Effect of a pocket compression device on hematomas, skin reactions, and comfort in patients receiving a cardiovascular implantable electronic device: a randomized controlled trial

被引:3
|
作者
Hu, Jingwen [1 ]
Zheng, Jie [2 ]
Liu, Xiaoxue [1 ]
Li, Guoliang [1 ]
Xiao, Xianghua [3 ]
机构
[1] Xi An Jiao Tong Univ, Dept Cardiol, Affiliated Hosp 1, Xian, Peoples R China
[2] Xi An Jiao Tong Univ, Clin Res Ctr, Affiliated Hosp 1, Xian, Peoples R China
[3] Xian City First Hosp, Shaanxi Eye Inst, Xian, Peoples R China
关键词
cardiovascular implantable electronic device; pocket compression device; hematomas; skin reactions; and comfort;
D O I
10.1007/s10840-021-00973-5
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aim/objective Aim to evaluate the safety and effectiveness of a pocket compression device in patients undergoing cardiovascular implantable electronic device (CIED) implantation. Material and methods This randomized controlled trial included 54 patients in the control group (elastic adhesive tape) and 54 patients in the experimental group (pocket compression device). A sandbag was positioned over the pocket after CIED implantation, using elastic adhesive tape in the control group and a novel pocket compression device in the experimental group. The primary outcome was the incidence of pocket hematomas, and the secondary outcomes were adverse skin reactions, the need for sandbag positional adjustments, and patient comfort. Results The distributions of age, sex, and clinical and procedure characteristics was not significantly different between the two groups (P>0.05). The occurrence of hematomas was lower when using the pocket compression device than in the control group (13.0% vs. 44.4%, P<0.001). The incidence rates of hematomas of grades 1 and the grade 2 hematomas were significantly lower in the experimental group than in the control group (13.0% vs. 31.5%, 0% vs. 11.1% respectively, [all P<0.05]). The incidence rates of adverse skin reactions and sandbag positional adjustments were much lower and the patient comfort score was significantly higher in the experimental group than in the control group (all P<0.001). Conclusion A pocket compression device can significantly decrease the incidence of pocket hematomas, adverse skin reactions, and positional adjustments when patients are undergoing CIED, which can improve patient comfort and decrease the nursing workload.
引用
收藏
页码:275 / 281
页数:7
相关论文
共 50 条
  • [41] A RANDOMIZED CONTROLLED TRIAL USING A SKIN STRETCHING DEVICE IN BURN SCAR RECONSTRUCTION
    Verhaegen, Pauline
    Jongen, Sandra
    van Trier, Toine
    Bloemen, Monica
    van der Wal, Martijn
    Nieuwenhuis, Marianne
    Middelkoop, Esther
    van Zuijlen, Paul
    WOUND REPAIR AND REGENERATION, 2010, 18 (06) : A97 - A97
  • [42] Outcomes in Patients With Cardiovascular Implantable Electronic Device Infection Managed With Chronic Antibiotic Suppression
    Tan, Eugene M.
    DeSimone, Daniel C.
    Sohail, M. Rizwan
    Baddour, Larry M.
    Wilson, Walter R.
    Steckelberg, James M.
    Virk, Abinash
    CLINICAL INFECTIOUS DISEASES, 2017, 64 (11) : 1516 - 1521
  • [43] Cardiovascular Implantable Electronic Device Leads in CKD and ESRD Patients: Review and Recommendations for Practice
    Saad, Theodore F.
    Hentschel, Dirk M.
    Koplan, Bruce
    Wasse, Haimanot
    Asif, Arif
    Patel, Daniel V.
    Salman, Loay
    Carrillo, Roger
    Hoggard, Jeff
    SEMINARS IN DIALYSIS, 2013, 26 (01) : 114 - 123
  • [44] Impact of delayed device re-implantation on outcomes of patients with cardiovascular implantable electronic device related infective endocarditis
    Arshad, Verda
    Baddour, Larry M.
    Lahr, Brian D.
    Khalil, Sarwat
    Tariq, Wajeeha
    Talha, Khawaja Muhammad
    Cha, Yong-Mei
    DeSimone, Daniel C.
    Sohail, M. Rizwan
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2021, 44 (08): : 1303 - 1311
  • [45] Stroke in patients with cardiovascular implantable electronic device infection undergoing transvenous lead removal
    Lee, Justin Z.
    Agasthi, Pradyumna
    Pasha, Ahmed K.
    Tarin, Claudia
    Tseng, Andrew S.
    Diehl, Nancy N.
    Hodge, David O.
    DeSimone, Christopher V.
    Killu, Ammar M.
    Brady, Peter A.
    Kancharla, Krishna
    Kusumoto, Fred M.
    Srivathsan, Komandoor
    Osborn, Michael J.
    Espinosa, Raul E.
    Rea, Robert F.
    Madhavan, Malini
    McLeod, Christopher J.
    Shen, Win-Kuang
    Cha, Yong-Mei
    Friedman, Paul A.
    Asirvatham, Samuel J.
    Mulpuru, Siva K.
    HEART RHYTHM, 2018, 15 (11) : 1593 - 1600
  • [46] Use of antimicrobial agent pocket irrigation for cardiovascular implantable electronic device infection prophylaxis: Results from an international survey
    Zheng, Qi
    Di Biase, Luigi
    Ferrick, Kevin J.
    Gross, Jay N.
    Guttenplan, Nils A.
    Kim, Soo G.
    Krumerman, Andrew K.
    Palma, Eugen C.
    Fisher, John D.
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2018, 41 (10): : 1298 - 1306
  • [47] New skin closure system facilitates wound healing after cardiovascular implantable electronic device surgery
    De Maria, Elia
    WORLD JOURNAL OF CLINICAL CASES, 2015, 3 (08) : 675 - 677
  • [48] New skin closure system facilitates wound healing after cardiovascular implantable electronic device surgery
    Elia De Maria
    World Journal of Clinical Cases, 2015, (08) : 675 - 677
  • [49] Efficacy of Hypothermic Compression Bandages in Cardiac Device Surgical Wounds: A Randomized Controlled Trial
    Valls, Alba Cano
    Niebla, Mireia
    Diago, Clara
    Domingo, Rebeca
    Tolosana, Jose Maria
    Perez, Silvia
    ADVANCES IN SKIN & WOUND CARE, 2024, 37 (09) : 8 - 8
  • [50] Effect of a Novel Thermostatic Device on Meibomian Gland Dysfunction: A Randomized Controlled Trial in Chinese Patients
    Li, Siyuan
    Yang, Ke
    Wang, Jingyi
    Li, Shang
    Zhu, Lei
    Feng, Jun
    Tian, Lei
    Jie, Ying
    OPHTHALMOLOGY AND THERAPY, 2022, 11 (01) : 261 - 270