Comparison of dose response functions for EBT3 model GafChromic™ film dosimetry system

被引:28
|
作者
Aldelaijan, Saad [1 ,2 ,3 ,4 ]
Devic, Slobodan [1 ,2 ,5 ]
机构
[1] Jewish Gen Hosp, Radiat Oncol Dept, Montreal, PQ H3T 1E2, Canada
[2] McGill Univ, Med Phys Unit, Montreal, PQ H4A 3J1, Canada
[3] Montreal Neurol Inst, Biol & Biomed Engn Dept, Montreal, PQ H3A 2B4, Canada
[4] King Faisal Specialist Hosp & Res Ctr, Biomed Phys Dept, Riyadh, Saudi Arabia
[5] McGill Univ, Jewish Gen Hosp, Segal Canc Ctr, Montreal, PQ, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
Radiochromic film dosimetry system; Response function; PROTOCOL;
D O I
10.1016/j.ejmp.2018.05.014
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: Different dose response functions of EBT3 model GafChromic (TM) film dosimetry system have been compared in terms of sensitivity as well as uncertainty vs. error analysis. We also made an assessment of the necessity of scanning film pieces before and after irradiation. Methods: Pieces of EBT3 film model were irradiated to different dose values in Solid Water (SW) phantom. Based on images scanned in both reflection and transmission mode before and after irradiation, twelve different response functions were calculated. For every response function, a reference radiochromic film dosimetry system was established by generating calibration curve and by performing the error vs. uncertainty analysis. Results: Response functions using pixel values from the green channel demonstrated the highest sensitivity in both transmission and reflection mode. All functions were successfully fitted with rational functional form, and provided an overall one-sigma uncertainty of better than 2% for doses above 2 Gy. Use of pre-scanned images to calculate response functions resulted in negligible improvement in dose measurement accuracy. Conclusion: Although reflection scanning mode provides higher sensitivity and could lead to a more widespread use of radiochromic film dosimetry, it has fairly limited dose range and slightly increased uncertainty when compared to transmission scan based response functions. Double-scanning technique, either in transmission or reflection mode, shows negligible improvement in dose accuracy as well as a negligible increase in dose uncertainty. Normalized pixel value of the images scanned in transmission mode shows linear response in a dose range of up to 11 Gy.
引用
收藏
页码:112 / 118
页数:7
相关论文
共 50 条
  • [31] Characterizing the response of Gafchromic EBT3 film in a 1.5 T magnetic field
    Roed, Y.
    Lee, H.
    Pinsky, L.
    Ibbott, G.
    RADIOTHERAPY AND ONCOLOGY, 2017, 123 : S403 - S403
  • [32] Sum signal dosimetry: A new approach for high dose quality assurance with Gafchromic EBT3
    Cusumano, Davide
    Fumagalli, Maria Luisa
    Ghielmetti, Francesco
    Rossi, Linda
    Grossi, Giuliano
    Lanzarotti, Raffaella
    Fariselli, Laura
    De Martin, Elena
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2017, 18 (02): : 181 - 190
  • [33] Investigating the Clinical Utility of Gafchromic EBT3 Film Dosimetry in An MR-Guided Linac
    Xhaferllari, I.
    Kim, J.
    Chetty, I.
    Liu, C.
    Du, D.
    Doemer, A.
    Glide-Hurst, C.
    Movsas, B.
    Dragovic, J.
    Wen, N.
    MEDICAL PHYSICS, 2018, 45 (06) : E380 - E380
  • [34] Gafchromic EBT3 film for absolute dosimetry in proton therapy based on averaging of beam quality
    Resch, A.
    Heyes, P.
    Fuchs, H.
    Georg, D.
    Hugo, P.
    RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S482 - S483
  • [35] Submillimeter Dose Measurement with Gafchromic EBT2 and EBT3 Films
    Corradini, N.
    Presilla, S.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [36] Energy response of the GAFCHROMIC EBT3 in diagnosis range
    Gotanda, T.
    Katsuda, T.
    Gotanda, R.
    Kuwano, T.
    Akagawa, T.
    Tanki, N.
    Tabuchi, A.
    Shimono, T.
    Kawaji, Y.
    Ideguchi, T.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, 2015, VOLS 1 AND 2, 2015, 51 : 752 - 755
  • [37] Surface dose measurement with Gafchromic EBT3 film for intensity modulated radiotherapy technique
    Akbas, Ugur
    Kesen, Nazmiye Donmez
    Koksal, Canan
    Okutan, Murat
    Demir, Bayram
    Becerir, Hatice Bilge
    3RD INTERNATIONAL CONFERENCE ON THEORETICAL AND EXPERIMENTAL STUDIES IN NUCLEAR APPLICATIONS AND TECHNOLOGY (TESNAT 2017), 2017, 154
  • [38] A diffuse reflectance spectrophotometer for radiation dosimetry of EBT3 GAFchromic films
    Piliero, Maria A.
    Pupillo, Francesco
    Presilla, Stefano
    Radiation Measurements, 2022, 154
  • [39] Comparing Homogeneity Between Gafchromic Film EBT2 and EBT3
    Mizuno, H.
    Sumida, I.
    Tanaka, A.
    Ogawa, K.
    MEDICAL PHYSICS, 2014, 41 (06) : 239 - 239
  • [40] Commissioning and evaluation of a radiochromic EBT3 film dosimetry system
    Khan, Muhammad Isa
    Tahir, Muhammad Bilal
    Rafique, Muhammad
    Iqbal, Tahir
    Zulfiqar, Sabiha
    Zahoor, Aliza
    Rehman, Jalil Ur
    Iqba, Khalid
    Chow, James
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2019, 18 (01) : 55 - 62