State entropy and bispectral index: correlation with end tidal sevoflurane concentrations

被引:0
|
作者
Rinaldi, S. [1 ]
Consales, G. [1 ]
De Gaudio, A. R. [1 ]
机构
[1] Univ Florence, Dept Crit Care, Anaesthesiol & Intens Care Unit, Florence, Italy
关键词
entropy; bispectral index; anesthesia monitoring; sevoflurane;
D O I
暂无
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Aim Aim of this study was to evaluate the difference between the correlations of state entropy (SE) and bispectral index (BIS) with different end tidal concentrations (Et) of sevoflurane. Methods. A total of 40 ASA I/II patients submitted to abdominal surgery were enrolled in the study. After premedication with fentanyl, anesthesia was induced with propofol and atracurium and maintained with sevoflurane in oxygen and air. The values of Et sevoflurane, SE and BIS were recorded during the operating room stay. The mean Et sevoflurane was calculated and the correlations of SE and BIS with Et sevoflurane were determined with Pearson test separately at values of Et sevoflurane below and higher the mean Et sevoflurane. Student's t-test was used to verify if the difference between the correlations was significant (P < 0.05). Results. The mean Et sevoflurane was 1.5%. During anesthesia the correlation with Et sevoflurane was -0.75 for SE and -0.70 for BIS. At Et sevoflurane higher than 1.5% the correlation with Et sevoflurane was -0.41 for SE and -0.14 for BIS. At Et sevoflurane below 1.5% the correlation with Et sevoflurane was -0.65 for SE, and -0.8 for BIS. The difference between the correlations of Et sevoflurane with SE and BIS resulted statistically significant. Conclusion. During sevoflurane anesthesia SE shows a better correlation than BIS with Et sevoflurane. Although at low Et sevoflurane the correlation is better with BIS, at high Et sevoflurane the correlation is better with SE.
引用
收藏
页码:39 / 48
页数:10
相关论文
共 50 条
  • [21] Correlation and agreement between bispectral index and state entropy of the electroencephalogram during propofol anaesthesia
    Bonhomme, V.
    Deflandre, E.
    Hans, P.
    BRITISH JOURNAL OF ANAESTHESIA, 2006, 97 (03) : 340 - 346
  • [22] The correlation of bispectral index with endtidal sevoflurane concentration and haemodynamic parameters in preschoolers
    McCann, ME
    Bacsik, J
    Davidson, A
    Auble, S
    Sullivan, L
    Laussen, P
    PAEDIATRIC ANAESTHESIA, 2002, 12 (06): : 519 - 525
  • [23] The correlation between bispectral index and airway reflexes with sevoflurane and halothane anaesthesia
    Davidson, A
    PEDIATRIC ANESTHESIA, 2004, 14 (03) : 241 - 246
  • [24] Determining entropy values equivalent to the bispectral index values during sevoflurane anaesthesia
    Balci, Canan
    Karabekir, Hamit
    Sivaci, Remziye
    ARCHIVES OF MEDICAL SCIENCE, 2010, 6 (03) : 370 - 374
  • [25] Comparison of Bispectral Index and Entropy values with electroencephalogram during surgical anaesthesia with sevoflurane
    Aho, A. J.
    Kamata, K.
    Jantti, V.
    Kulkas, A.
    Hagihira, S.
    Huhtala, H.
    Yli-Hankala, A.
    BRITISH JOURNAL OF ANAESTHESIA, 2015, 115 (02) : 258 - 266
  • [26] Does Nitrous Oxide Affect Bispectral Index and State Entropy When Added to a Propofol Versus Sevoflurane Anesthetic?
    Ozcan, Mehmet S.
    Ozcan, Mukadder D.
    Khan, Qaiser S.
    Thompson, David M.
    Chetty, Pramod K.
    JOURNAL OF NEUROSURGICAL ANESTHESIOLOGY, 2010, 22 (04) : 309 - 315
  • [27] Entropy indices vs the bispectral index™ for estimating nociception during sevoflurane anaesthesia
    Takamatsu, I
    Ozaki, M
    Kazama, T
    BRITISH JOURNAL OF ANAESTHESIA, 2006, 96 (05) : 620 - 626
  • [28] Minimum alveolar concentrations of sevoflurane for maintaining bispectral index below 50 in children
    Tsuruta, Shouhei
    Satsumae, Tsuyoshi
    Mizutani, Taro
    Inomata, Shinichi
    Shimizu, Takeru
    Takahashi, Shinji
    Tanaka, Makoto
    PEDIATRIC ANESTHESIA, 2011, 21 (11) : 1124 - 1127
  • [29] Bispectral index (BIS™) monitoring and end-tidal isoflurane concentrations during anesthesia for liver transplantation.
    Schumann, R.
    Hudcova, J.
    Anderson, C.
    Bonney, I.
    LIVER TRANSPLANTATION, 2007, 13 (06) : S64 - S65
  • [30] Comparison between Bispectral Index and Patient State Index as Measnres of the Electroencephalographic Effects of Sevoflurane
    Soehle, Martin
    Ellerkmann, Richard K.
    Grube, Matthias
    Kuech, Matthias
    Wirz, Stefan
    Hoeft, Andreas
    Bruhn, Joergen
    ANESTHESIOLOGY, 2008, 109 (05) : 799 - 805