The quality assurance of volumetric modulated arc therapy (VMAT) plans for early stage prostate cancer: a technical note

被引:3
|
作者
Elith, Craig A. [1 ,2 ]
Dempsey, Shane E. [2 ]
Cao, Fred [1 ]
Farshadi, Afrooz [1 ]
Warren-Forward, Helen M. [2 ]
机构
[1] British Columbia Canc Agcy, Fraser Valley Ctr, Surrey, BC, Canada
[2] Univ Newcastle, Sch Hlth Sci, Callaghan, NSW, Australia
来源
JOURNAL OF MEDICAL RADIATION SCIENCES | 2014年 / 61卷 / 04期
关键词
ArcCHECK; IMSure; quality assurance; VMAT;
D O I
10.1002/jmrs.78
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
As radiation therapy transitions from intensity modulated radiation therapy (IMRT) to volumetric modulated arc therapy (VMAT) it is important to consider the quality assurance (QA) of VMAT plans in light of what has previously been learned and developed in IMRT QA. This technical note assesses if IMRT based plan QA software, which has reduced the need in IMRT for phantom dose measurements on the linear accelerator, can be incorporated into VMAT QA processes. Twenty prostate cases were retrospectively planned using VMAT with one arc to deliver a prescription of 74 Gy in 37 fractions. A plan QA was performed using both IMSure (version 3.3), a software-based IMRT QA program, and ArcCHECK (version 6.2.3.5713), a phantom-based VMAT QA tool. Outcomes assessed included the time needed to perform the QA of both the IMSure and ArcCHECK QA methods, and agreement between planned dose and QA measured dose. On average per case, the ArcCHECK technique needed 31.5 min to perform the VMAT plan QA, while IMSure required 3.5 min to perform the same QA. All 20 cases passed dosimetric QA using ArcCHECK. However, using IMSure, three cases failed dosimetric QA using the departments existing IMRT QA criteria. This research has demonstrated that the IMRT QA software IMSure may be incorporated into the QA of VMAT plans, however the criteria to assess the dosimetry of the VMAT plans may need to be different to that for IMRT cases. The implication of this research for radiation therapists is to be critically aware of the differences between the plan QA requirements and methods for IMRT and those required for VMAT.
引用
收藏
页码:261 / 266
页数:6
相关论文
共 50 条
  • [21] Evidence based quality assurance of volumetric modulated arc therapy
    Tuemmler, H.
    Bellmann, C.
    Schilling, P.
    Schreiber, A.
    RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S117 - S117
  • [22] COMMISSIONING OF VOLUMETRIC MODULATED ARC THERAPY (VMAT)
    Bedford, James L.
    Warrington, Alan P.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 73 (02): : 537 - 545
  • [23] Image-Guided Volumetric Modulated Arc Therapy (IG-VMAT) For Prostate Cancer
    Song, Yulin
    Mueller, Boris
    Obcemea, Ceferino
    Mychalczak, Borys
    5TH EUROPEAN CONFERENCE OF THE INTERNATIONAL FEDERATION FOR MEDICAL AND BIOLOGICAL ENGINEERING, PTS 1 AND 2, 2012, 37 : 223 - +
  • [24] Comparison of Dynamic Volumetric Modulated Arc Therapy (VMAT) versus Static IMRT for Early-stage Larynx Cancer
    Riegel, A. C.
    Antone, J.
    Potters, L.
    Schwartz, D. L.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2011, 81 (02): : S520 - S520
  • [25] Comparisons of volumetric modulated arc therapy (VMAT) quality assurance (QA) systems: sensitivity analysis to machine errors
    Liang, Bin
    Liu, Bo
    Zhou, Fugen
    Yin, Fang-Fang
    Wu, Qiuwen
    RADIATION ONCOLOGY, 2016, 11
  • [26] Comparisons of volumetric modulated arc therapy (VMAT) quality assurance (QA) systems: sensitivity analysis to machine errors
    Bin Liang
    Bo Liu
    Fugen Zhou
    Fang-fang Yin
    Qiuwen Wu
    Radiation Oncology, 11
  • [27] Patient Specific Quality Assurance of Volumetric Modulated Arc Therapy (VMAT): Portal Dosimetry Compared with ArcCheck Phantom
    Utitsarn, K.
    Watthanasarn, T.
    Pimthong, J.
    Krongkietlearts, K.
    Pihusut, C.
    Chaloemchawalit, W.
    Jitmon, J.
    Khuanchana, S.
    MEDICAL PHYSICS, 2020, 47 (06) : E826 - E826
  • [28] A retrospective analysis for patient-specific quality assurance of volumetric-modulated arc therapy plans
    Li, Guangjun
    Wu, Kui
    Peng, Guang
    Zhang, Yingjie
    Bai, Sen
    MEDICAL DOSIMETRY, 2014, 39 (04) : 309 - 313
  • [29] Volumetric modulated arc therapy (VMAT) in the treatment of oesophageal cancer patients
    Martini, S.
    Iorio, G. C.
    Arcadipane, F.
    Franco, P.
    Ricardi, U.
    RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S776 - S777
  • [30] Impact of Parameters for Volumetric Modulated Arc Therapy (VMAT) for Rectal Cancer
    Runxiao, L.
    Yankun, C.
    Rong, Q.
    Yuzhi, S.
    Kai, S.
    Mingchang, M.
    MEDICAL PHYSICS, 2015, 42 (06) : 3439 - 3439