Case Report: Inappropriate Discharge of an Implantable Cardioverter-Defibrillator Caused by the Combined Count Criterion

被引:0
|
作者
Wolpert C. [1 ,2 ]
Jung W. [1 ]
Spehl S. [1 ]
Korte T. [1 ]
Lüderitz B. [1 ]
机构
[1] Department of Medicine-Cardiology, University of Bonn, Bonn
[2] Dept. of Medicine-Cardiology, University of Bonn, 53105 Bonn
关键词
Combined count; Defibrillator; Inappropriate;
D O I
10.1023/A:1009716907852
中图分类号
学科分类号
摘要
We describe a rare case of an inappropriate defibrillator discharge during a sinus tachycardia with single fast sensed beats fulfilling the ventricular fibrillation rate criterion, activating the "combined count criterion," although the prevalent sinus tachycardia was inside a ventricular tachycardia rate zone, programmed to monitor only mode. The "combined count criterion" is a safety feature that ensures defibrillation therapy in case of a tachyarrhythmia inside the ventricular tachycardia detection zone, that is temporarily crossing the ventricular fibrillation detection rate. This case reports shows that there is a low but unpredictable risk of device therapy even in a zone programmed to monitor only mode. This phenomenon should be considered when a monitor only zone is installed for revelation of slower tachyarrhythmias.
引用
收藏
页码:53 / 56
页数:3
相关论文
共 50 条
  • [1] Inappropriate discharge of an implantable cardioverter-defibrillator caused by the combined count criterion
    Wolpert, C
    Jung, W
    Spehl, S
    Korte, T
    Lüderitz, B
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 1998, 2 (01) : 53 - 56
  • [2] Troubleshooting of inappropriate therapies in patients with implantable cardioverter-defibrillator - a case report
    Walter, T
    Kluge, P
    Neugebauer, A
    Pfeiffer, D
    HERZ KREISLAUF, 1997, 29 (10): : 293 - 296
  • [3] Inappropriate Shocks With Subcutaneous Implantable Cardioverter-Defibrillator in a Young Patient: A Case Report
    Khan, Zahid
    Sethumadhavan, Dinesh
    Rayner, Tom
    Kyaw, Sithu
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2023, 15 (02)
  • [4] Inappropriate therapy of an implantable cardioverter-defibrillator - is it always an undesired event? A case report
    Krzyzanowski, Krystian
    Michalkiewicz, Dariusz
    Wierzbowski, Robert
    Gnilka, Anna
    Ryczek, Robert
    Orski, Zbigniew
    Kubik, Leszek
    KARDIOLOGIA POLSKA, 2009, 67 (08) : 927 - 931
  • [5] ERRONEOUS DISCHARGE OF AN IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR CAUSED BY AN ELECTRIC RAZOR
    SEIFERT, T
    BLOCK, M
    BORGGREFE, M
    BREITHARDT, G
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1995, 18 (08): : 1592 - 1594
  • [6] PREDICTORS OF INAPPROPRIATE IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR DISCHARGES
    Catanzaro, John N.
    Makaryus, Amgad
    Makaryus, John
    Sison, Cristina
    Vavasis, Christos
    Donaldson, David
    Altman, Erik
    Beldner, Stuart
    Jadonath, Ram
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2011, 57 (14) : E172 - E172
  • [7] Implantable Cardioverter-Defibrillator Shock: Appropriate or Inappropriate?
    Nault, Michael A.
    McIntyre, William F.
    Simpson, Christopher S.
    Redfearn, Damian P.
    Abdollah, Hoshiar
    Brennan, F. James
    Baranchuk, Adrian
    ANNALS OF NONINVASIVE ELECTROCARDIOLOGY, 2010, 15 (02) : 181 - 183
  • [8] Duration of discharge of neuromuscular incapacitating device and inappropriate implantable cardioverter-defibrillator detections
    Calton, Rajneesh
    Cameron, Douglas
    Masse, Stephane
    Nanthakumar, Kumaraswamy
    CIRCULATION, 2007, 115 (20) : E472 - E474
  • [9] Inappropriate implantable cardioverter-defibrillator shocks in MADIT II
    Daubert, James P.
    Zareba, Wojciech
    Cannom, David S.
    McNitt, Scott
    Rosero, Spencer Z.
    Wang, Paul
    Schuger, Claudio
    Steinberg, Jonathan S.
    Higgins, Steven L.
    Wilber, David J.
    Klein, Helmut
    Andrews, Mark L.
    Hall, W. Jackson
    Moss, Arthur J.
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2008, 51 (14) : 1357 - 1365
  • [10] Methods of minimizing inappropriate implantable cardioverter-defibrillator shocks
    Schaer B.
    Osswald S.
    Current Cardiology Reports, 2000, 2 (4) : 346 - 352