Metal artifact reduction in patients with dental implants using multispectral three-dimensional data acquisition for hybrid PET/MRI

被引:28
|
作者
Gunzinger J.M. [1 ]
Delso G. [2 ]
Boss A. [3 ]
Porto M. [1 ]
Davison H. [1 ]
von Schulthess G.K. [1 ]
Huellner M. [1 ,4 ]
Stolzmann P. [1 ,4 ]
Veit-Haibach P. [1 ,3 ]
Burger I.A. [1 ,3 ]
机构
[1] Department of Medical Radiology, Division of Nuclear Medicine, University Hospital Zurich, Ramistr. 100, Zurich
[2] Global MR Applications and Workflow, GE Healthcare, Zurich
[3] Department of Medical Radiology, Institute of Diagnostic and Interventional Radiology, University Hospital Zurich, Ramistr. 100, Zurich
[4] Department of Medical Radiology, Institute of Neuroradiology, University Hospital Zurich, Ramistr. 100, Zurich
关键词
Attenuation correction; Image noise; MAVRIC; Signal voids;
D O I
10.1186/s40658-014-0102-z
中图分类号
学科分类号
摘要
Background: Hybrid positron emission tomography/magnetic resonance imaging (PET/MRI) shows high potential for patients with oropharyngeal cancer. Dental implants can cause substantial artifacts in the oral cavity impairing diagnostic accuracy. Therefore, we evaluated new MRI sequences with multi-acquisition variable-resonance image combination (MAVRIC SL) in comparison to conventional high-bandwidth techniques and in a second step showed the effect of artifact size on MRI-based attenuation correction (AC) with a simulation study. Methods: Twenty-five patients with dental implants prospectively underwent a trimodality PET/CT/MRI examination after informed consent was obtained under the approval of the local ethics committee. A conventional 3D gradient-echo sequence (LAVA-Flex) commonly used for MRI-based AC of PET (acquisition time of 14 s), a T1w fast spin-echo sequence with high bandwidth (acquisition time of 3.2 min), as well as MAVRIC SL sequence without and with increased phase acceleration (MAVRIC, acquisition time of 6 min; MAVRIC-fast, acquisition time of 3.5 min) were applied. The absolute and relative reduction of the signal void artifact was calculated for each implant and tested for statistical significance using the Wilcoxon signed-rank test. The effect of artifact size on PET AC was simulated in one case with a large tumor in the oral cavity. The relative difference of the maximum standardized uptake value (SUVmax) in the tumor was calculated for increasing artifact sizes centered over the second molar. Results: The absolute reduction of signal void from LAVA-Flex sequences to the T1-weighted fast spin-echo (FSE) sequences was 416 mm2 (range 4 to 2,010 mm2) to MAVRIC 481 mm2 (range 12 to 2,288 mm2) and to MAVRIC-fast 486 mm2 (range 39 to 2,209 mm2). The relative reduction in signal void was significantly improved for both MAVRIC and MAVRIC-fast compared to T1 FSE (−75%/− 78% vs. − 62%, p < 0.001 for both). The relative error for SUVmax was negligible for artifacts of 0.5-cm diameter (−0.1%), but substantial for artifacts of 5.2-cm diameter (−33%). Conclusions: MAVRIC-fast could become useful for artifact reduction in PET/MR for patients with dental implants. This might improve diagnostic accuracy especially for patients with tumors in the oropharynx and substantially improve accuracy of PET quantification. © 2014, Gunzinger et al; licensee Springer.
引用
收藏
页码:1 / 14
页数:13
相关论文
共 50 条
  • [1] A Multispectral Three-Dimensional Acquisition Technique for Imaging Near Metal Implants
    Koch, Kevin M.
    Lorbiecki, John. E.
    Hinks, R. Scott
    King, Kevin F.
    MAGNETIC RESONANCE IN MEDICINE, 2009, 61 (02) : 381 - 390
  • [2] Three-dimensional shape completion using deep convolutional neural networks: Application to truncation compensation and metal artifact reduction in PET/MRI attenuation correction
    Arabi, Hossein
    Zaidi, Habib
    2019 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2019,
  • [3] Reduction of metal artifact in three-dimensional computed tomography (3D CT) with dental impression materials
    Park, W. S.
    Kim, K. D.
    Shin, H. K.
    Lee, S. H.
    2007 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-16, 2007, : 3496 - +
  • [4] New perspective for dental implants using three-dimensional computed graphics
    Cavalcanti, MGP
    Ruprecht, A
    Vannier, MW
    JOURNAL OF DENTAL RESEARCH, 2002, 81 : B135 - B135
  • [5] Polymer gel dosimetry using a three-dimensional MRI acquisition technique
    Baras, P
    Seimenis, I
    Kipouros, P
    Papagiannis, P
    Angelopoulos, A
    Sakelliou, L
    Pappas, E
    Baltas, D
    Karaiskos, P
    Sandilos, P
    Vlachos, L
    MEDICAL PHYSICS, 2002, 29 (11) : 2506 - 2516
  • [6] Functional optoacoustic imaging of moving objects using microsecond-delay acquisition of multispectral three-dimensional tomographic data
    Xosé Luís Deán-Ben
    Erwin Bay
    Daniel Razansky
    Scientific Reports, 4
  • [7] Functional optoacoustic imaging of moving objects using microsecond-delay acquisition of multispectral three-dimensional tomographic data
    Dean-Ben, Xose Luis
    Bay, Erwin
    Razansky, Daniel
    SCIENTIFIC REPORTS, 2014, 4
  • [8] Multimodality image fusion of three-dimensional PET, MRI imaging, and CT data of the thorax in tumor patients
    Beier, J
    Bittner, R
    Rohlfing, T
    Mueller, F
    Wust, P
    Felix, R
    RADIOLOGY, 1999, 213P : 261 - 261
  • [9] THREE-DIMENSIONAL HYBRID-ENCODED MRI USING COMPRESSED SENSING
    Wang, Haifeng
    Liang, Dong
    King, Kevin F.
    Ying, Leslie
    2012 9TH IEEE INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING (ISBI), 2012, : 398 - 401
  • [10] Metal artifact reduction using virtual monochromatic images for patients with pedicle screws implants on CT
    Dong, Yue
    Shi, Ai Jun
    Wu, Jian Lin
    Wang, Ru Xin
    Sun, Li Fei
    Liu, Ai Lian
    Liu, Yi Jun
    EUROPEAN SPINE JOURNAL, 2016, 25 (06) : 1754 - 1763