Intraoperative electrode burns

被引:11
|
作者
Russell M.J. [1 ]
Gaetz M. [1 ]
机构
[1] Aaken Laboratories, Davis, CA 95616
关键词
Burns; Electric; Electrodes; Intraoperative; Monitoring; Neuromonitoring;
D O I
10.1023/B:JOCM.0000025254.73456.db
中图分类号
学科分类号
摘要
Objectives. The purpose of this paper is to increase awareness of electrode burns that occur in the operating room (OR), to discuss their causes, warning signs, and methods that may reduce their occurrence. Two general sources of electrical burns are described: 1) Excessive radio-frequency current density usually originating from a monopolar electrosurgical unit (ESU) and 2) Direct current (DC) originating from a battery or equipment malfunction. Methods. Case examples associated with intraoperative neuromonitoring are provided and discussed in the context of the relevant literature. Seven cases are described, five associated with excessive radio-frequency current density and two associated with direct current originating from a battery or equipment malfunction. Conclusions. Recommendations on prevention and warning signs associated with these burns are presented in addition to procedures to follow if a burn is suspected. © 2004 Kluwer Academic Publishers.
引用
收藏
页码:25 / 32
页数:7
相关论文
共 50 条
  • [1] Intraoperative positioning for circumferential extremity burns
    Shalom, A
    Wong, L
    BURNS, 2001, 27 (06) : 643 - 645
  • [2] ELECTRODE BURNS DURING LOCAL HYPERTHERMIA
    STOTT, DG
    WALLBANK, WA
    BRITISH JOURNAL OF ANAESTHESIA, 1993, 70 (03) : 370 - 371
  • [3] Intraoperative burns secondary to warmed IV bags: A warning
    Rosenfield, LK
    Pitlyk, PJ
    ANESTHESIOLOGY, 1999, 90 (02) : 616 - 618
  • [4] Software burns new path for electrode design
    不详
    PROFESSIONAL ENGINEERING, 2001, 14 (06) : 48 - 48
  • [5] A complication of intraoperative facial nerve monitoring: Facial skin burns
    Haenggeli, A
    Richter, M
    Lehmann, W
    Dulguerov, P
    AMERICAN JOURNAL OF OTOLOGY, 1999, 20 (05): : 679 - 682
  • [6] Intraoperative Correction of Cochlear Implant Electrode Translocation
    Morrel, William G.
    Manzoor, Nauman F.
    Dawant, Benoit M.
    Noble, Jack H.
    Labadie, Robert F.
    AUDIOLOGY AND NEURO-OTOLOGY, 2022, 27 (02) : 104 - 108
  • [7] Study on precise localization of intraoperative matrix electrode
    术中矩阵式电极的精确定位研究
    Jiang, Tao (taoojiang@163.com), 2017, West China Hospital, Sichuan Institute of Biomedical Engineering (34):
  • [8] Laryngeal mask airway use in children with acute burns: intraoperative airway management
    McCall, JE
    Fischer, CG
    Schomaker, E
    Young, JM
    PAEDIATRIC ANAESTHESIA, 1999, 9 (06): : 515 - 520
  • [9] Narcotrend and bispectral index for monitoring intraoperative anesthetic depth in patients with severe burns
    Zhang, Yongqiang
    Yang, Shuangmei
    Liu, Jun
    Wang, Yunfei
    Gao, Baoqin
    Cui, Wenjing
    Cao, Shu'e
    Zhang, Wei
    INTERNATIONAL JOURNAL OF CLINICAL AND EXPERIMENTAL MEDICINE, 2016, 9 (11): : 22020 - 22026
  • [10] TRANSTRAGAL, TRANSTYMPANIC ELECTRODE PLACEMENT FOR INTRAOPERATIVE ELECTROCOCHLEOGRAPHIC MONITORING
    PRASS, RL
    KINNEY, SE
    LUDERS, H
    OTOLARYNGOLOGY-HEAD AND NECK SURGERY, 1987, 97 (04) : 343 - 350