Case Report: Inappropriate Implantable Cardioverter Defibrillator Discharge from Sensing of External Alternating Current Leak

被引:0
|
作者
Ashok Garg
Manish Wadhwa
Keith Brown
Cathy Luckett
Tim Vaughn
Ulrika Birgersdotter-Green
Gregory Feld
机构
[1] University of California,Department of Medicine, Cardiology Division, Cardiac Electrophysiology Program
[2] School of Medicine,undefined
关键词
implantable defibrillator; sensing; ventricular fibrillation; electromagnetic interference; alternating current;
D O I
暂无
中图分类号
学科分类号
摘要
Implantable cardioverter defibrillators (ICDs) are now an accepted and effective therapy for treatment of survivors of sudden cardiac death (SCD) and prevention of SCD in high-risk patients. Normal ICD function and delivery of therapy depends on appropriate sensing and detection of myocardial electrical potentials. Electromagnetic interference resulting in ICD malfunction is a well-documented phenomenon, however, there are less well-known external sources of interference, which may cause life threatening ICD malfunction. We report a unique case of repeated inappropriate ICD shocks in a ten-year old boy caused by the ICD sensing alternating current from an unexpected external source.
引用
收藏
页码:181 / 184
页数:3
相关论文
共 50 条
  • [1] Inappropriate implantable cardioverter defibrillator discharge from sensing of external alternating current leak
    Garg, A
    Wadhwa, M
    Brown, K
    Luckett, C
    Vaughn, T
    Birgersdotter-Green, U
    Feld, G
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2002, 7 (02) : 181 - 184
  • [2] Inappropriate implantable cardioverter-defibrillator shock due to external alternating current leak: Report of two cases
    Chan, NY
    Ho, LLWL
    EUROPACE, 2005, 7 (02): : 193 - 196
  • [3] TERMINATION OF ELECTROCUTION INJURY WITH AN INAPPROPRIATE IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR SHOCK DUE TO EXTERNAL ALTERNATING CURRENT LEAK
    Yetis, Begum
    Atar, Ilyas
    Ozin, Bulent
    Muderrisoglu, Haldun
    INTERNATIONAL JOURNAL OF CARDIOLOGY, 2010, 140 : S90 - S90
  • [4] Inappropriate Implantable Cardioverter-Defibrillator Shocks Attributed to Alternating-Current Leak in a Swimming Pool
    Makaryus, John N.
    Angert-Gilman, Julia
    Yacoub, Mena
    Patel, Apoor
    Goldner, Bruce
    TEXAS HEART INSTITUTE JOURNAL, 2014, 41 (01) : 61 - 63
  • [5] Implantable cardioverter defibrillator - An unusual case of inappropriate discharge during showering
    Manolis, AG
    Katsivas, AG
    Vassilopoulos, CV
    Louvros, NE
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2000, 4 (01) : 265 - 268
  • [6] Implantable Cardioverter Defibrillator—An Unusual Case of Inappropriate Discharge During Showering
    Athanasios G. Manolis
    Apostolos G. Katsivas
    Charis V. Vassilopoulos
    Nikolaos E. Louvros
    Journal of Interventional Cardiac Electrophysiology, 2000, 4 : 265 - 268
  • [7] Case Report: Inappropriate Discharge of an Implantable Cardioverter-Defibrillator Caused by the Combined Count Criterion
    Wolpert C.
    Jung W.
    Spehl S.
    Korte T.
    Lüderitz B.
    Journal of Interventional Cardiac Electrophysiology, 1997, 2 (1) : 53 - 56
  • [8] Inappropriate sensing of pectoral myopotentials by an implantable cardioverter defibrillator system
    Karanam, S
    John, RM
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2004, 27 (05): : 688 - 689
  • [9] Troubleshooting of inappropriate therapies in patients with implantable cardioverter-defibrillator - a case report
    Walter, T
    Kluge, P
    Neugebauer, A
    Pfeiffer, D
    HERZ KREISLAUF, 1997, 29 (10): : 293 - 296
  • [10] Inappropriate shock delivery by implantable cardioverter defibrillator due to electrical leak from bathroom tap
    Camsari, Ahmet
    Yilmaz, Dilek Cicek
    Ozcan, Turkay
    CLINICAL RESEARCH IN CARDIOLOGY, 2010, 99 (11) : 761 - 763