Fatigue testing of a proximal femoral hip component

被引:3
|
作者
Macdonald, W
Carlsson, LV
Gathercole, N
Jacobsson, CM
机构
[1] Univ Bath, Dept Engn & Appl Sci, Mat Res Ctr, SMS Consultants, Bath BA2 7AY, Avon, England
[2] Univ Gothenburg, Sahlgrens Univ Hosp, Dept Orthopaed, Gothenburg, Sweden
[3] Univ Gothenburg, Inst Surg Sci, Dept Biomat Res, Gothenburg, Sweden
关键词
hip replacement; fatigue testing; femoral component; GFRE bushings; GOT implant;
D O I
10.1243/09544110360579358
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The Gothenburg Osseointegrated Titanium (GOT) implant is a novel total hip replacement including a metaphyseal loading proximal femoral component fixed in the retained femoral neck. Endurance testing was performed under conditions analogous to ISO 7206-4: 1989. The cement-free implant is not fixed distally within the intramedullary canal, so distal embedding (as specified in the standard) would have been unrealistic. Instead glass-fibre-reinforced epoxy (GFRE) bushings were used to model reduced bone support mid-length at the medial cortex and distally at the lateral cortex. Such support simulated proximal bone loss, realistically reproducing the effect of osteolysis or fixation failure. Under such conditions the component survived unbroken for 10 million cycles at 3.0 kN peak load.
引用
收藏
页码:137 / 145
页数:9
相关论文
共 50 条
  • [1] Fatigue failure of the femoral component of a cementless total hip arthroplasty
    Morgan-Hough, CVJ
    Tavakkolizadeh, A
    Purkayastha, S
    JOURNAL OF ARTHROPLASTY, 2004, 19 (05): : 658 - 660
  • [2] Material optimisation of the femoral component of a hip prosthesis based on the fatigue notch fatigue approach
    Hedia, HS
    Barton, DC
    Fisher, J
    BIO-MEDICAL MATERIALS AND ENGINEERING, 1997, 7 (02) : 83 - 98
  • [3] A REPRODUCIBLE FATIGUE TEST METHOD FOR THE FEMORAL COMPONENT OF TOTAL HIP PROSTHESES
    DUCHEYNE, P
    WEVERS, M
    DEMEESTER, P
    JOURNAL OF BIOMECHANICS, 1982, 15 (10) : 809 - 809
  • [4] Fatigue behavior of a titanium femoral hip prosthesis with proximal sleeve-stem modularity
    Krygier, Jan Janusz, 1600, John Wiley & Sons Inc, New York, NY, United States (05):
  • [5] HIP SALVAGE BY PROXIMAL FEMORAL REPLACEMENT
    SIM, FH
    CHAO, EYS
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1981, 63 (08): : 1228 - 1238
  • [6] Influence of Femoral Component Design on Proximal Femoral Bone Mass After Total Hip Replacement A Randomized Controlled Trial
    Slullitel, Pablo A.
    Mahatma, Mohit M.
    Farzi, Mohsen
    Grammatopoulos, George
    Wilkinson, J. Mark
    Beaule, Paul E.
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2021, 103 (01): : 74 - 83
  • [7] Effect of proximal femoral bone support on the fixation of a press-fit noncemented total hip replacement femoral component
    Sangiorgio, Sophia N.
    Ebramzadeh, Edward
    Borkowski, Sean L.
    Oakes, Daniel A.
    Reid, Jeremy J.
    Bengs, Benjamin C.
    JOURNAL OF APPLIED BIOMATERIALS & FUNCTIONAL MATERIALS, 2013, 11 (01): : 26 - 34
  • [8] FATIGUE BEHAVIOR OF A TITANIUM FEMORAL HIP-PROSTHESIS WITH PROXIMAL SLEEVE-STEM MODULARITY
    KRYGIER, JJ
    DUJOVNE, AR
    BOBYN, JD
    JOURNAL OF APPLIED BIOMATERIALS, 1994, 5 (03) : 195 - 201
  • [9] Revision total hip replacement using the bias proximal porous-coated femoral component
    Suominen, S
    Santavirta, S
    ANNALES CHIRURGIAE ET GYNAECOLOGIAE, 1996, 85 (04) : 347 - 351
  • [10] Total hip arthroplasty with femoral osteotomy for proximal femoral deformity
    Papagelopoulos, PJ
    Trousdale, RT
    Lewallen, DG
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1996, (332) : 151 - 162