Evaluation of delivered dose to a moving target by 4D dose reconstruction in gated volumetric modulated arc therapy

被引:8
|
作者
Chung, Hyekyun [1 ,2 ]
Jung, Jinhong [1 ]
Jeong, Chiyoung [1 ]
Kwak, Jungwon [1 ]
Park, Jin-hong [1 ]
Kim, Su Ssan [1 ]
Yoon, Sang Min [1 ]
Song, Si Yeol [1 ]
Kim, Jong Hoon [1 ]
Choi, Eun Kyung [1 ]
Cho, Seungryong [2 ]
Cho, Byungchul [1 ]
机构
[1] Univ Ulsan, Dept Radiat Oncol, Coll Med, Asan Med Ctr, Seoul, South Korea
[2] Korea Adv Inst Sci & Technol, Dept Nucl & Quantum Engn, Daejeon, South Korea
来源
PLOS ONE | 2018年 / 13卷 / 09期
基金
新加坡国家研究基金会;
关键词
STEREOTACTIC BODY RADIOTHERAPY; CELL LUNG-CANCER; PROSPECTIVE PHASE-II; RADIATION-THERAPY; PANCREATIC-CANCER; DOSIMETRIC IMPACT; BREATHING MOTION; LIVER-TUMORS; CARCINOMA; RAPIDARC;
D O I
10.1371/journal.pone.0202765
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Purpose To develop a 4D dose reconstruction method and to evaluate the delivered dose in respiratory-gated volumetric modulated arc therapy (VMAT). Materials and methods A total 112 treatment sessions of gated VMAT for 30 stereotactic body radiotherapy (SBRT) patients (10 lung, 10 liver, and 10 pancreas) were evaluated. For respiratory-gated SBRT, 4DCT was acquired, and the CT data at the end-exhale phase was used for a VMAT plan. The delivered dose was reconstructed using a patient's respiratory motion and machine motion acquired during the beam delivery. The machine motion was obtained from the treatment log file, while the target position was estimated from an external respiratory marker position. The target position was divided into 1-mm position bins, and sub-beams with beam isocenters corresponding to each position bin were created in a motion mimicking plan, reflecting motion data including MLC leaf positions and gantry angle and target position data during beam treatment. The reconstructed 4D dose was compared with the dose of the original plan using these dosimetric parameters; the maximum dose (D-max) and mean dose (D-mean) of gross target volume (GTV) or organs at risk (spinal cord, esophagus, heart, duodenum, kidney, spinal cord, and stomach). The minimum dose (Dmin) to GTV was also calculated to verify cold spots in tumors. Results There was no significant difference of dose parameters regard to the GTV in all tumors. For the liver cases, there were significant differences in the D-max of duodenum (-4.2 +/- 1.4%), stomach (-3.5 +/- 4.2%), left kidney (-4.1 +/- 2.8%), and right kidney (-3.2 +/- 1.3%), and in the D-mean of duodenum (-3.8 +/- 1.4%), stomach (-3.9 +/- 2.2%), left kidney (-3.1 +/- 2.8%), and right kidney (-4.1 +/- 2.6%). For the pancreas cases, there were significant differences in the D-max of stomach (2.1 +/- 3.0%), and in the D-mean of liver (1.5 +/- 0.6%), duodenum (-1.0 +/- 1.4%), stomach (2.1 +/- 1.6%), and right kidney (-1.3 +/- 0.9%). The average gamma pass rates were 97.6 +/- 4.8% for lung cases, 99.6 +/- 0.5% for liver cases, and 99.5 +/- 0.5% for pancreas cases. Most cases showed insignificant dose variation, with gamma pass rates higher than 98%, except for two lung cases with gamma pass rates of 86.9% and 90.6%. The low gamma pass rates showed larger global motion ranges resulting from the baseline shift during beam delivery. Conclusion The actual delivered dose in thoracic and abdominal VMAT under breathing motion was verified by 4D dose reconstruction using typical treatment equipment and software. The proposed method provides a verification method for the actual delivered dose and could be a dosimetric verification QA tool for radiation treatment under various respiratory management techniques.
