Three-Dimensional Accuracy of a Digitally Coded Healing Abutment Implant Impression System

被引:16
|
作者
Ng, Simon D. [1 ]
Tan, Keson B. [2 ]
Teoh, K. H. [1 ]
Cheng, Ansgar C. [2 ]
Nicholls, Jack I. [3 ]
机构
[1] Natl Dent Ctr Singapore, Dept Restorat Dent, Singapore, Singapore
[2] Natl Univ Singapore, Fac Dent, Singapore 119083, Singapore
[3] Univ Washington, Sch Dent, Dept Restorat Dent, Seattle, WA 98195 USA
关键词
accuracy; coordinate measuring machine; digital measurement; implant impressions; three-dimensional measurement; SUPPORTED PROSTHESES; RESTORATION; CONNECTION; DENTISTRY; DESIGN;
D O I
10.11607/jomi.3433
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: This study examined the three-dimensional (3D) accuracy of the Encode Impression System (EN) in transferring the locations of two implants from master models to test models and compared this to the direct impression (DI) technique. The effect of interimplant angulation on the 3D accuracy of both impression techniques was also evaluated. Materials and Methods: Seven sectional polymethyl methacrylate mandibular arch master models were fabricated with implants in the first premolar and first molar positions. The implants were placed parallel to each other or angulated mesiodistally or buccolingually with total divergent angles of 10, 20, or 30 degrees. Each master model was secured onto an aluminum block containing a gauge block, which defined the local coordinate references. Encode healing abutments were attached to the implants before impressions were made for the EN test models; pickup impression copings were attached for the DI test models. For the seven test groups of each impression technique, a total of 70 test models were fabricated (n = 5). The EN test models were sent to Biomet 3i for implant analog placement. The centroid of each implant or implant analog and the angular orientation of the long axis relative to the x- and y-axes were measured with a coordinate measuring machine. Statistical analyses were performed. Results: Impression technique had a significant effect on y distortion, global linear distortion, and absolute xz and yz angular distortions. Interimplant angulation had significant effects on x and y distortions. However, neither impression technique nor interimplant angulation had a significant effect on z distortion. Conclusions: Distortions were observed with both impression techniques. However, the results suggest that EN was less accurate than DI.
引用
收藏
页码:927 / 936
页数:10
相关论文
共 50 条
  • [1] The Accuracy of an Implant Impression Technique Using Digitally Coded Healing Abutments
    Eliasson, Alf
    Ortorp, Anders
    CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2012, 14 : e30 - e38
  • [2] Three-Dimensional Accuracy of Implant and Abutment Level Impression Techniques: Effect on Marginal Discrepancy
    Alikhasi, Marzieh
    Siadat, Hakimeh
    Monzavi, Abbas
    Momen-Heravi, Fatemeh
    JOURNAL OF ORAL IMPLANTOLOGY, 2011, 37 (06) : 649 - 657
  • [3] Accuracy of implant impression techniques with a scannable healing abutment
    Jung, Hong-Taek
    Kim, Ha-Young
    Song, So-Yeon
    Park, Jin-Hong
    Lee, Jeong-Yol
    JOURNAL OF PROSTHETIC DENTISTRY, 2022, 128 (04): : 729 - 734
  • [4] Three-Dimensional Accuracy of Plastic Transfer Impression Copings for Three Implant Systems
    Teo, Juin Wei
    Tan, Keson B.
    Nicholls, Jack I.
    Wong, Keng Mun
    Uy, Joanne
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2014, 29 (03) : 577 - 584
  • [5] ACCURACY OF IMPLANT IMPRESSIONS WITHOUT IMPRESSION COPINGS: A THREE-DIMENSIONAL ANALYSIS
    Kwon, Joo-Hyun
    Son, Yong-Ha
    Han, Chong-Hyun
    Kim, Sunjai
    JOURNAL OF PROSTHETIC DENTISTRY, 2011, 105 (06): : 367 - 373
  • [6] Three-dimensional accuracy of innovative implant-level impression techniques with plastic snap-on impression copings
    Asli, Hamid Neshandar
    Hemmati, Yasamin Babaee
    Falahchai, Mehran
    DENTAL AND MEDICAL PROBLEMS, 2021, 58 (03) : 351 - 357
  • [7] Accuracy of Digital vs Conventional Implant Impression Approach: A Three-Dimensional Comparative In Vitro Analysis
    Basaki, Kinga
    Alkumru, Hasan
    De Souza, Grace
    Finer, Yoav
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2017, 32 (04) : 792 - 799
  • [8] Accuracy of digitally coded healing abutments: A systematic review
    Talesara, Vaibhav
    Bennani, Vincent
    Aarts, John
    Ratnayake, Jithendra
    Khurshid, Zohaib
    Brunton, Paul
    SAUDI DENTAL JOURNAL, 2023, 35 (08) : 891 - 903
  • [9] Evaluation of the Three-Dimensional Accuracy of Implant Impression Techniques in Two Simulated Clinical Conditions by Optical Scanning
    Sabouhi, Mahmoud
    Bajoghli, Farshad
    Abolhasani, Majid
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2015, 30 (01) : 26 - 34
  • [10] Three-Dimensional Accuracy of Digital Impression versus Conventional Method: Effect of Implant Angulation and Connection Type
    Alikhasi, Marzieh
    Siadat, Hakime
    Nasirpour, Alireza
    Hasanzade, Mahya
    INTERNATIONAL JOURNAL OF DENTISTRY, 2018, 2018