Long-term testing of an intracranial pressure monitoring device

被引:16
|
作者
Kroin, JS
McCarthy, RJ
Stylos, L
Miesel, K
Ivankovich, AD
Penn, RD
机构
[1] Rush Med Coll, Dept Neurosurg, Chicago, IL 60612 USA
[2] Rush Med Coll, Dept Anesthesiol, Chicago, IL 60612 USA
[3] Mt Sinai Sch Med, Dept Neurosurg, New York, NY USA
[4] Medtron Inc, Minneapolis, MN USA
关键词
cerebrospinal fluid; intracranial pressure; head injury; hydrocephalus; dog;
D O I
10.3171/jns.2000.93.5.0852
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Object. Long-term monitoring of intracranial pressure (ICP) is limited by the lack of an implantable sensor with low drift. The goal of this study was to demonstrate that a new capacitive transducer system will produce accurate and stable ICP records over extended periods. Methods. Intracranial pressure sensors were implanted into the frontal white matter of four dogs. Tn addition, a fluid-filled catheter was placed in the cisterna magna (CM) to measure cerebrospinal fluid (CSF) pressure. The animals were tested using standard physiological maneuvers such as jugular vein compression, head elevation, and CSF withdrawal from and saline injection into the CM to verify that the ICP sensor precisely matched CSF pressure changes. The mean ICP pressure and CM pressure were compared for months to demonstrate that the transducer system produced minimal drift over time. The change in the ICP sensor record closely duplicated that of the CSF waveform in the CM in response to well-known physiological stimuli. More important, mean ICP pressure remained within 3 mm Hg of CM pressure for months, with a mean difference of less than 0.3 mm Hg, Histological examination of the dog brains revealed only minimal tissue reaction to the presence of the sensor. Conclusions. The authors demonstrate a new implantable solid-state sensor that reliably measures ICP for months, with minimal drift. The clinical application of this sensor and its telemetry is for long-term monitoring of patients with head injury, mass lesions, and hydrocephalus.
引用
收藏
页码:852 / 858
页数:7
相关论文
共 50 条
  • [31] Long-term monitoring of intracranial pressure in freely-moving rats using telemetry system
    Eftekhari, Sajedeh
    Johansen, Katrine P.
    Bruun, Signe Rath
    Jensen, Rigmor H.
    CEPHALALGIA, 2019, 39 : 143 - 143
  • [32] Non-Surgical, In-Stent Membrane Bioelectronics for Long-Term Intracranial Pressure Monitoring
    Lee, Jimin
    Bateman, Allison
    Kim, Mi Hyeon
    Rigo, Bruno
    Kim, Hodam
    Lee, Jaeho
    Choi, Yun Hyeok
    Herbert, Robert
    Lee, Deok Hee
    Yeo, Woon-Hong
    ADVANCED HEALTHCARE MATERIALS, 2025,
  • [33] LONG-TERM BLOOD-PRESSURE MONITORING
    ZHAROV, EI
    VERTKIN, AL
    SOLOVYEV, VV
    PROKHOROVICH, EA
    MARTYNOV, DA
    KARDIOLOGIYA, 1990, 30 (06) : 118 - 122
  • [34] A METHOD OF PRODUCING A LOADING DEVICE FOR LONG-TERM TESTING
    BAKIROV, MY
    SHUKYUROV, YG
    ISMAILOVA, RS
    INDUSTRIAL LABORATORY, 1985, 51 (09): : 854 - 855
  • [35] DEVICE FOR LONG-TERM DURABILITY TESTING OF TRANSMISSION OILS
    Peterka, Bohuslav
    TRENDS IN AGRICULTURAL ENGINEERING 2010, 2010, : 473 - 475
  • [36] Vibration Device for Long-Term Testing of Car Seats
    Srb, P.
    Fliegel, V.
    MODERN METHODS OF CONSTRUCTION DESIGN, 2014, : 189 - 195
  • [37] INTEGRATED IMPLANTABLE DEVICE FOR LONG-TERM GLUCOSE MONITORING
    WILKINS, E
    ATANASOV, P
    MUGGENBURG, BA
    BIOSENSORS & BIOELECTRONICS, 1995, 10 (05): : 485 - 494
  • [38] A Wearable Device for Online and Long-Term ECG Monitoring
    Longoni, Marco
    Carrera, Diego
    Rossi, Beatrice
    Fragneto, Pasqualina
    Pessione, Marco
    Boracchi, Giacomo
    PROCEEDINGS OF THE TWENTY-SEVENTH INTERNATIONAL JOINT CONFERENCE ON ARTIFICIAL INTELLIGENCE, 2018, : 5838 - 5840
  • [39] Integrated implantable device for long-term glucose monitoring
    Dept. of Chem./Nuclear Engineering, University of New Mexico, Albuquerque, NM 87131-1341, United States
    BIOSENS. BIOELECTRON., 5 (485-494):
  • [40] A PROPOSAL OF A NOVEL CARDIORESPIRATORY LONG-TERM MONITORING DEVICE
    Lapi, S.
    Biagi, E.
    Borgioli, G.
    Calzolai, M.
    Masotti, L.
    Fontana, G.
    BIOSIGNALS 2011, 2011, : 38 - +