined by complexation with pyrocatechol violet to form a colloidal lake with a maximum absorbance at 580 nm. Interference from iron(III) is masked by reduction to iron(II) and complexation with 1,10-phenanthroline. The monitor operates on a 30-min analytical cycle, i.e., with 48 results per day, and each cycle includes an autocalibration step. Over the course of a 3-month field trial the reliability was greater than 90%. The performance conforms to industry specifications for both overall accuracy (bias < ± 10%) and repeatability (relative standard deviation < ± 10%), the linear range is 0-1000 μg l-1 and the practical limit of detection is 45 μg l-1. © 1990.