Applications of CFD Technique in the Design and Flow Analysis of Implantable Axial Flow Blood Pump

被引:0
|
作者
Hui-min Fan
Fang-wen Hong
Guo-ping Zhang
Liang Ye
Zhong-Min Liu
机构
[1] Shanghai East Hospital Affiliated to Tongji University,
[2] China Ship Scientific Research Center,undefined
来源
Journal of Hydrodynamics | 2010年 / 22卷
关键词
axial flow blood pump; CFD; PIV;
D O I
暂无
中图分类号
学科分类号
摘要
Based on established numerical methods and hydrodynamic performance testing facilities, CFD technique are applied to improve the design of the implantable axial flow blood pump and the flow analysis. (1) Applying brushless machine magnet steel, reducing its thickness while increasing the length, the flow channel’s cross-section is increased, with no space connection between the large and the small rotators, and with a cone transition segment from the bearing to the principal axis, the flow is made smoother. The rotating speed is lowered by 1000 rpm - 1200 rpm under the same flow rate and pressure head, and thus the hemolysis can be avoided. (2) Different outlet stator guiding vanes are selected for the same blood pump for analyses of hydrodynamic performances and flow fields. An excellent design not only can regulate the rotating flow field into an axial one, reduce the circulation loss and improve the pump efficiency, but also can avoid backflow, vortex and secondary flow at the pump outlet, and thus the thrombus can be prevented. (3) The calf live tests show that some residual clots exist at the inner wall of the outlet connection bends, which are analyzed and explained by CFD techniques and the corresponding improvements are proposed. All results are verified by hydrodynamic performance tests and PIV flow field tests, and consistent conclusions are obtained.
引用
收藏
页码:518 / 525
页数:7
相关论文
共 50 条
  • [21] Noninvasive Miniaturized Mass-flow Meter using a Curved Cannula for Implantable Axial Flow Blood Pump
    Kosaka, Ryo
    Nishida, Masahiro
    Maruyama, Osamu
    Yamane, Takashi
    2011 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2011, : 1343 - 1346
  • [22] The Optimal Blading Design and CFD Analysis of Axial Flow Fan
    Lee, Chan
    Noh, Eui Jong
    Kim, Seung Wook
    Yang, Sang Ho
    International Journal of Fluid Machinery and Systems, 2022, 15 (03): : 367 - 375
  • [23] CFD ANALYSIS OF A CANNED PUMP ROTOR CONSIDERING AN ANNULAR FLUID WITH AXIAL FLOW
    Untaroiu, Alexandrina
    Dimond, Timothy W.
    Allaire, Paul E.
    Armentrout, Richard
    PROCEEDINGS OF THE ASME TURBO EXPO 2009, VOL 6, PTS A AND B, 2009, : 1013 - 1022
  • [24] Proposal of hemodynamically improved design of an axial flow blood pump for LVAD
    Vikas Kannojiya
    Arup Kumar Das
    Prasanta Kumar Das
    Medical & Biological Engineering & Computing, 2020, 58 : 401 - 418
  • [25] Proposal of hemodynamically improved design of an axial flow blood pump for LVAD
    Kannojiya, Vikas
    Das, Arup Kumar
    Das, Prasanta Kumar
    MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2020, 58 (02) : 401 - 418
  • [27] Off-pump Replacement of the INCOR Implantable Axial-flow Pump
    Nakashima, Kuniki
    Kirsch, Matthias E. W.
    Vermes, Emmanuelle
    Rosanval, Odile
    Loisance, Daniel
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2009, 28 (02): : 199 - 201
  • [28] Modeling and identification of an axial flow blood pump
    Choi, SJ
    Boston, JR
    Thomas, D
    Antaki, JF
    PROCEEDINGS OF THE 1997 AMERICAN CONTROL CONFERENCE, VOLS 1-6, 1997, : 3714 - 3715
  • [29] PERFORMANCE CHARACTERIZATION AND CFD ANALYSIS OF AN AXIAL BLOOD PUMP
    Chan, W. K.
    Wong, Y. W.
    Koh, S. Y.
    Chong, V.
    JOURNAL OF MECHANICS IN MEDICINE AND BIOLOGY, 2005, 5 (01) : 151 - 163
  • [30] Multi-condition design optimization of groove flow control technique in an axial-flow pump
    Li, Jinghong
    Zhang, Rui
    Xu, Hui
    Feng, Jiangang
    INTERNATIONAL JOURNAL OF TURBO & JET-ENGINES, 2024, 40 (s1) : s17 - s32