Effect of the saturation pulse duration on chemical exchange saturation transfer in amide proton transfer MR imaging: a phantom study

被引:12
|
作者
Wada T. [1 ]
Togao O. [2 ]
Tokunaga C. [1 ]
Funatsu R. [1 ]
Kobayashi K. [1 ]
Nakamura Y. [1 ]
机构
[1] Division of Radiology, Department of Medical Technology, Kyushu University Hospital, 3-1-1, Maidashi, Higashi-ku, Fukuoka
[2] Department of Clinical Radiology, Graduate School of Medical Sciences, Kyushu University, 3-1-1, Maidashi, Higashi-ku, Fukuoka
基金
日本学术振兴会;
关键词
Amide proton transfer (APT) imaging; Chemical exchange saturation transfer (CEST); Chicken egg white; Parallel RF transmission; Saturation pulse duration;
D O I
10.1007/s12194-015-0326-1
中图分类号
学科分类号
摘要
Amide proton transfer (APT) contrast imaging is based on the chemical exchange saturation transfer (CEST) of protons between the amide groups and bulk water. Here, we demonstrate the effect of the saturation pulse duration on CEST in APT imaging with use of a clinical MR scanner. Four samples were prepared from chicken egg white diluted with H2O. Experiments were performed on a 3T clinical MR scanner with use of a body coil for two-channel parallel radiofrequency transmission. APT images were acquired at six frequency offsets (± 3.0, ± 3.5, ± 4.0 ppm) with respect to the water resonance as well as one far off-resonant frequency (−160 ppm) for signal normalization. The CEST effect was defined as asymmetry of the magnetization transfer ratio at 3.5 ppm. We measured the CEST effects in the egg white samples with different concentrations at seven saturation pulse durations. The influence of the extension of repetition time (TR) on the CEST effect was also evaluated. The CEST effect was not influenced by the change in TR. The CEST effect was increased significantly with the concentration when the duration was ≥1.0 s (P < 0.01). The CEST effect was highly correlated with the concentration at all saturation pulse durations, and its increase ratio was higher at longer saturation pulse durations. In conclusion, a long saturation pulse duration is useful for the sensitive detection of mobile proteins and peptides in APT imaging. © 2015, Japanese Society of Radiological Technology and Japan Society of Medical Physics.
引用
收藏
页码:15 / 21
页数:6
相关论文
共 50 条
  • [1] Effect of Saturation Pulse Duration and Power on pH-weighted Amide Proton Transfer Imaging: A Phantom Study
    Tokunaga, Chiaki
    Wada, Tatsuhiro
    Togao, Osamu
    Yamashita, Yasuo
    Kobayashi, Kouji
    Kato, Toyoyuki
    MAGNETIC RESONANCE IN MEDICAL SCIENCES, 2023, 22 (04) : 487 - 495
  • [2] A fast and novel method for amide proton transfer-chemical exchange saturation transfer multislice imaging
    Schuere, Jan-Ruediger
    Pilatus, Ulrich
    Deichmann, Ralf
    Hattingen, Elke
    Shrestha, Manoj
    NMR IN BIOMEDICINE, 2021, 34 (07)
  • [3] Adaptive denoising for chemical exchange saturation transfer MR imaging
    Breitling, Johannes
    Deshmane, Anagha
    Goerke, Steffen
    Korzowski, Andreas
    Herz, Kai
    Ladd, Mark E.
    Scheffler, Klaus
    Bachert, Peter
    Zaiss, Moritz
    NMR IN BIOMEDICINE, 2019, 32 (11)
  • [4] Quantifying amide proton exchange rate and concentration in chemical exchange saturation transfer imaging of the human brain
    Heo, Hye-Young
    Han, Zheng
    Jiang, Shanshan
    Schar, Michael
    van Zijl, Peter C. M.
    Zhou, Jinyuan
    NEUROIMAGE, 2019, 189 : 202 - 213
  • [5] B1 Power Optimization for Chemical Exchange Saturation Transfer Imaging: A Phantom Study Using Egg White for Amide Proton Transfer Imaging Applications in the Human Brain
    Kanazawa, Yuki
    Fushimi, Yasutaka
    Sakashita, Naotaka
    Okada, Tomohisa
    Arakawa, Yoshiki
    Miyazaki, Mitsue
    MAGNETIC RESONANCE IN MEDICAL SCIENCES, 2018, 17 (01) : 86 - 94
  • [6] Chemical Exchange Saturation Transfer MR Imaging Potential Clinical Applications
    Polders, Daniel Louis
    Hoogduin, Johannes Marinus
    PET CLINICS, 2013, 8 (03) : 245 - 257
  • [7] Direct saturation-corrected chemical exchange saturation transfer MRI of glioma: Simplified decoupling of amide proton transfer and nuclear overhauser effect contrasts
    Zhou, Iris Yuwen
    Wang, Enfeng
    Cheung, Jerry S.
    Lu, Dongshuang
    Ji, Yang
    Zhang, Xiaoan
    Fulci, Giulia
    Sun, Phillip Zhe
    MAGNETIC RESONANCE IN MEDICINE, 2017, 78 (06) : 2307 - 2314
  • [8] Chemical exchange saturation transfer imaging and spectroscopy
    Zhou, Jinyuan
    van Zijl, Peter C. M.
    PROGRESS IN NUCLEAR MAGNETIC RESONANCE SPECTROSCOPY, 2006, 48 (2-3) : 109 - 136
  • [9] Chemical Exchange Saturation Transfer for Epilepsy Imaging
    Brown, Joshua D.
    Kim, Hahnsung
    Sun, Philip Zhe
    Hu, Ranliang
    APPLIED RADIOLOGY, 2024, 53 (03)
  • [10] Quantitative Chemical Exchange Saturation Transfer Imaging of Amide Proton Transfer Differentiates between Cerebellopontine Angle Schwannoma and Meningioma: Preliminary Results
    Koike, Hirofumi
    Morikawa, Minoru
    Ishimaru, Hideki
    Ideguchi, Reiko
    Uetani, Masataka
    Hiu, Takeshi
    Matsuo, Takayuki
    Miyoshi, Mitsuharu
    INTERNATIONAL JOURNAL OF MOLECULAR SCIENCES, 2022, 23 (17)