Axial Length Measurement in Asteroid Hyalosis: Comparing Laser Interferometry and Ultrasound A-scan

被引:0
|
作者
Moghimi, Sasan [1 ]
Kiumehr, Saman
Amoozadeh, Javad [1 ]
Fakhraie, Ghasem [1 ]
机构
[1] Univ Tehran Med Sci, Farabi Eye Hosp, Eye Res Ctr, Tehran, Iran
来源
IRANIAN JOURNAL OF OPHTHALMOLOGY | 2009年 / 21卷 / 01期
关键词
Laser Interferometry; Ultrasound; Asteroid Hyalosis; Axial Length; PARTIAL COHERENCE INTERFEROMETRY; DIABETES-MELLITUS; BIOMETRY; POPULATION; HYALITIS; EYES;
D O I
暂无
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To compare laser interferometry (LI) and ultrasound A-scan in measurement of axial length (AL) in eyes with asteroid hyalosis Methods: Twenty three patients with unilateral moderate to severe asteroid hyalosis determined by B-scan ultrasonography images participated in the study. The AL measurements were obtained by LI and ultrasound A-scan in both asteroid hyalosis and noninvolved eyes. The measurements were performed by the two methods as well as for affected and contralateral normal eyes and were statistically compared using the paired T-test. A P-value less than 0.05 was considered significant. Results: The ultrasound AL measurements were taken in every case but measurement with the LI method was not possible in 4 eyes with asteroid hyalosis (17.3%). There was a high correlation between LI and ultrasound A-scan measurements in both asteroid hyalosis and noninvolved eyes (r(2)=0.97, P<0.001 and r(2)=0.95, P<0.001 respectively). There was no significant difference between AL taken by either method in asteroid hyalosis and noninvolved eyes. Conclusion: The failure rate of AL measurement using LI was relatively high in eyes with asteroid hyalosis. There was a high correlation between AL measured by ultrasound A-scan and LI method in both asteroid hyalosis and noninvolved eyes.
引用
收藏
页码:29 / 34
页数:6
相关论文
共 50 条
  • [21] Measurement of the axial eye length by wavelength tuning interferometry
    Lexer, F
    Hitzenberger, CK
    Kulhavy, M
    Fercher, AF
    LASERS IN OPHTHALMOLOGY IV, PROCEEDINGS OF, 1996, 2930 : 202 - 206
  • [22] Assessment of a new through-the-eyelid technique for 'A' scan ultrasound ocular axial length measurement
    Laws, F
    Laws, D
    Wood, I
    Clark, D
    OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 1998, 18 (05) : 408 - 414
  • [23] IDENTIFICATION AND MEASUREMENT OF ALL COMPARTMENTS OF HUMAN-SKIN INVIVO BY A-SCAN ULTRASOUND
    EDWARDS, C
    ALABOOSI, M
    MARKS, R
    JOURNAL OF INVESTIGATIVE DERMATOLOGY, 1986, 87 (01) : 137 - 137
  • [24] Axial Length Measurement by Immersion B-Scan
    Daly, Suzanne
    Coleman, D. Jackson
    Trief, Danielle
    Silverman, Ronald H.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2019, 60 (09)
  • [25] Axial length measurement in silicone oil-filled eyes using laser Doppler interferometry
    Chae, JB
    Park, HR
    Yoon, YH
    RETINA-THE JOURNAL OF RETINAL AND VITREOUS DISEASES, 2004, 24 (04): : 655 - 657
  • [26] LOCALIZED PARTICLE CONCENTRATION MEASUREMENT IN SLURRY FLOWS USING A-SCAN ULTRASOUND TECHNIQUE
    Furlan, John M.
    Mundla, Venkat
    Kadambi, Jaikrishnan
    Hoyt, Nathaniel
    Visintainer, Robert
    Addie, Greame
    PROCEEDINGS OF THE ASME FLUIDS ENGINEERING DIVISION SUMMER CONFERENCE, VOL 1, PTS A-C, 2009, : 555 - 565
  • [27] Evaluation of Axial Length Measurement of the Eye Using Partial Coherence Interferometry and Ultrasound in Cases of Macular Disease
    Kojima, Takashi
    Tamaoki, Akeno
    Yoshida, Norihiko
    Kaga, Tatsushi
    Suto, Chikako
    Ichikawa, Kazuo
    OPHTHALMOLOGY, 2010, 117 (09) : 1750 - 1754
  • [28] ULTRASOUND VELOCITIES FOR AXIAL EYE LENGTH MEASUREMENT
    HOFFER, KJ
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1994, 20 (05): : 554 - 562
  • [29] Ultrasound axial length measurement in biphakic eyes
    Hoffer, KJ
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2003, 29 (05): : 961 - 965
  • [30] Comparison of Intraocular Lens Calculation Formulas Measured by Immersion-Type A-Scan Ultrasound and Partial Coherence Interferometry
    Yun, Il Suk
    Park, Aram
    Choi, Shin Young
    Yang, Hong Seok
    JOURNAL OF THE KOREAN OPHTHALMOLOGICAL SOCIETY, 2016, 57 (06): : 905 - 916