Experimental investigations and mathematical simulation of the operation of ionizing-radiation diamond detectors

被引:0
|
作者
E. V. Gladchenkov
K. V. Zakharchenko
R. F. Ibragimov
A. F. Kaperko
V. A. Kolyubin
V. P. Kulagin
P. G. Nedosekin
E. M. Tyurin
机构
[1] National Research University,
[2] Higher School of Economics,undefined
[3] National Research Nuclear University,undefined
[4] Moscow Engineering Physics Institute (MEPhI),undefined
关键词
D O I
暂无
中图分类号
学科分类号
摘要
A mathematical model of an ionizing-radiation monitor based on diamond detectors has been developed. Experimental investigations of the monitor were carried out, and, using their results, the model was verified and optimized. It was demonstrated that the developed model makes it possible to calculate data of sensor measurements with an accuracy of 10% or better. These results will be used to reconstruct cosmicradiation spectra according to the output data of the monitor.
引用
收藏
页码:339 / 344
页数:5
相关论文
共 50 条
  • [31] STABILITY OF SPECTROMETRIC CHARACTERISTICS OF DETECTORS OF LOW-PENETRATION IONIZING-RADIATION
    VYDAI, YT
    ZAGARII, LB
    TSIRLIN, YA
    INSTRUMENTS AND EXPERIMENTAL TECHNIQUES, 1990, 33 (02) : 311 - 313
  • [32] An Investigation of Nanocrystalline Semiconductor Assemblies as a Material Basis for Ionizing-Radiation Detectors
    Kim, Geehyun
    Huang, James
    Hammig, Mark D.
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2009, 56 (03) : 841 - 848
  • [33] Doping silicon with lanthanides as a method for the optimization of parameters of ionizing-radiation detectors
    Afanas'eva, N.P.
    Brinkevich, D.I.
    Prosolovich, V.S.
    Yankovskij, Yu.N.
    Pribory i Tekhnika Eksperimenta, 2002, 45 (03): : 24 - 26
  • [34] Doping Silicon with Lanthanides as a Method for the Optimization of Parameters of Ionizing-Radiation Detectors
    N. P. Afanas'eva
    D. I. Brinkevich
    V. S. Prosolovich
    Yu. N. Yankovskii
    Instruments and Experimental Techniques, 2002, 45 : 314 - 316
  • [35] BULK RESPONSE OF SOME POLYMERIC NUCLEAR TRACK DETECTORS TO IONIZING-RADIATION
    LUCK, HB
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH, 1983, 212 (1-3): : 371 - 373
  • [36] YALO3-CE FAST SCINTILLATORS FOR DETECTORS OF IONIZING-RADIATION
    BARYSHEVSKII, VG
    KORZHIK, MV
    MOROZ, VI
    PAVLENKO, VB
    FEDOROV, AA
    SMIRNOV, SA
    EGORYCHEVA, OA
    KACHANOV, VA
    INSTRUMENTS AND EXPERIMENTAL TECHNIQUES, 1992, 35 (03) : 440 - 442
  • [37] NONLINEAR SIGNAL TRANSFORMATION IN THICKNESS GAUGING WITH MULTIPLE IONIZING-RADIATION DETECTORS
    NEDAVNII, OI
    SOVIET JOURNAL OF NONDESTRUCTIVE TESTING-USSR, 1987, 23 (10): : 683 - 685
  • [38] FUTURE POTENTIAL OF CHARGE-COUPLED-DEVICES AS DETECTORS OF IONIZING-RADIATION
    ACTON, P
    AGNEW, G
    COTTON, R
    HEDGES, S
    MCKEMEY, AK
    ROBBINS, M
    ROY, T
    WATTS, SJ
    DAMERELL, CJS
    ENGLISH, RL
    GILLMAN, AR
    LINTERN, AL
    SU, D
    WICKENS, FJ
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1991, 305 (03): : 504 - 511
  • [39] Doping silicon with lanthanides as a method for the optimization of parameters of ionizing-radiation detectors
    Afanas'eva, NP
    Brinkevich, DI
    Prosolovich, VS
    Yankovskii, YN
    INSTRUMENTS AND EXPERIMENTAL TECHNIQUES, 2002, 45 (03) : 314 - 316
  • [40] RESPONSE OF EXPERIMENTAL EPILEPTIC FOCUS TO FOCAL IONIZING-RADIATION
    BARCIASALORIO, JL
    VANACLOCHA, V
    CERDA, M
    CIUDAD, J
    LOPEZGOMEZ, L
    APPLIED NEUROPHYSIOLOGY, 1987, 50 (1-6) : 359 - 364