A comparative study of three small-field detectors for patient specific stereotactic arc dosimetry

被引:0
|
作者
Luke K. Webb
Emma K. Inness
Paul H. Charles
机构
[1] Princess Alexandra Hospital,Radiation Oncology
[2] Queensland University of Technology,Science & Engineering Faculty
关键词
Scintillator; Stereotactic radiotherapy; Small fields; Patient dose verification;
D O I
暂无
中图分类号
学科分类号
摘要
This paper examines the difference in patient specific dosimetry using three different detectors of varying active volume, density and composition, for quality assurance of stereotactic treatments. A PTW 60017 unshielded electron diode, an Exradin W1 scintillator, and a PTW 31014 PinPoint small volume ionisation chamber were setup in a Lucy 3D QA phantom, and were positioned at the isocentre of an Elekta Axesse, with beam modulator collimator, using Exactrac and a HexaPODTM couch. Dose measurements were acquired for 43 stereotactic arcs, and compared to BrainLAB iPlan version 3.0.0 treatment planning system (TPS) calculations using a pencil beam algorithm. It was found that for arcs with field sizes >15\documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$>15$$\end{document} mm, the properties of a detector have minimal impact on the measured doses, with all three detectors agreeing with the TPS (to within 5%). However, for field sizes <15\documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$<15$$\end{document} mm, only the scintillator was found to yield results to within 5% of the TPS. The dose discrepancies were found to increase with decreasing field size. It is recommended that for field sizes <15\documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$<15$$\end{document} mm, a water equivalent dosimeter like the Exradin W1 scintillator be used in order to minimise detector composition perturbations in the measured doses.
引用
收藏
页码:217 / 223
页数:6
相关论文
共 50 条
  • [1] A comparative study of three small-field detectors for patient specific stereotactic arc dosimetry
    Webb, Luke K.
    Inness, Emma K.
    Charles, Paul H.
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2018, 41 (01) : 217 - 223
  • [2] Characterization of small-field stereotactic radiosurgery beams with modern detectors
    Tyler, Madelaine
    Liu, Paul Z. Y.
    Chan, Kin Wa
    Ralston, Anna
    McKenzie, David R.
    Downes, Simon
    Suchowerska, Natalka
    PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (21): : 7595 - 7608
  • [3] Measurement of Dosimetric Parameters of Stereotactic Cones with Small-Field Detectors
    Tamang, H. B. Moktan
    Setianegara, J.
    Soultan, D.
    Li, H.
    Jiang, H.
    MEDICAL PHYSICS, 2024, 51 (10) : 7973 - 7973
  • [4] Small-field dosimetry with multiple detectors and Monte Carlo calculations
    Parker, B
    Shin, A
    Liu, H
    MEDICAL PHYSICS, 2002, 29 (06) : 1372 - 1372
  • [5] On Monolithic Silicon Array Detectors for Small-Field Photon Beam Dosimetry
    Biasi, Giordano
    Davis, Jeremy
    Petasecca, Marco
    Guatelli, Susanna
    Perevertaylo, Vladimir
    Kron, Tomas
    Rosenfeld, Anatoly B.
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2018, 65 (09) : 2640 - 2649
  • [6] Effect of angular dependence for small-field dosimetry using seven different detectors
    Kawata, Kohei
    Ono, Tomohiro
    Hirashima, Hideaki
    Tsuruta, Yusuke
    Fujimoto, Takahiro
    Nakamura, Mitsuhiro
    Nakata, Manabu
    MEDICAL PHYSICS, 2023, 50 (03) : 1274 - 1289
  • [7] Assessment of new small-field detectors against standard-field detectors for practical stereotactic beam data acquisition
    McKerracher, C
    Thwaites, DI
    PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (09): : 2143 - 2160
  • [8] Influence of Jaw Setting in the Determination of Stereotactic Small-Field Output Factors with Different Detectors
    George, Seby
    Ponmalar, Y. Retna
    Godson, Henry Finlay
    Kumar, A. Sathish
    Ravindran, B. Paul
    JOURNAL OF MEDICAL PHYSICS, 2022, 47 (01) : 65 - 72
  • [9] Small-Field Dosimetry in Single-Isocenter Stereotactic Irradiation for Brain Metastases Using a Detector-Specific Field Output Correction Factor
    Ono, T.
    Nakamura, M.
    Uto, M.
    Mizowaki, T.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [10] On the Monte Carlo simulation of small-field micro-diamond detectors for megavoltage photon dosimetry
    Andreo, Pedro
    Palmans, Hugo
    Marteinsdottir, Maria
    Benmakhlouf, Hamza
    Carlsson-Tedgren, Asa
    PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (01): : L1 - L10