Intraoperative electrode burns

被引:11
|
作者
Russell M.J. [1 ]
Gaetz M. [1 ]
机构
[1] Aaken Laboratories, Davis, CA 95616
关键词
Burns; Electric; Electrodes; Intraoperative; Monitoring; Neuromonitoring;
D O I
10.1023/B:JOCM.0000025254.73456.db
中图分类号
学科分类号
摘要
Objectives. The purpose of this paper is to increase awareness of electrode burns that occur in the operating room (OR), to discuss their causes, warning signs, and methods that may reduce their occurrence. Two general sources of electrical burns are described: 1) Excessive radio-frequency current density usually originating from a monopolar electrosurgical unit (ESU) and 2) Direct current (DC) originating from a battery or equipment malfunction. Methods. Case examples associated with intraoperative neuromonitoring are provided and discussed in the context of the relevant literature. Seven cases are described, five associated with excessive radio-frequency current density and two associated with direct current originating from a battery or equipment malfunction. Conclusions. Recommendations on prevention and warning signs associated with these burns are presented in addition to procedures to follow if a burn is suspected. © 2004 Kluwer Academic Publishers.
引用
收藏
页码:25 / 32
页数:7
相关论文
共 50 条
  • [21] Intraoperative assessment of an implantable electrode array for cavernous nerve stimulation
    Burnett, Arthur L.
    Teloken, Patrick E.
    Briganti, Alberto
    Whitehurst, Todd
    Montorsi, Francesco
    JOURNAL OF SEXUAL MEDICINE, 2008, 5 (08): : 1949 - 1954
  • [22] Analysis of Intraoperative Radiographic Electrode Placement During Cochlear Implantation
    Gnagi, Sharon H.
    Baker, Trenton R.
    Pollei, Taylor R.
    Barrs, David M.
    OTOLOGY & NEUROTOLOGY, 2015, 36 (06) : 1045 - 1047
  • [23] Addressing the Barriers to Bioimpedance Spectroscopy Use in Major Burns: Alternate Electrode Placement
    Kenworthy, Pippa
    Grisbrook, Tiffany L.
    Phillips, Michael
    Gibson, William
    Wood, Fiona M.
    Edgar, Dale W.
    JOURNAL OF BURN CARE & RESEARCH, 2017, 38 (06): : E952 - E959
  • [24] INTRAOPERATIVE ENDOCARDIAL ACTIVATION-MAPPING AND INTRAOPERATIVE ENDOCARDIAL PACE-MAPPING USING A NEW BALLOON ELECTRODE
    FANN, JI
    LOEB, JM
    LOCICERO, J
    FREDERIKSEN, JW
    MORAN, JM
    SURGICAL FORUM, 1984, 35 : 273 - 275
  • [25] New Backstrap Vagus Electrode for Continuous Intraoperative Neuromonitoring in Thyroid Surgery
    Lamade, Wolfram
    Ulmer, Christoph
    Rieber, Fabian
    Friedrich, Colin
    Koch, Klaus P.
    Thon, Klaus-Peter
    SURGICAL INNOVATION, 2011, 18 (03) : 206 - 213
  • [26] Use of Double Electrode Pads for Intraoperative Monitoring of the Recurrent Laryngeal Nerve
    Choby, Garret
    Trojanowski, Andrzej
    Johnson, Samuel
    Goldenberg, David
    ANNALS OF OTOLOGY RHINOLOGY AND LARYNGOLOGY, 2010, 119 (04): : 233 - 235
  • [27] Voltage Matrix Algorithm for Intraoperative Detection of Cochlear Implant Electrode Misplacement
    Franke-Trieger, Annett
    Lailach, Susen
    Strahl, Stefan B.
    Murrmann, Katrin
    Schebsdat, Erik
    Neudert, Marcus
    Zahnert, Thomas
    AUDIOLOGY AND NEUROTOLOGY, 2024,
  • [28] Intraoperative evaluation of the spiral nerve cuff electrode on the femoral nerve trunk
    Polasek, K. H.
    Schiefer, M. A.
    Pinault, G. C. J.
    Triolo, R. J.
    Tyler, D. J.
    JOURNAL OF NEURAL ENGINEERING, 2009, 6 (06)
  • [29] Intraoperative Neuromonitoring Using a Single Transcartilage Needle Electrode During Thyroidectomy
    Lee, Hyoung S.
    Seo, Sang G.
    Kim, Dong Y.
    Kim, Sung W.
    Choi, Young
    Lee, Kang D.
    LARYNGOSCOPE, 2021, 131 (02): : 448 - 452
  • [30] A reusable, self-adhesive electrode for intraoperative stimulation in the lower limbs
    Triolo, RJ
    Moss, JD
    Bhadra, N
    JOURNAL OF REHABILITATION RESEARCH AND DEVELOPMENT, 2001, 38 (05): : 527 - 532