引用
收藏
页数:18
相关论文
共 50 条
  • [31] GATED DELIVERY OF VOLUMETRIC MODULATED ARC THERAPY WITH RAPIDARC: PRE-CLINICAL EVALUATION BASED ON DOSE MEASUREMENTS AND LOG FILES ANALYSIS
    Nicolini, G.
    de Palma, E. Vanetti
    Clivio, A.
    Fogliata, A.
    Cozzi, L.
    RADIOTHERAPY AND ONCOLOGY, 2010, 96 : S151 - S152
  • [32] Real-time dose reconstruction during volumetric modulated arc therapy with dynamic MLC tracking
    Ravkilde, T.
    O'Brien, R.
    Keall, P. J.
    Poulsen, P. R.
    RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S172 - S173
  • [33] Improvement of Dose Distributions in Volumetric Modulated Arc Therapy (RapidArc) with Jaw Tracking Technique for Lateralized Target
    Diaz Moreno, R.
    Alarcon, A.
    Apaza, O.
    Paredes, L.
    Venencia, C.
    MEDICAL PHYSICS, 2019, 46 (06) : E561 - E561
  • [34] Dynamically accumulated dose and 4D accumulated dose for moving tumors
    Li, Heng
    Li, Yupeng
    Zhang, Xiaodong
    Li, Xiaoqiang
    Liu, Wei
    Gillin, Michael T.
    Zhu, X. Ronald
    MEDICAL PHYSICS, 2012, 39 (12) : 7359 - 7367
  • [35] 4D Treatment Dose Reconstruction for Scanned Ion Beam Therapy
    Richter, D.
    Saito, N.
    Chaudhri, N.
    Haertig, M.
    Combs, S.
    Habermehl, D.
    Herfarth, K.
    Jaekel, O.
    Durante, M.
    Bert, C.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2013, 87 (02): : S183 - S183
  • [36] 4D dose reconstruction for prostate SBRT
    Ehrbar, Stefanie
    Ehrbar, S.
    Hagen, T.
    Lang, S.
    Poulsen, P.
    Guckenberger, M.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2019, 195 (12) : 1128 - 1128
  • [37] INFLUENCE OF TREATMENT COUCH ATTENUATION ON DOSE DELIVERED BY RAPID ARC TM - VOLUMETRIC MODULATED ARC THERAPY MEASURED BY A 2D ION CHAMBER ARRAY
    Gomola, I.
    Depuydt, T.
    Bocanek, J.
    RADIOTHERAPY AND ONCOLOGY, 2008, 88 : S387 - S387
  • [38] Evaluation of 3D and 4D Delivered Dose with AAA and XVMC Dose Calculation Algorithms in Stereotactic Body Radiation Therapy for Lung Cancer
    Matsugi, K.
    Sawada, A.
    Nakamura, M.
    Miyabe, Y.
    Yamauchi, C.
    Hiraoka, M.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [39] Comparison of testicular dose delivered by intensity-modulated radiation therapy (IMRT) and volumetric-modulated arc therapy (VMAT) in patients with prostate cancer
    Martin, Jeffrey M.
    Handorf, Elizabeth A.
    Price, Robert A.
    Cherian, George
    Buyyounouski, Mark K.
    Chen, David Y.
    Kutikov, Alexander
    Johnson, Matthew E.
    Ma, Chung-Ming Charlie
    Horwitz, Eric M.
    MEDICAL DOSIMETRY, 2015, 40 (03) : 186 - 189
  • [40] A Probability Density Function Based Dose Calculation for Gated Volumetric Modulated Arc Therapy in Lung Stereotactic Body Radiotherapy
    Zhang, Y.
    Yang, Y.
    Li, X.
    Li, T.
    Heron, D.
    Huq, M.
    MEDICAL PHYSICS, 2013, 40 (06)