SKIN IMPEDANCE MEASUREMENTS USING SIMPLE AND COMPOUND ELECTRODES

被引:40
|
作者
WOO, EJ
HUA, P
WEBSTER, JG
TOMPKINS, WJ
PALLASARENY, R
机构
[1] UNIV WISCONSIN, DEPT ELECT & COMP ENGN, 1415 JOHNSON DR, MADISON, WI 53706 USA
[2] KONKUK UNIV DANWO DONG 322, COLL MED, DEPT BIOMED ENGN, CHOONGJU 380701, SOUTH KOREA
[3] SIEMENS GAMMASON INC, APPL RES GRP, HOFFMAN ESTATES, IL 60195 USA
[4] ETS ING TELECOM, DIV INSTRUMENTAC & BIOENGN, E-08080 BARCELONA, SPAIN
关键词
COMPOUND ELECTRODE; ELECTRODE CONFIGURATION; SKIN IMPEDANCE;
D O I
10.1007/BF02446200
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
We have studied the effect of the electrode configuration on the measurement of body impedance and found that the electrode configuration greatly affects the impedance measurement using the four-electrode method. We studied the characteristics of the compound electrode and found that the compound electrode provides the four-electrode method in a compact form. A new method of measuring the skin impedance using simple electrodes at low frequencies was developed. At high frequencies where the effect of internal tissue impedance is not negligible, we used the compensation method using compound electrodes, because they measure the voltage right under the skin. At 50 kHz, we measured the real part of the skin impedance of less than 80-OMEGA on the thorax. We propose a simple instrument which can measure accurate skin impedance at various frequencies.
引用
收藏
页码:97 / 102
页数:6