The Active Tube Clearance System A Novel Bedside Chest-Tube Clearance Device

被引:0
|
作者
Shalli, Shanaz [1 ]
Boyle, Edward M. [2 ]
Saeed, Diyar [1 ]
Fukamachi, Kiyotaka [1 ]
Cohn, William E. [3 ]
Gillinov, A. Marc [4 ]
机构
[1] Cleveland Clin, Lerner Res Inst, Dept Biomed Engn ND20, Cleveland, OH 44106 USA
[2] St Charles Hosp, Bend, OR USA
[3] Texas Heart Inst, Dept Cardiovasc Surg, Houston, TX 77025 USA
[4] Cleveland Clin, Heart & Vasc Inst, Dept Thorac & Cardiovasc Surg, Cleveland, OH 44106 USA
关键词
Hemothorax; Clogging; Pneumothorax; Outcome; Chest tube; Drainage; Occlusion; Thoracostomy;
D O I
10.1097/imi.0b013e3181cf7ce3
中图分类号
R61 [外科手术学];
学科分类号
摘要
Objective: Chest-tube clogging can lead to complications after heart and lung surgery. Surgeons often choose large-diameter chest tubes or place more than one chest tube when concerned about the potential for clogging. The purpose of this report is to describe the design and function of a proprietary active tube clearance system, a novel device that clears clots and debris from chest tubes. Device Description: The active tube clearance system is a novel chest tube clearance apparatus developed to maintain chest tube patency. Chest tube clearance is achieved by advancing the specially designed clearance member back and forth within the chest tube under sterile conditions, breaking down and pulling clots back toward the drainage receptacle, thereby leaving the inner portion of the chest tube clear of any obstructing material. Conclusions: By maintaining chest tube patency, chest tube drainage can be performed more safely, and this apparatus may possibly lead to the use of smaller chest tubes and less invasive insertion techniques.
引用
收藏
页码:42 / 47
页数:6
相关论文
共 50 条
  • [21] HEART PERFORATION - A RARE COMPLICATION AFTER CHEST-TUBE TREATMENT
    FERNANDEZ, ED
    NEUDECK, F
    PIOTROWSKI, J
    CHIRURG, 1995, 66 (09): : 920 - 922
  • [22] Costs associated with chest-tube placement for pneumothorax complicating percutaneous chest biopsy
    Goodacre, BW
    Ozkan, O
    Wittich, GR
    Kavanagh, PV
    Zwischenberger, JB
    VanSonnenberg, E
    RADIOLOGY, 1997, 205 : 140 - 140
  • [23] Acute diaphragmatic paralysis caused by chest-tube trauma to phrenic nerve
    E. Nahum
    Joseph Ben-Ari
    Tommy Schonfeld
    Gadi Horev
    Pediatric Radiology, 2001, 31 : 444 - 446
  • [24] EFFECT OF CHEST-TUBE SUCTION ON GAS-FLOW THROUGH A BRONCHOPLEURAL FISTULA
    POWNER, DJ
    CLINE, CD
    RODMAN, GH
    CRITICAL CARE MEDICINE, 1985, 13 (02) : 99 - 101
  • [25] Buffalo Chest-Tube Thoracostomy Gone Too Far: A Case Report
    Samuel, A.
    Mahmood, N.
    AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2020, 201
  • [26] Acute diaphragmatic paralysis caused by chest-tube trauma to phrenic nerve
    Nahum, E
    Ben-Ari, J
    Schonfeld, T
    Horev, G
    PEDIATRIC RADIOLOGY, 2001, 31 (06) : 444 - 446
  • [27] Planar elastica inside a curved tube with clearance
    Chen, Jen-San
    Li, Chia-Wei
    INTERNATIONAL JOURNAL OF SOLIDS AND STRUCTURES, 2007, 44 (18-19) : 6173 - 6186
  • [28] PUMPING AND CLEARANCE FUNCTION OF THE EUSTACHIAN-TUBE
    HONJO, I
    HAYASHI, M
    ITO, S
    TAKAHASHI, H
    AMERICAN JOURNAL OF OTOLARYNGOLOGY, 1985, 6 (03) : 241 - 244
  • [29] AN AUDIT OF WARD-BASED CHEST-TUBE INSERTION AT A TERTIARY TEACHING HOSPITAL
    Bedecarratz, Nicole Lafontaine
    Joosten, Simon
    Irving, Louis
    Hew, Mark
    RESPIROLOGY, 2012, 17 : 23 - 23
  • [30] Modeling Tube Clearance and Bounding the Effect of Friction in Concentric Tube Robot Kinematics
    Ha, Junhyoung
    Fagogenis, Georgios
    Dupont, Pierre E.
    IEEE TRANSACTIONS ON ROBOTICS, 2019, 35 (02) : 353 - 370