Fatigue testing of a proximal femoral hip component

被引:3
|
作者
Macdonald, W
Carlsson, LV
Gathercole, N
Jacobsson, CM
机构
[1] Univ Bath, Dept Engn & Appl Sci, Mat Res Ctr, SMS Consultants, Bath BA2 7AY, Avon, England
[2] Univ Gothenburg, Sahlgrens Univ Hosp, Dept Orthopaed, Gothenburg, Sweden
[3] Univ Gothenburg, Inst Surg Sci, Dept Biomat Res, Gothenburg, Sweden
关键词
hip replacement; fatigue testing; femoral component; GFRE bushings; GOT implant;
D O I
10.1243/09544110360579358
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The Gothenburg Osseointegrated Titanium (GOT) implant is a novel total hip replacement including a metaphyseal loading proximal femoral component fixed in the retained femoral neck. Endurance testing was performed under conditions analogous to ISO 7206-4: 1989. The cement-free implant is not fixed distally within the intramedullary canal, so distal embedding (as specified in the standard) would have been unrealistic. Instead glass-fibre-reinforced epoxy (GFRE) bushings were used to model reduced bone support mid-length at the medial cortex and distally at the lateral cortex. Such support simulated proximal bone loss, realistically reproducing the effect of osteolysis or fixation failure. Under such conditions the component survived unbroken for 10 million cycles at 3.0 kN peak load.
引用
收藏
页码:137 / 145
页数:9
相关论文
共 50 条
  • [31] Reorientational Proximal Femoral Osteotomies for Arthrogrypotic Hip Contractures
    Pieter van Bosse, Harold Jacob
    Saldana, Roger E.
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2017, 99 (01): : 55 - 64
  • [32] Total hip arthroplasty in patients with proximal femoral deformity
    Berry, DJ
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1999, (369) : 262 - 272
  • [33] Total hip arthroplasty for complications of proximal femoral fractures
    Tabsh, I
    Waddell, JP
    Morton, J
    JOURNAL OF ORTHOPAEDIC TRAUMA, 1997, 11 (03) : 166 - 169
  • [34] Varus proximal femoral osteotomy for hip dysplasia in adults
    Ansari, A.
    Jones, S.
    Hashemi-Nejad, A.
    Catterall, A.
    HIP INTERNATIONAL, 2008, 18 (03) : 200 - 206
  • [35] PROXIMAL FEMORAL ALLOGRAFTS IN REVISION HIP-ARTHROPLASTY
    ALLAN, DG
    LAVOIE, GJ
    MCDONALD, S
    OAKESHOTT, R
    GROSS, AE
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1991, 73 (02): : 235 - 240
  • [36] ACCIDENTAL DRILL BIT SCORING OF A TOTAL HIP-ARTHROPLASTY FEMORAL COMPONENT WITH SUBSEQUENT FATIGUE FRACTURE
    HARPER, MC
    RALSTON, M
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1983, (173) : 173 - 177
  • [37] Total hip arthroplasty with femoral osteotomy and modular prosthesis for proximal femoral deformity
    Deng, Xiaowen
    Liu, Jun
    Qu, Tao
    Li, Xusheng
    Zhen, Ping
    Gao, Qiuming
    Xue, Yun
    Liu, Peng
    Cao, Guoding
    He, Xiaole
    JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2019, 14 (01)
  • [38] Total hip arthroplasty with femoral osteotomy and modular prosthesis for proximal femoral deformity
    Xiaowen Deng
    Jun Liu
    Tao Qu
    Xusheng Li
    Ping Zhen
    Qiuming Gao
    Yun Xue
    Peng Liu
    Guoding Cao
    Xiaole He
    Journal of Orthopaedic Surgery and Research, 14
  • [39] Age-Associated Changes in Proximal Femur Morphology Affect Femoral Component Sizing in Cementless Hip Arthroplasty
    Veldman, Hidde D.
    Boymans, Tim A. E. J.
    van Steenbergen, Liza N.
    Heyligers, Ide C.
    GERIATRIC ORTHOPAEDIC SURGERY & REHABILITATION, 2022, 13
  • [40] Expandable Proximal Femoral Nail versus Gamma Proximal Femoral Nail for the treatment of hip reverse oblique fractures
    Yaniv Warschawski
    Ran Ankori
    Tal Frenkel Rutenberg
    Ely L. Steinberg
    Ran Atzmon
    Michael Drexler
    Archives of Orthopaedic and Trauma Surgery, 2022, 142 : 777 - 785