Dose calculation validation of VMC++ for photon beams

被引:27
|
作者
Gardner, J. [1 ]
Siebers, J.
Kawrakow, I.
机构
[1] Virginia Commonwealth Univ, Dept Radiat Oncol, Richmond, VA 23298 USA
[2] Natl Res Council Canada, Ottawa, ON K1A 0R6, Canada
关键词
D O I
10.1118/1.2714473
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The radiation therapy specific Voxel Monte Carlo (VMC++) dose calculation algorithm achieves a dramatic improvement in MC dose calculation efficiency for radiation therapy treatment planning dose evaluation compared with other MC algorithms. This work aims to validate VMC++ for radiation therapy photon beam planning. VMC++ was validated with respect to the well-benchmarked EGS-based DOSXYZnrc by comparing depth dose and lateral profiles for field sizes ranging from I x 1 to 40 x 40 cm(2) for 6 and 18 MV beams in a homogeneous water phantom and in a simulated bone-lung-bone phantom. Patient treatment plan dose distributions were compared for five prostate plans and five head-and-neck (H/N) plans, all using intensity-modulated radiotherapy beams. For all tests, the same incident particles were used in both codes to isolate differences due to modeling of the radiation source. Voxel-by-voxel observed differences were analyzed to distinguish between systematic and purely statistical differences. Dose-volume-histogram-derived dose indices were compared for the patient plans. For the homogeneous water phantom and the bone-lung-bone phantom, the depth dose curve predicted by VMC++ agreed with that predicted by DOSXYZnrc within expected statistical uncertainty in all voxels except the surface voxel of the water phantom, where VMC++ predicted a lower dose. When the electron cutoff parameter was decreased for both codes, the surface voxel agreed within expected statistical uncertainty. For prostate plans, the most severe difference between the codes resulted in 55% of the voxels showing a systematic difference of 0.32% of maximum dose. For H/N plans, the largest difference observed resulted in 2% of the voxels showing a systematic difference of 0.98% of maximum dose. For the prostate plans, the most severe difference in the planning target volume D-95 was 0.4%, the rectum D-35 was 0.2%, the rectum D-17 was 0.2%, the bladder D-50 was 0.3% and the bladder D-25 was 0.3%. For the H/N plans, the most severe difference in the gross tumor volume D-98 was 0.4%, the clinical target volume D-90 was 0.2%, the nodes D-90 was 0.2%, the parotids D-95 was 0.8%, and the cord D-2 was 0.8%. All of these differences are clinically insignificant. VMC++ showed an average efficiency gain over DOSXYZnrc of at least an order of magnitude without introducing significant systematic bias. VMC++ can be used for photon beam MC patient dose computations without a clinically significant loss in accuracy. (c) 2007 American Association of Physicists in Medicine.
引用
收藏
页码:1809 / 1818
页数:10
相关论文
共 50 条
  • [41] Dose calculation accuracy in the build-up region of flattening filter-free photon beams
    De Puysseleyr, A.
    Lechner, W.
    Georg, D.
    De Wagter, C.
    RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S71 - S72
  • [42] A METHOD FOR DOSE CALCULATION FOR HIGH-ENERGY PHOTON BEAMS BASED ON MEASUREMENTS PERFORMED AT REFERENCE DEPTH
    BRIDIER, A
    DUTREIX, A
    ACTA ONCOLOGICA, 1993, 32 (04) : 425 - 433
  • [43] The accuracy of dose calculation in wedge fields of the Analytical Anisotropic Algorithm for 6 and 10 MV photon beams
    Rutchantuek, S.
    Stansook, N.
    Changkaew, P.
    Sakulsingharoj, S.
    RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S179 - S180
  • [44] Dose calculations for external photon beams in radiotherapy
    Ahnesjö, A
    Aspradakis, MM
    PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (11): : R99 - R155
  • [45] SURFACE DOSE MEASUREMENTS IN CLINICAL PHOTON BEAMS
    NILSSON, B
    SORCINI, B
    ACTA ONCOLOGICA, 1989, 28 (04) : 537 - 542
  • [46] Scatter dose model for therapeutic photon beams
    Nizin, PS
    Mooij, RB
    PROCEEDINGS OF THE 22ND ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4, 2000, 22 : 1827 - 1830
  • [47] DOSE BUILDUP FOR OBLIQUELY INCIDENT PHOTON BEAMS
    GERBI, BJ
    MEIGOONI, AS
    KHAN, FM
    MEDICAL PHYSICS, 1987, 14 (03) : 393 - 399
  • [48] CORRECTIONS TO SCATTER DOSE IN LUNG FOR PHOTON BEAMS
    RICE, RK
    BJARNGARD, BE
    CHIN, LM
    SIDDON, RL
    MEDICAL PHYSICS, 1987, 14 (04) : 708 - 708
  • [49] Calculation of effective doses for broad parallel photon beams
    Kim, CH
    Reece, WD
    Poston, JW
    HEALTH PHYSICS, 1999, 76 (02): : 156 - 161
  • [50] Dosimetric verification of a commercial Monte Carlo treatment planning system (VMC++) for a 9 MeV electron beam
    Schiapparelli, P.
    Zefiro, D.
    Taccini, G.
    MEDICAL PHYSICS, 2009, 36 (05) : 1759 - 1767