Time-Profile Measurement of an Emulsion Using Multiphoton Ionization Time-of-Flight Mass Spectrometry in Combination with a Microscope

被引:0
|
作者
Yosuke Shimo
Tomohiro Uchimura
机构
[1] University of Fukui,Department of Materials Science and Engineering, Graduate School of Engineering
来源
Analytical Sciences | 2016年 / 32卷
关键词
MPI-TOFMS; O/W emulsion;
D O I
暂无
中图分类号
学科分类号
摘要
Multiphoton ionization time-of-flight mass spectrometry was applied to the measurement of an oil-in-water emulsion that contained toluene as a dispersed phase. Before the measurement, the sample was sufficiently creamed, and then stirred for a short period of time for dispersion. As a result, several intense spikes appeared on the time profile constructed from the peak area for toluene. In the present study, an optical microscope was used to observe the capillary column for sample introduction, and small toluene droplets along with their aggregates were found in the images of the emulsion flowing through the capillary. The aggregates produced intense spikes comprised of multiple plots, which could be easily marked by applying a moving median filter. In the present study, droplets with minimum diameters as small as 2.9 μm, which corresponds to 13 fL, could be calculated as detectable spikes.
引用
收藏
页码:1059 / 1063
页数:4