Validation of TONOPORT V blood-pressure measuring monitor in adults

被引:0
|
作者
A Haensel
K Utech
W Langewitz
机构
[1] Outpatient Clinic,Department of Internal Medicine
[2] University Hospital Basel,Department of Psychosomatic Medicine
[3] University Hospital Basel,undefined
来源
关键词
blood-pressure measuring device; Tonoport V; ambulatory blood pressure measuring;
D O I
暂无
中图分类号
学科分类号
摘要
A new automatic blood-pressure (BP) measuring device TONOPORT V was evaluated according to the International Protocol for Validation of Blood Pressure Measuring Devices in adults by the European Society of Hypertension. BP values measured by the TONOPORT V were compared to BP readings from two independent observers. A total of 33 patients (20 males, 13 females) provided systolic and diastolic BP readings in the normotensive, borderline hypertensive, and hypertensive range. Their age varied between 30 and 83 years, and their arm circumference between 23 and 36 cm. The device showed a mean (±s.d.) deviation from observer measurements of −0.7 (4.6) mmHg for systolic and −0.8 (4.4) mmHg for diastolic BP. The accuracy of the device did not vary according to BP values or other patient characteristics. The device passed all phases of the protocol and can be recommended following the regulation rules of the European Society of Hypertension.
引用
收藏
页码:745 / 749
页数:4
相关论文
共 50 条
  • [31] BLOOD-PRESSURE SCREENING IN ADULTS
    不详
    AMERICAN FAMILY PHYSICIAN, 1994, 50 (08) : 1729 - 1732
  • [32] AMBULATORY BLOOD-PRESSURE MONITORING DURING PREGNANCY - VALIDATION OF THE TM-2420 MONITOR
    CLARK, S
    HOFMEYR, GJ
    COATS, AJS
    REDMAN, CWG
    OBSTETRICS AND GYNECOLOGY, 1991, 77 (01): : 152 - 155
  • [33] EFFECT OF MEASURING AMBULATORY BLOOD-PRESSURE ON SLEEP AND ON BLOOD-PRESSURE DURING SLEEP
    DAVIES, RJO
    JENKINS, NE
    STRADLING, JR
    BRITISH MEDICAL JOURNAL, 1994, 308 (6932): : 820 - 823
  • [34] FINAPRES - A NONINVASIVE DEVICE TO MONITOR BLOOD-PRESSURE
    PORTER, KB
    OBRIEN, WF
    KIEFERT, V
    KNUPPEL, RA
    OBSTETRICS AND GYNECOLOGY, 1991, 78 (03): : 430 - 433
  • [35] EVALUATION OF A CONTINUOUS NONINVASIVE BLOOD-PRESSURE MONITOR
    ABENSTEIN, JP
    STENSRUD, PE
    SPACKMAN, TN
    TARHAN, S
    ANESTHESIOLOGY, 1993, 79 (3A) : A504 - A504
  • [36] EVALUATION OF AN INEXPENSIVE ARTERIAL BLOOD-PRESSURE MONITOR
    BISGARD, GE
    GILLILAN, SW
    VETERINARY MEDICINE & SMALL ANIMAL CLINICIAN, 1973, 68 (09): : 1005 - 1007
  • [37] MEASURING ANKLE SYSTOLIC BLOOD-PRESSURE - VALIDATION OF THE DINAMAP 1846SX
    MUNDT, KA
    CHAMBLESS, LE
    BURNHAM, CB
    HEISS, G
    ANGIOLOGY, 1992, 43 (07) : 555 - 566
  • [38] EVALUATION OF THE DINAMAP CONTINUOUS BLOOD-PRESSURE MONITOR
    MANOLIO, TA
    FISHEL, SC
    BEATTIE, C
    TORRES, J
    CHRISTOPHERSON, R
    MERRITT, WT
    WHELTON, PK
    AMERICAN JOURNAL OF HYPERTENSION, 1988, 1 (03) : S161 - S167
  • [39] ASSESSMENT OF BLOOD-PRESSURE MEASURING TECHNIQUES
    MCKAY, DW
    RAJU, MK
    CAMPBELL, NRC
    MEDICAL EDUCATION, 1992, 26 (03) : 208 - 212
  • [40] VALUE OF MEASURING BLOOD-PRESSURE IN PREGNANCY
    DANIEL, DG
    BRITISH MEDICAL JOURNAL, 1993, 307 (6913): : 1214 - 1215