Determination of uranium in urine by inductively coupled plasma mass spectrometry with pneumatic nebulization

被引:33
|
作者
Caddia, M [1 ]
Iversen, BS [1 ]
机构
[1] Commiss European Communities, Joint Res Ctr, Med Serv, I-21020 Ispra, Italy
关键词
uranium; urine; inductively coupled plasma mass spectrometry; pneumatic nebulization;
D O I
10.1039/a704444e
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An analytical method using inductively coupled argon plasma mass spectrometry with pneumatic nebulization sample introduction for the determination of uranium in urine is described. The urine samples were diluted 1+19 (1% HNO3 in Milli-Q water) before analysis, Standard additions calibration with iridium (Ir-193) as an internal standard was used to compensate for undesirable matrix effects. The additions to the diluted samples were 0.1, 0.5, 1.0 and 2.0 ng l(-1), equivalent to 2, 10, 20 and 40 ng l(-1) in undiluted urine, The detection limit of the method is 0.32 ng l(-1), which is sufficiently low to measure uranium in non-exposed subjects. The precision at the 8.9 ng l(-1) level was 3.9% (RSD), and at the 28.9 ng l(-1) level 2.0%. The recovery for 20 ng l(-1) added to ten different samples was 101.5%, Urine samples from 18 nonexposed subjects were analysed. The mean concentration found was 16.1 ng l(-1) (median = 15.1 ng l(-)1) with 14.9 ng 1-' (median = 13.5 ng l(-1)) for male and 17.2 ng I-1 (median = 17.6 ng l(-1)) for female subjects.
引用
收藏
页码:309 / 313
页数:5
相关论文
共 50 条
  • [31] DIRECT INJECTION NEBULIZATION FOR INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY
    WIEDERIN, DR
    SMITH, FG
    HOUK, RS
    ANALYTICAL CHEMISTRY, 1991, 63 (03) : 219 - 225
  • [32] Determination of selenium in urine by inductively coupled plasma mass spectrometry: interferences and optimization
    Gammelgaard, B
    Jons, O
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1999, 14 (05) : 867 - 874
  • [33] COMPARISON OF PNEUMATIC NEBULIZATION AND HYDRIDE GENERATION INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY FOR ISOTOPIC ANALYSIS OF SELENIUM
    JANGHORBANI, M
    TING, BTG
    ANALYTICAL CHEMISTRY, 1989, 61 (07) : 701 - 708
  • [34] The determination of mercury in whole blood and urine by inductively coupled plasma mass spectrometry
    Nixon, DE
    Burritt, MF
    Moyer, TP
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1999, 54 (08) : 1141 - 1153
  • [35] Determination of trace levels of elements in urine by inductively coupled plasma mass spectrometry
    Nakagawa, J
    Tsuchiya, Y
    Yashima, Y
    Tezuka, M
    Fujimoto, Y
    JOURNAL OF HEALTH SCIENCE, 2004, 50 (02) : 164 - 168
  • [36] Determination of Toxic and Essential Elements in Urine by Inductively Coupled Plasma Mass Spectrometry
    Veikhman, G. A.
    Gilyova, O. V.
    Stenno, E. V.
    Ulanova, T. S.
    SOVREMENNYE TEHNOLOGII V MEDICINE, 2016, 8 (03) : 120 - 124
  • [37] Determination of mercury in whole blood and urine by inductively coupled plasma mass spectrometry
    Fong, Bonnie Mei Wah
    Siu, Tak Shing
    Lee, Joseph Sai Kit
    Tam, Sidney
    JOURNAL OF ANALYTICAL TOXICOLOGY, 2007, 31 (05) : 281 - 287
  • [38] Matrix elimination for the determination of uranium in seawater using inductively coupled plasma mass spectrometry
    Wood, Jordana
    Gill, Gary
    Choe, Key-Young
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2015, 249
  • [39] A fast method for the determination of uranium in river water with inductively coupled plasma mass spectrometry
    Li, YF
    Tian, LD
    Li, FX
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2002, 30 (11) : 1400 - 1400
  • [40] Uranium quantification in semen by inductively coupled plasma mass spectrometry
    Todorov, Todor I.
    Ejnik, John W.
    Guandalini, Gustavo
    Xu, Hanna
    Hoover, Dennis
    Anderson, Larry
    Squibb, Katherine
    McDiarmid, Melissa A.
    Centeno, Jose A.
    JOURNAL OF TRACE ELEMENTS IN MEDICINE AND BIOLOGY, 2013, 27 (01) : 2 - 6