CALIBRATION OF RF FIELD-STRENGTH METERS

被引:6
|
作者
DEVREEDE, JPM
TEUNISSE, GM
MUNTER, K
HOITINK, H
HOLLAND, K
ODWYER, M
BERGSTEN, L
PURANEN, L
机构
[1] PHYS TECH BUNDESANSTALT,D-38023 BRAUNSCHWEIG,GERMANY
[2] ABB RES CTR,EMC BADEN LTD,CH-5405 BADEN DATWILL,SWITZERLAND
[3] NATL PHYS LAB,TEDDINGTON TW11 0LW,MIDDX,ENGLAND
[4] RADIO FREQUENCY TECHNOL,DUBLIN 8,IRELAND
[5] SWEDISH NATL TESTING & RES INST,S-50115 BORAS,SWEDEN
[6] FINNISH CTR RADIAT & NUCL SAFETY,SF-00881 HELSINKI,FINLAND
关键词
D O I
10.1109/19.377850
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
An intercomparison of field strength meters was performed during 1991-1993 over the frequency range 100 kHz-1000 MHz and field strength range 1-10 V/m using different calibration techniques. Four field sensors, electric as well as magnetic, have been used. Seven institutes participated in this intercomparison using at least three different calibration methods. A calibration uncertainty of less than 1 dB can be obtained if certain procedures are followed. The results are applicable in the fields of EMC testing and health care (nonionizing radiation).
引用
收藏
页码:351 / 354
页数:4
相关论文
共 50 条