SOMATOSENSORY EVOKED-POTENTIAL MONITORING DURING INTRACRANIAL SURGERY

被引:13
|
作者
DJURIC, S [1 ]
MILENKOVIC, Z [1 ]
KLOPCICSPEVAK, M [1 ]
SPASIC, M [1 ]
机构
[1] UNIV NOVI SAD, CTR CLIN, MED REHABIL CLIN, NEUROL & NEUROSURG CLIN, YU-21000 NOVI SAD, YUGOSLAVIA
关键词
INTRACRANIAL ANEURYSM; INTRAOPERATIVE MONITORING; SOMATOSENSORY EVOKED POTENTIALS; CENTRAL CONDUCTION TIME;
D O I
10.1007/BF01541787
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
In the neurosurgical approach to intracranial aneurysms which are often accompanied by arterial spasm and cortical ischaemia, monitoring procedures aim to obtain useful information on cerebral function. SEPs evoked by stimulation of the median nerve at the wrist and of the tibial nerve at the medial malleolus were registered in 45 patients with intracranial aneurysms during neurosurgical procedures. Our results show SEP abnormalities during different stages of neurosurgical procedures in 36 patients out of the monitored 45. Significant abnormalities of SEPs with respect to the control group were decrease of the amplitude of N 20-P 25 complex, lengthening of the absolute latency of the waves N 20-and P 25 and lengthening of the central conduction time (CCT) (N 13-N 20). The greatest SEP abnormalities were registered during the neurosurgical approach to aneurysm and during the clipping procedure. However, the changes were reversible in the majority of the patients. The aim of this paper was to focus on early detection of some cerebral function disturbances during the neurosurgical procedure as well as the prevention of possible brain damage.