Contact mechanics and elastohydrodynamic lubrication in a novel metal-on-metal hip implant with an aspherical bearing surface

被引:22
|
作者
Meng, Qingen [1 ]
Gao, Leiming [1 ]
Liu, Feng [1 ]
Yang, Peiran [2 ]
Fisher, John [1 ]
Jin, Zhongmin [1 ]
机构
[1] Univ Leeds, Sch Mech Engn, Inst Med & Biol Engn, Leeds LS2 9JT, W Yorkshire, England
[2] Qingdao Technol Univ, Sch Mech Engn, Qingdao, Peoples R China
基金
中国国家自然科学基金;
关键词
Contact mechanics; Elastohydrodynamic lubrication; Metal-on-metal hip implants; Aspherical bearing surfaces; WEAR PARTICLES; RESURFACING CUPS; ION RELEASE; IN-VITRO; DEFORMATION; JOINTS; PREDICTION; PROSTHESES; CLEARANCE; DIAMETER;
D O I
10.1016/j.jbiomech.2009.11.018
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Diameter and diametral clearance of the bearing surfaces of metal-on-metal hip implants and structural supports have been recognised as key factors to reduce the dry contact and hydrodynamic pressures and improve lubrication performance. On the other hand, application of aspherical bearing surfaces can also significantly affect the contact mechanics and lubrication performance by changing the radius of the curvature of a bearing surface and consequently improving the conformity between the head and the cup. In this study, a novel metal-on-metal hip implant employing a specific aspherical bearing surface, Alpharabola, as the acetabular surface was investigated for both contact mechanics and elastohydrodynamic lubrication under steady-state conditions. When compared with conventional spherical bearing surfaces, a more uniform pressure distribution and a thicker lubricant film thickness within the loaded conjunction were predicted for this novel Alpharabola hip implant. The effects of the geometric parameters of this novel acetabular surface on the pressure distribution and lubricant thickness were investigated. A significant increase in the predicted lubricant film thickness and a significant decrease in the dry contact and hydrodynamic pressures were found with appropriate combinations of these geometric parameters, compared with the spherical bearing surface. (C) 2009 Elsevier Ltd. All rights reserved.
引用
收藏
页码:849 / 857
页数:9
相关论文
共 50 条
  • [41] The Otto Aufranc Award - Wear and lubrication of metal-on-metal hip implants
    Chan, FW
    Bobyn, JD
    Medley, JB
    Krygier, JJ
    Tanzer, M
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1999, (369) : 10 - 24
  • [42] In Silico Contact Pressure of Metal-on-Metal Total Hip Implant with Different Materials Subjected to Gait Loading
    Jamari, J.
    Ammarullah, Muhammad Imam
    Santoso, Gatot
    Sugiharto, S.
    Supriyono, Toto
    van der Heide, Emile
    METALS, 2022, 12 (08)
  • [43] Outcomes of the Recalled Articular Surface Replacement Metal-on-Metal Hip Implant System: A Systematic Review
    Laaksonen, Inari
    Donahue, Gabrielle S.
    Madanat, Rami
    Makela, Keijo T.
    Malchau, Henrik
    JOURNAL OF ARTHROPLASTY, 2017, 32 (01): : 341 - 346
  • [44] The effect of cup orientation and coverage on contact mechanics and range of motion of metal-on-metal hip resurfacing arthroplasty
    Wang, Ling
    Williams, Sophie
    Udofia, Itoro
    Isaac, Graham
    Fisher, John
    Jin, Zhongmin
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2012, 226 (H11) : 877 - 886
  • [45] Comparison of numerical methods for elastohydrodynamic lubrication analysis of metal-on-metal hip implants: multi-grid verses Newton-Raphson
    Gao, L. M.
    Meng, Q. E.
    Wang, F. C.
    Yang, P. R.
    Jin, Z. M.
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART J-JOURNAL OF ENGINEERING TRIBOLOGY, 2007, 221 (J2) : 133 - 140
  • [46] Wear Simulation of Metal-on-Metal Hip Replacements With Frictional Contact
    Mattei, Lorenza
    Di Puccio, Francesca
    JOURNAL OF TRIBOLOGY-TRANSACTIONS OF THE ASME, 2013, 135 (02):
  • [47] Contact mechanics and lubrication analyses of ceramic-on-metal total hip replacements
    Meng, Qingen
    Liu, Feng
    Fisher, John
    Jin, Zhongmin
    TRIBOLOGY INTERNATIONAL, 2013, 63 : 51 - 60
  • [48] Wear of surface engineered metal-on-metal hip prostheses
    J. Fisher
    X. Q. Hu
    T. D. Stewart
    S. Williams
    J. L. Tipper
    E. Ingham
    M. H. Stone
    C. Davies
    P. Hatto
    J. Bolton
    M. Riley
    C. Hardaker
    G. H. Isaac
    G. Berry
    Journal of Materials Science: Materials in Medicine, 2004, 15 : 225 - 235
  • [49] Survivorship of a Metal-on-Metal Total Hip Implant With Modular Titanium Adapter
    Semaan, Derek J.
    Rutledge-Jukes, Heath
    Berend, Keith R.
    Lombardi, Adolph V., Jr.
    Adams, Joanne B.
    Crawford, David A.
    JOURNAL OF ARTHROPLASTY, 2022, 37 (07): : S560 - S565
  • [50] Wear of surface engineered metal-on-metal hip prostheses
    Fisher, J
    Hu, XQ
    Stewart, TD
    Williams, S
    Tipper, JL
    Ingham, E
    Stone, MH
    Davies, C
    Hatto, P
    Bolton, J
    Riley, M
    Hardaker, C
    Isaac, GH
    Berry, G
    JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE, 2004, 15 (03) : 225 - 235