Optical Microfiber Reader for Enzyme-Linked Immunosorbent Assay

被引:38
|
作者
Sun, Li-Peng [1 ]
Huang, Yan [1 ]
Huang, Tiansheng [1 ]
Yuan, Zihao [1 ]
Lin, Wenfu [1 ]
Sun, Zhen [1 ]
Yang, Mingjin [1 ]
Xiao, Peng [1 ]
Ma, Jun [1 ]
Wang, Wei [1 ]
Zhang, Yi [2 ]
Liu, Zonghua [2 ]
Guan, Bai-Ou [1 ]
机构
[1] Jinan Univ, Inst Photon Technol, Guangdong Prov Key Lab Opt Fiber Sensing & Commun, Guangzhou 511443, Guangdong, Peoples R China
[2] Jinan Univ, Dept Biomed Engn, Guangzhou 510632, Guangdong, Peoples R China
基金
中国国家自然科学基金;
关键词
C-REACTIVE PROTEIN; COLLOIDAL GOLD IMMUNOASSAY; ULTRASENSITIVE DETECTION; MULTIPLEXED DETECTION; ELISA; BIOSENSORS; SERUM; SENSITIVITY; BIOMARKERS; DISEASE;
D O I
10.1021/acs.analchem.9b04119
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In clinical diagnosis, accurate and reliable measurement technologies for the detection of disease biomarkers at ultralow concentrations can provide guidance for the initiation of treatment and potentially improve survival for patients. Here, we demonstrate an optical microfiber reader for enhanced analytical sensitivity in enzyme-linked immunosorbent assays (ELISA) that enables the detection of tiny changes of the refractive index (RI) induced by the catalyzed oxidation of substrate, owing to the strong interaction between the evanescent field and surrounding medium. By employing the microfiber reader for the C-reaction protein (CRP) and interleukin-6 (IL-6) assays after the enzymatic signal amplification in ELISA, we experimentally investigate the biosensing capacity of the device. As a result, log linear relations of CRP and IL-6 detection in PBS and human serum between the concentration and spectral response were obtained at both nanogram and picogram levels, respectively, and anti-CRP/HRP detection as low as 9.75 pg/mL was achieved, which was undetectable by the conventional spectrophotometry. With a stable, accurate, and color-free detection capacity, this optical microfiber reader has a promising prospect in early disease diagnosis and clinical treatment.
引用
收藏
页码:14141 / 14148
页数:8
相关论文
共 50 条
  • [31] Development of an enzyme-linked immunosorbent assay for octylphenol
    Li, Zhuokun
    Liu, Liqiang
    Song, Shanshan
    Guo, Shidong
    Kuang, Hua
    Xu, Chuanlai
    FOOD AND AGRICULTURAL IMMUNOLOGY, 2014, 25 (03) : 397 - 410
  • [32] Enzyme-Linked Immunosorbent Assay of Chlorampenicol in Foodstuff
    Fedorova, M. D.
    Andreeva, I. P.
    Vilegzhanina, E. S.
    Komarov, A. A.
    Rubtsova, M. Yu.
    Samsonova, J. V.
    Egorov, A. M.
    APPLIED BIOCHEMISTRY AND MICROBIOLOGY, 2010, 46 (08) : 795 - 801
  • [33] Evaluation of musk by enzyme-linked immunosorbent assay
    Ahn, KS
    Hahn, BS
    Lee, JP
    Chang, SY
    Lee, HK
    Jeong, CS
    Kim, YS
    BIOLOGICAL & PHARMACEUTICAL BULLETIN, 2002, 25 (04) : 418 - 421
  • [34] Sandwich enzyme-linked immunosorbent assay for naringin
    Qu, Huihua
    Wang, Xueqian
    Qu, Baoping
    Kong, Hui
    Zhang, Yue
    Shan, Wenchao
    Cheng, Jinjun
    Wang, Qingguo
    Zhao, Yan
    ANALYTICA CHIMICA ACTA, 2016, 903 : 149 - 155
  • [35] Enzyme-linked immunosorbent assay and immunobullous disease
    Heymann, Warren R.
    Smith, Kathleen J.
    Brown, Stuart
    JOURNAL OF THE AMERICAN ACADEMY OF DERMATOLOGY, 2009, 60 (04) : 676 - 678
  • [36] Development of an enzyme-linked immunosorbent assay for podophyllotoxin
    Xu, DM
    Yu, XY
    Liu, YQ
    Feng, JT
    Pan, LG
    Liu, XJ
    He, J
    Zhang, X
    INTERNATIONAL IMMUNOPHARMACOLOGY, 2005, 5 (11) : 1583 - 1592
  • [37] Cascade enzyme-linked immunosorbent assay (CELISA)
    Lee, Young-mi
    Jeong, Yujin
    Kang, Hyo Jin
    Chung, Sang J.
    Chung, Bong Hyun
    BIOSENSORS & BIOELECTRONICS, 2009, 25 (02): : 332 - 337
  • [38] An enzyme-linked immunosorbent assay for the detection of bifenthrin
    Li Bo
    Shi Hai-Yan
    Wang Ming-Hua
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2008, 36 (01) : 34 - 38
  • [39] An Enzyme-linked Immunosorbent Assay for Detection of Imidacloprid
    Peng Fang-Yi
    Jiang Hai-Rong
    Chen Yuan-Xiang
    Chen Sheng-Zhen
    Lin Zhi-Hua
    Liao Pu
    He Miao
    Shi Han-Chang
    Cai Qiang
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2010, 38 (12) : 1737 - 1741
  • [40] Enzyme-linked immunosorbent assay for the pyrethroid permethrin
    Shan, GM
    Leeman, WR
    Stoutamire, DW
    Gee, SJ
    Chang, DPY
    Hammock, BD
    JOURNAL OF AGRICULTURAL AND FOOD CHEMISTRY, 2000, 48 (09) : 4032 - 4040