Plastic microchip flow cytometer based on 2- and 3-dimensional hydrodynamic flow focusing

被引:0
|
作者
S. Chung
S. J. Park
J. K. Kim
C. Chung
D. C. Han
J. K. Chang
机构
[1] Seoul National University,School of Mechanical and Aerospace Engineering
[2] Seoul National University,Digital Bio Technology, Co., Rm. 1304, Bldg 136
[3] Seoul National University,School of Electrical Engineering and Computer Science
来源
关键词
Fluid Dynamic; Computational Fluid Dynamic; Calculation Result; Laser Scanning; Flow Cytometer;
D O I
暂无
中图分类号
学科分类号
摘要
This paper deals with design and fabrication of 2- and 3- dimensional microchip type flow cytometers hydrodynamically driven. Flow rate and pressure distrubution have been calculated from Poiseuille flow theory and results were confirmed by computational fluid dynamics. The flow focusing has been evaluated in 2- and 3-dimensional microchip type flow cytometers fabricated with Polydimetylsiloxane molding process. The shape of focused flow was detected with a laser scanning microscope and favourably compared with calculations based on Poiseuille flow assumptions and CFD (Computational Fluid Dynamics) simulation. Simulation and calculation results showed good correlations to each other, and also to the visualized shape of focused flow in microchip. The 3-dimensional focused flow showed the more stable focusing status in laser induced fluorescence measurements than that in 2-dimensional case.
引用
收藏
页码:525 / 533
页数:8
相关论文
共 50 条
  • [31] HYDRODYNAMIC CHROMATOGRAPHY - 3-DIMENSIONAL LAMINAR DISPERSION IN RECTANGULAR CONDUITS WITH TRANSVERSE FLOW
    TAKAHASHI, T
    GILL, WN
    CHEMICAL ENGINEERING COMMUNICATIONS, 1980, 5 (5-6) : 367 - 385
  • [32] STABILITY OF LAMINAR FLOW BETWEEN PARALLEL PLANES FOR 2- + 3-DIMENSIONAL FINITE DISTURBANCES
    MEKSYN, D
    ZEITSCHRIFT FUR PHYSIK, 1964, 178 (02): : 159 - +
  • [33] 3-DIMENSIONAL FLOW PATTERNS IN 2-DIMENSIONAL WAKES
    TRIANTAFYLLOU, GS
    JOURNAL OF FLUIDS ENGINEERING-TRANSACTIONS OF THE ASME, 1992, 114 (03): : 356 - 361
  • [34] Measurements of scattered light on a microchip flow cytometer with integrated polymer based optical elements
    Wang, Z
    El-Ali, J
    Engelund, M
    Gotsæd, T
    Perch-Nielsen, IR
    Mogensen, KB
    Snakenborg, D
    Kutter, JP
    Wolff, A
    LAB ON A CHIP, 2004, 4 (04) : 372 - 377
  • [35] Development of Microfluidic Impedance Flow Cytometry Based on Three-Dimensional Hydrodynamic Focusing
    Chen, Xiao
    Li, Yimin
    Wang, Mingruihong
    Qin, Xuzhen
    Wang, Junbo
    Huo, Xiaoye
    Chen, Jian
    2024 IEEE SENSORS, 2024,
  • [36] ANALYSIS OF FLOW INTO DRAINTILE IN 3-DIMENSIONAL FLOW FIELD
    HAZENBERG, G
    PANU, US
    JOURNAL OF HYDROLOGY, 1991, 122 (1-4) : 321 - 333
  • [37] 3-DIMENSIONAL FLOW IN A ROTOR - A STRICTLY 3-DIMENSIONAL CALCULATION OF FLOW IN TURBOMACHINES BY THE S1-S2 METHOD
    BUISINE, D
    MICHEAU, P
    JOURNAL DE MECANIQUE THEORIQUE ET APPLIQUEE, 1987, 6 (04): : 525 - 546
  • [38] SOLUTION OF 3-DIMENSIONAL PLASTIC-FLOW PROBLEM IN THIN SHEET ROLLING
    POLUKHIN, VP
    CHERNILE.VE
    RUSSIAN METALLURGY-METALLY-USSR, 1970, (06): : 72 - &
  • [39] Optical guiding-based cell focusing for Raman flow cell cytometer
    Verma, Ravi Shanker
    Ahlawat, Sunita
    Uppal, Abha
    ANALYST, 2018, 143 (11) : 2648 - 2655
  • [40] Three-dimensional hydrodynamic flow and particle focusing using four vortices Dean flow
    Byung Hang Ha
    Kang Soo Lee
    Jin Ho Jung
    Hyung Jin Sung
    Microfluidics and Nanofluidics, 2014, 17 : 647 - 655