A Quadruple Hole Resonant Cavity based Optical Pressure Sensor for Blood Pressure Measurement

被引:0
|
作者
Shivam Upadhyay
机构
[1] Government Mahila Engineering College,Department of Electronics and Communication
来源
关键词
Sensor; Photonic Crystal (PCs); Finite Difference Time Domain (FDTD); Plane Wave Expansion (PWE); Blood Pressure (BP);
D O I
暂无
中图分类号
学科分类号
摘要
This day and age, blood pressure (BP) is one of the serious cardiovascular diseases. For the continuous monitoring of BP, a small, light weight and cuff less device is needed. A new photonic crystal based optical pressure sensor is designed for observing of blood pressure. Concept of developing photonic pressure sensor is based on surface sensing principle, when an exterior tension is applied on the outer face of sensor, it is spread evenly in the form of strain and vary the geometrical parameters of sensor hence the resonance wavelength is shifted and is sensed by the optical sensor. The sensor is designed through a 2D-ordered triangular lattice structure. To obtain very high-quality factor and good light confinement, quadruple hole resonant cavities with four scattering rods are introduced into the linear waveguide. The designed sensor can monitor pressure from 0 kPa or 0 mmHg to 50 kPa or 375 mmHg and its operating ranges lies between the optical C and S band. It has very high quality factor 1721, high sensitivity 25 µm/KPa and low minimum detectable pressure 3.98e-7 nN. Over all, the proposed approach provides a new sensing system for the human blood pressure monitoring which have high selectivity, low cost, light weight and easy to use. The band gap of the PCs structure has been computed by plane wave expansion (PWE) method. The transmitted power and resonant wavelength are calculated with the help of finite difference time domain (FDTD) algorithm.
引用
收藏
页码:1647 / 1658
页数:11
相关论文
共 50 条
  • [31] Subcutaneous blood pressure monitoring with an implantable optical sensor
    Theodor, Michael
    Ruh, Dominic
    Fiala, Jens
    Foerster, Katharina
    Heilmann, Claudia
    Manoli, Yiannos
    Beyersdorf, Friedhelm
    Zappe, Hans
    Seifert, Andreas
    BIOMEDICAL MICRODEVICES, 2013, 15 (05) : 811 - 820
  • [32] Subcutaneous blood pressure monitoring with an implantable optical sensor
    Michael Theodor
    Dominic Ruh
    Jens Fiala
    Katharina Förster
    Claudia Heilmann
    Yiannos Manoli
    Friedhelm Beyersdorf
    Hans Zappe
    Andreas Seifert
    Biomedical Microdevices, 2013, 15 : 811 - 820
  • [33] Impact of Clinic-Based Blood Pressure Approaches on Blood Pressure Measurement
    Juraschek, Stephen P.
    Ishak, Anthony
    Mukamal, Kenneth J.
    Cohen, Marc L.
    Beach, Jennifer L.
    AMERICAN JOURNAL OF HYPERTENSION, 2020, 33 (01) : 26 - 30
  • [34] Development and investigation of high resolution resonant pressure sensor with optical interrogation
    Józwik, M
    Gorecki, C
    Sabac, A
    Heinis, D
    Dean, T
    Jacobelli, A
    OPTICAL MEASUREMENT SYSTEMS FOR INDUSTRIAL INSPECTION IV, PTS 1 AND 2, 2005, 5856 : 734 - 739
  • [35] A Resonant Pressure Microsensor with a Wide Pressure Measurement Range
    Xiang, Chao
    Lu, Yulan
    Cheng, Chao
    Wang, Junbo
    Chen, Deyong
    Chen, Jian
    MICROMACHINES, 2021, 12 (04)
  • [36] Soft measurement of cavity pressure based on ultrasonic signal
    Ying, J. (yingji_zju@yahoo.cn), 1600, Huazhong University of Science and Technology (41):
  • [37] Functionalized single wall carbon nanotube sensor in a perturbed microwave resonant cavity based toxin/pollutant gas pressure sensor
    Tooski, S. B.
    JOURNAL OF APPLIED PHYSICS, 2010, 107 (03)
  • [38] Continuous optical blood pressure measurement in the perioperative setting
    Ghamri, Y.
    Axis, A.
    Braun, F.
    Pierrel, N.
    Sola, J.
    Schoettker, P.
    SWISS MEDICAL WEEKLY, 2018, 148 : 3S - 3S
  • [39] Optical fiber pressure sensor based on F-P cavity in the oil and gas well
    Zhao, Qingchao
    Liu, Xiaohui
    Ma, Long
    Zhao, Wenan
    Wang, Hongzhong
    2017 INTERNATIONAL SYMPOSIUM ON RESOURCE EXPLORATION AND ENVIRONMENTAL SCIENCE (REES 2017), 2017, 64
  • [40] Continuous blood pressure measurement based on pulse transit time: a comparison to ambulatory blood pressure measurement
    Kuechler, G.
    Gehring, J.
    Wesely, M.
    Thimel, K.
    Haberl, R.
    EUROPEAN HEART JOURNAL, 2016, 37 : 1267 - 1267