Potential effect of solvent and slit width on some properties of room temperature fluorescence of hydroxy polycyclic aromatic hydrocarbons

被引:0
|
作者
Bassam F. Alfarhani
Rana A. Hammza
Ali S. Alzaidy
机构
[1] University of Al-Qadisiyah,Department of Chemistry, College of Sciences
[2] Al-Nahrain University,Department of Chemistry, College of Science
来源
Chemical Papers | 2021年 / 75卷
关键词
Slit width; Solvents; PAH; Metabolite; Fluorescence;
D O I
暂无
中图分类号
学科分类号
摘要
Spectral properties of the hydroxy polycyclic aromatic hydrocarbons (2-naphthol, 1-hydroxypyrene and 1-hydroxyfluorene) were studied in solvents (methanol, ethanol, 1-propanol and 1-hexanol) room temperature emission and excitation spectroscopy. Different values of slit width have been used (1.5, 3, 5, 10 and 15 nm). The results show that the FWHM and the intensity of the fluorescence signal affect using different silt width; generally, higher slit width will exhibit larger intensity value with lower details of fluorescence signal. With using different solvent, the results show wide range of differences between different solvents and these effects due to the different solvation process.