摘要
Simulations using Monte Carlo packages are increasingly used in nuclear imaging for positron emission tomographic and single-photon emission computed tomographic applications, either for modeling imaging systems or developing algorithms for analysis improvement of image quantifications. This work implemented a GEANT4-based single- photon emission computed tomographic simulation system, GATE, to evaluate the performance of two scatter correction methods: the triple-energy window method using trapezoidal approximation and the triple-energy window method using triangular approximation. A cold-spot hot-background phantom and a resolution phantom were simulated to generate projection images for scatter correction. Both methods were not found to have strictly valid assumptions. A comparative assessment of these two methods was made based on image contrast improvement, image mottle level, and image resolution improvement. Results showed that triangular approximation was better than trapezoidal approximation for contrast improvement with greater image mottle level and some improvement in the resolution of the resolution phantom for both approximations, but trapezoidal approximation with these criteria was better than triangular approximation. A compromise between the contrast improvement and the image mottle level exists. Neither of the two methods performed best for all three criteria.
引用
收藏
页码:317 / 324
页数:8
相关论文
共 50 条
  • [1] Optimization of scatter correction method in samarium-153 single-photon emission computed tomography using triple-energy window: A monte carlo simulation study
    Asmi, Hicham
    Bentayeb, Farida
    Bouzekraoui, Youssef
    Bonutti, Faustino
    INDIAN JOURNAL OF NUCLEAR MEDICINE, 2020, 35 (03): : 210 - 215
  • [2] Comparison of the dual-energy window method and the triple-energy window method for quantitative single photon emission computed tomography
    Ito, Y.
    Kuniyoshi, M.
    Suetake, M.
    Mochizuki, K.
    Matsumoto, K.
    Endo, K.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2016, 43 : S647 - S647
  • [3] Determination of the energy windows for the triple energy window scatter correction method in Gadolinium-159 single photon emission computed tomography using Monte Carlo simulation
    Bouzekraoui Y.
    Bentayeb F.
    Asmi H.
    Bonutti F.
    Iranian Journal of Medical Physics, 2019, 16 (06) : 405 - 409
  • [4] Clinical value of triple-energy window scatter correction in simultaneous dual-isotope single-photon emission tomography with I-123-BMIPP and Tl-201
    Yang, JT
    Yamamoto, K
    Sadato, N
    Tsuchida, T
    Takahashi, N
    Hayashi, N
    Yonekura, Y
    Ishii, Y
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1997, 24 (09): : 1099 - 1106
  • [5] Clinical value of triple-energy window scatter correction in simultaneous dual-isotope single-photon emission tomography with 123I-BMIPP and 201TI
    Yang J.-T.
    Yamamoto K.
    Sadato N.
    Tsuchida T.
    Takahashi N.
    Hayashi N.
    Yonekura Y.
    Ishii Y.
    European Journal of Nuclear Medicine, 1997, 24 (9) : 1099 - 1106
  • [6] Attenuation correction single-photon emission computed tomography myocardial perfusion imaging
    Bateman, TM
    Cullom, SJ
    SEMINARS IN NUCLEAR MEDICINE, 2005, 35 (01) : 37 - 51
  • [7] Positron emission tomography and single-photon emission computed tomography brain imaging in the evaluation of dementia
    Van Heertum, RL
    Tikofsky, RS
    SEMINARS IN NUCLEAR MEDICINE, 2003, 33 (01) : 77 - 85
  • [8] BONE IMAGING USING SINGLE-PHOTON EMISSION COMPUTED-TOMOGRAPHY
    GATES, GF
    WESTERN JOURNAL OF MEDICINE, 1994, 161 (04): : 410 - 411
  • [9] Parathyroid imaging using single-photon emission computed tomography/computed tomography: single institution experience
    Guerrouj, H.
    Benrais, N.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2012, 39 : S287 - S287
  • [10] Energy Window and Contrast Optimization for Single-photon Emission Computed Tomography Bremsstrahlung Imaging with Yttrium-90
    Bouzekraoui, Youssef
    Bentayeb, Farida
    Asmi, Hicham
    Bonutti, Faustino
    INDIAN JOURNAL OF NUCLEAR MEDICINE, 2019, 34 (02): : 125 - 128