The characteristic absolute efficiency response curves of a high purity germanium detector (HPGe) for different counting geometries have been established in the energy range 50–1500 keV by measuring the absolute efficiencies using both mono-energetic and multi-gamma emitting radionuclide point calibrated sources supplied by IAEA. Several fitting functions proposed in the literature were assessed for interpolation within the intermediate energy range of interest. The values of the function parameters have been determined by using the linear least square methods. The problems associated with the measurements of experimental efficiency data at small source–detector distances and the importance of the correlation matrix in the estimation of precise uncertainties have been shown. It was found that the inclusion of correlation matrices in the propagation of error formulae plays a significant role up to 450 keV gamma-ray energy and results in a drastic reduction of errors associated with the predicted efficiencies. The discrepancy at closer counting geometries in the absence of true gamma-gamma coincidence corrections is found to reach to about 30%.
引用
收藏
页码:457 / 462
页数:5
相关论文
共 50 条
  • [31] STUDY OF MATHEMATICAL FUNCTIONS TO FIT DETECTOR EFFICIENCY CURVE IN GAMMA RAY ENERGY RANGE FROM 59.54 keV UP TO 1408.01 keV
    Thabet, Abouzeid A.
    Salem, Bohaysa A.
    Badawi, Mohamed S.
    NUCLEAR TECHNOLOGY & RADIATION PROTECTION, 2021, 36 (03): : 224 - 233
  • [32] Simulation of the response of a segmented high-purity germanium detector for gamma emission tomography of nuclear fuel
    Andersson, Peter
    Rathore, Vikram
    Senis, Lorenzo
    Anastasiadis, Anastasios
    Sunden, Erik Andersson
    Atak, Haluk
    Holcombe, Scott
    Hakansson, Ane
    Jansson, Peter
    Nyberg, Johan
    SN APPLIED SCIENCES, 2020, 2 (02):
  • [33] Simulation of the response of a segmented high-purity germanium detector for gamma emission tomography of nuclear fuel
    Peter Andersson
    Vikram Rathore
    Lorenzo Senis
    Anastasios Anastasiadis
    Erik Andersson Sundén
    Haluk Atak
    Scott Holcombe
    Ane Håkansson
    Peter Jansson
    Johan Nyberg
    SN Applied Sciences, 2020, 2
  • [34] Optimisation design of cylindrical containers for improving the detection efficiency of a high-purity germanium detector using the LabSOCS
    Yang, Baolu
    Ha, Yiming
    Li, An
    Zhou, Hongjie
    Wang, Feng
    Li, Weiming
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2013, 298 (03) : 1673 - 1677
  • [35] Assessment of the Soil Self-Attenuation Correction Factor to Determine the Efficiency of a High-Purity Germanium Detector
    Reza Pourimani
    Monire Mohebian
    S. M. Modarresi
    Iranian Journal of Science and Technology, Transactions A: Science, 2020, 44 : 311 - 317
  • [36] Assessment of the Soil Self-Attenuation Correction Factor to Determine the Efficiency of a High-Purity Germanium Detector
    Pourimani, Reza
    Mohebian, Monire
    Modarresi, S. M.
    IRANIAN JOURNAL OF SCIENCE AND TECHNOLOGY TRANSACTION A-SCIENCE, 2020, 44 (A1): : 311 - 317
  • [37] Optimisation design of cylindrical containers for improving the detection efficiency of a high-purity germanium detector using the LabSOCS
    Baolu Yang
    Yiming Ha
    An Li
    Hongjie Zhou
    Feng Wang
    Weiming Li
    Journal of Radioanalytical and Nuclear Chemistry, 2013, 298 : 1673 - 1677
  • [38] Tables of elastic scattering cross sections of photons in the energy range 50-1500 keV for all elements in the range 13≤Z≤104
    Chatterjee, BK
    Roy, SC
    JOURNAL OF PHYSICAL AND CHEMICAL REFERENCE DATA, 1998, 27 (06) : 1011 - 1215
  • [39] X-RAY LINEAR ABSORPTION-COEFFICIENTS FOR SILICON AND GERMANIUM IN THE ENERGY-RANGE 5 TO 50 KEV
    HILDEBRANDT, G
    ACTA CRYSTALLOGRAPHICA SECTION A, 1979, 35 (JUL): : 696 - 697
  • [40] ACCURATE MEASUREMENT OF RELATIVE EFFICIENCY OF GE(LI) GAMMA-RAY DETECTORS IN ENERGY RANGE 500 TO 1500 KEV
    FREEMAN, JM
    JENKIN, JG
    NUCLEAR INSTRUMENTS & METHODS, 1966, 43 (02): : 269 - &