Use of the optical access trocar for safe and rapid entry in various laparoscopic procedures

被引:0
|
作者
A. String
E. Berber
A. Foroutani
J.R. Macho
J.M. Pearl
A.E. Siperstein
机构
[1] Department of General Surgery,
[2] The Cleveland Clinic Foundation,undefined
[3] 9500 Euclid Avenue,undefined
[4] Cleveland,undefined
[5] OH 44195,undefined
[6] USA,undefined
[7] Department of Surgery,undefined
[8] University of California San Francisco,undefined
[9] 1600 Divisadero St.,undefined
[10] Mount Zion Medical Center,undefined
[11] San Francisco,undefined
[12] CA,undefined
[13] 94115,undefined
[14] USA,undefined
来源
Surgical Endoscopy | 2001年 / 15卷
关键词
D O I
暂无
中图分类号
学科分类号
摘要
Background: The use of the Veress needle in laparoscopy to create the pneumoperitoneum has inherent risks; it may cause vascular and visceral injuries. The open technique is compromised by the leakage of carbon dioxide and can also be time consuming. One alternative is to enter the abdomen using an optical trocar under direct view. Our aim was to determine whether the optical access trocar can be used to effect a safe and rapid entry in various laparoscopic procedures. Methods: Over a 4-year period, the Optiview trocar was used for initial entry in 650 laparoscopic procedures. The procedures included cholecystectomy (n = 282), transabdominal inguinal hernia repair (n = 76), radiofrequency ablation of liver tumors (n = 73), adrenalectomy (n = 54), appendectomy (n = 41), colorectal surgery (n = 39), and various other procedures (n = 85). The following parameters were analyzed: presence of previous abdominal operations, site and duration of entry, and complications. Results: Of the 650 patients, 156 (24%) had had previous abdominal operations. In 25 cases, previous trocar sites were reused for optical access. The optical trocar was inserted at the umbilicus in 495 patients (76%), in the right upper quadrant in 77 (12%), in the left upper quadrant in 26 (4%), in the upper midline in eight (1%), in the right lower quadrant in six (0.9%), and in the left lower quadrant in three (0.5%). In 35 patients undergoing posterior adrenalectomy, optical trocars were used to enter Gerota's space. Mean (SD) entry times were 92 (45) sec at the umbilical site, 114 (30) sec at the back, and 77 (35) sec at the remaining sites. Complications (0.3%) included one injury to the bowel and one injury to the gallbladder; however, they were recognized and repaired immediately. Conclusions: To our knowledge, this report comprises the largest series in which the optical access trocar was used for laparoscopic surgery. This device provides the basis for a safe and fast technique for initial trocar placement: it also has the potential to reduce costs. Thanks to our favorable experience, the optical trocar method has become the standard technique for abdominal access in our laparoscopic practice since 1995.
引用
收藏
页码:570 / 573
页数:3
相关论文
共 50 条
  • [31] A safe and simple method for routine open access in laparoscopic procedures
    T. Lafullarde
    R. Van Hee
    T. Gys
    Surgical Endoscopy, 1999, 13 : 769 - 772
  • [32] The laparoscopic sound: A simple instrument for rapid and safe secondary trocar insertion in retroperitoneal laparoscopy
    Kyriakidis, AA
    EUROPEAN UROLOGY, 1996, 30 (04) : 471 - 474
  • [33] THE USE OF A DISPOSABLE LAPAROSCOPIC TROCAR FOR THE INJECTION AND SAFE EVACUATION OF ECHINOCOCCUS CYSTS OF THE LIVER
    VROMEN, A
    ZAMIR, O
    BEGLEIBTER, N
    FREUND, HR
    SURGERY GYNECOLOGY & OBSTETRICS, 1993, 177 (05): : 517 - 518
  • [34] Initial entry via the left upper quadrant with an optical trocar in laparoscopic bariatric surgery
    Amiki, Manabu
    Ishiyama, Yasuhiro
    Harada, Tatsunosuke
    Mochizuki, Ichitaro
    Tomizawa, Yuki
    Ito, Shingo
    Oneyama, Masataka
    Hara, Yoshiaki
    Narita, Kazuhiro
    Tachimori, Yuji
    Goto, Manabu
    Sekikawa, Koji
    Kuba, Motoko
    ASIAN JOURNAL OF ENDOSCOPIC SURGERY, 2022, 15 (02) : 463 - 466
  • [35] Optical trocar access for initial trocar placement in laparoscopic gastrointestinal surgery: A propensity score-matching analysis
    Tanaka, Chie
    Fujiwara, Michitaka
    Kanda, Mitsuro
    Murotani, Kenta
    Iwata, Naoki
    Hayashi, Masamichi
    Kobayashi, Daisuke
    Yamada, Suguru
    Kodera, Yasuhiro
    ASIAN JOURNAL OF ENDOSCOPIC SURGERY, 2019, 12 (01) : 37 - 42
  • [37] Optimizing first trocar access for robot-assisted radical prostatectomy: Optical trocar access through the upper abdominal quadrant using the Kii Fios First Entry trocar
    Shimbo, Masaki
    Endo, Fumiyasu
    Tominaga, Koki
    Sano, Masayuki
    Nishino, Takato
    Kyono, Yoko
    Komatsu, Kenji
    Ohyama, Takehiro
    Sakurai, Masato
    Narimoto, Kazutaka
    Matsushita, Kazuhito
    Hattori, Kazunori
    ASIAN JOURNAL OF ENDOSCOPIC SURGERY, 2021, 14 (03) : 443 - 450
  • [38] Primary midline peritoneal access with optical trocar is safe and effective in morbidly obese patients
    Sabeti, Nari
    Tarnoff, Michael
    Kim, Julie
    Shikora, Scott
    SURGERY FOR OBESITY AND RELATED DISEASES, 2009, 5 (05) : 610 - 614
  • [39] Learning curve of optical trocar access during laparoscopic pelvic surgery: A prospective study
    Gaia, G.
    Sighinolfi, M. Chiara
    Rocco, B.
    Cannoletta, M.
    Cannoletta, M.
    Sampogna, V
    Lamarca, A.
    Alboni, C.
    ACTAS UROLOGICAS ESPANOLAS, 2023, 47 (10): : 675 - 680
  • [40] Supraumbilical primary trocar insertion for laparoscopic access: the relationship between points of entry and retroperitoneal vital vasculature by imaging
    Stanhiser, Jamie
    Goodman, Linnea
    Soto, Enrique
    Al-Aref, Ibraheem
    Wu, Jenny
    Gojayev, Anar
    Nutter, Benjamin
    Falcone, Tommaso
    AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 2015, 213 (04) : 506.e1 - 506.e5