Subpectoral Implantation of Internal Pulse Generators for Deep Brain Stimulation: Technical Note for Improved Cosmetic Outcomes

被引:7
|
作者
White-Dzuro, Gabrielle A. [1 ]
Lake, Wendell [2 ]
Neimat, Joseph S. [3 ]
机构
[1] Univ Washington, Dept Neurol Surg, 325 9th Ave Box 359924, Seattle, WA 98104 USA
[2] Univ Wisconsin Madison, Dept Neurol Surg, Madison, WI USA
[3] Vanderbilt Univ, Sch Med, Dept Neurol Surg, Nashville, TN 37235 USA
关键词
Deep brain stimulation; Essential tremor; Internal pulse generator; Parkinson's disease; Subpectoral; CARDIOVERTER-DEFIBRILLATORS; PACEMAKERS; POCKET; SITE;
D O I
10.1093/ons/opx018
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
BACKGROUND: Deep brain stimulation is increasingly used to treat a variety of disorders. As the prevalence of this technology increases, greater demands are placed on neurosurgical practitioners to improve cosmetic results, maximize patient comfort, and minimize complication rates. We have increasingly employed subpectoral implantation of internal pulse generators (IPGs) to improve patient satisfaction. OBJECTIVE: To determine the complication rates of subpectorally placed IPGs as compared to those placed in a subcutaneous location. METHODS: We reviewed a series of 301 patients from a single institution. Complication rates including infection, hematoma, and lead fracture were recorded. Rates were compared for subcutaneously and subpectorally located devices. RESULTS: Of the records reviewed, we found 301 patients who underwent 308 procedures for initial IPG implantation. Of these, 275 were subpectoral IPG implantation, 19 were infraclavicular subcutaneous implantation, and 14 were subcutaneous implantation in the abdomen. A total of 6 IPG pocket infections occurred, 2 subpectoral and 4 infraclavicular subcutaneous. Of the IPG infections, 2 of the infraclavicular subcutaneous devices had associated erosions. Two patients had their devices relocated from a subpectoral pocket to a subcutaneous pocket in the abdomen due to discomfort. Two patients in the subpectoral group suffered from hematoma requiring evacuation. Two patients in the infraclavicular subcutaneous group had lead fracture occur. CONCLUSIONS: Subpectoral implantation of deep brain stimulation IPGs is a viable alternative with a low complication rate. This techniquemay offer a lower rate of infection and wound erosion.
引用
收藏
页码:529 / 534
页数:6
相关论文
共 50 条
  • [21] Extensor bar for deep brain stimulation electrode insertion: Technical note
    Oliveira, J. O., Jr.
    Listik, S.
    Machado, A.
    Walter, A.
    Barata, S.
    Cukiert, A.
    MOVEMENT DISORDERS, 2010, 25 (07) : S441 - S442
  • [22] Cosmetic considerations for placement of deep brain stimulation pulse generator: the submammary subfascial approach
    Jason Yuen
    Khaled Alameddine
    Eugene S. Bah
    Kendall H. Lee
    Basel A. Sharaf
    Acta Neurochirurgica, 2023, 165 : 735 - 739
  • [23] Cosmetic considerations for placement of deep brain stimulation pulse generator: the submammary subfascial approach
    Yuen, Jason
    Alameddine, Khaled
    Bah, Eugene S.
    Lee, Kendall H.
    Sharaf, Basel A.
    ACTA NEUROCHIRURGICA, 2023, 165 (03) : 735 - 739
  • [24] Implantable Pulse Generators for Deep Brain Stimulation: Challenges, Complications, and Strategies for Practicality and Longevity
    Sarica, Can
    Iorio-Morin, Christian
    Aguirre-Padilla, David H.
    Najjar, Ahmed
    Paff, Michelle
    Fomenko, Anton
    Yamamoto, Kazuaki
    Zemmar, Ajmal
    Lipsman, Nir
    Ibrahim, George M.
    Hamani, Clement
    Hodaie, Mojgan
    Lozano, Andres M.
    Munhoz, Renato P.
    Fasano, Alfonso
    Kalia, Suneil K.
    FRONTIERS IN HUMAN NEUROSCIENCE, 2021, 15
  • [25] Impact of two different rechargeable pulse generators on parkinsonian patients with deep brain stimulation
    Bogaert, C.
    der Linden, C.
    MOVEMENT DISORDERS, 2022, 37 : S136 - S136
  • [26] Risk of tissue damage and deep brain stimulation with external devices: A technical note
    Trottenberg, T
    Winter, C
    Alesch, F
    Kupsch, A
    ANNALS OF NEUROLOGY, 2004, 56 (02) : 310 - 311
  • [27] Meta-analysis and review of rechargeable implantable pulse generators for spinal cord stimulation and deep brain stimulation
    Ozgan, Denizhan
    Krieg, Sandro M.
    Jakobs, Martin
    BRAIN AND SPINE, 2024, 4
  • [28] Recharge PSYCH A Study on Rechargeable Implantable Pulse Generators in Deep Brain Stimulation for Psychiatric Disorders
    Jakobs, Martin
    Hajiabadi, Mohammad Mehdi
    Agiorre-Padilla, David Hernan
    Giaccobe, Peter
    Unterberg, Andreas W.
    Lozano, Andres M.
    WORLD NEUROSURGERY, 2023, 170 : E331 - E339
  • [29] Management of Pulse Generators in a Breast Cancer Patient with in Situ Subthalamic Nucleus Deep Brain Stimulation
    Son, Byung-chul
    Kim, Joong-Seok
    Park, Woo-Chan
    Ko, Hak-cheol
    JOURNAL OF NEUROLOGICAL SURGERY PART A-CENTRAL EUROPEAN NEUROSURGERY, 2019, 80 (03) : 223 - 227
  • [30] Lead Cap Localization using Ultrasound in Deep Brain Stimulation Surgery: Technical Note
    Englot, D. J.
    Chang, E. F.
    Larson, P. S.
    MINIMALLY INVASIVE NEUROSURGERY, 2011, 54 (01) : 48 - 49