High Stress Conditions Do Not Increase Wear of Thin Highly Crosslinked UHMWPE

被引:29
|
作者
Kelly, Natalie H. [1 ]
Rajadhyaksha, Amar D. [2 ]
Wright, Timothy M. [1 ]
Maher, Suzanne A. [1 ]
Westrich, Geoffrey H. [2 ]
机构
[1] Hosp Special Surg, Dept Biomech, New York, NY 10021 USA
[2] Hosp Special Surg, Hip & Knee Serv, New York, NY 10021 USA
关键词
TOTAL HIP-ARTHROPLASTY; MOLECULAR-WEIGHT POLYETHYLENE; FEMORAL-HEAD; CONVENTIONAL POLYETHYLENE; ACETABULAR COMPONENTS; MECHANICAL-PROPERTIES; IMPINGEMENT; LINERS; DISLOCATION; REPLACEMENT;
D O I
10.1007/s11999-009-1154-6
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Introduction of highly crosslinked polyethylene has increased interest in large femoral heads, because thin acetabular liners can be used while maintaining low wear rates and larger heads decrease the incidence of instability. However, crosslinking and subsequent thermal treatments can cause decreased mechanical properties that might obviate the reduced wear under extreme conditions. To examine whether increased contact pressures would adversely affect wear in thin liners, we tested thin and thick highly crosslinked liners (3.8 mm thickness/44-mm head and 7.9 mm thickness/36-mm head, respectively) to 5 million cycles on a hip simulator under near impingement conditions. Conventional polyethylene liners (7.9 mm thickness/36-mm head) served as controls. Large femoral heads with highly crosslinked polyethylene liners as thin as 3.8 mm in thickness do not wear at a higher rate than a thicker liner of the same material, even when subjected to large contact pressures such as occur under near-impingement conditions. Crosslinked polyethylene may allow for liners that are thinner than has been traditionally accepted. This conclusion, however, is based solely on wear test results with idealized cup position, no intentional edge loading, no head subluxation, and no artificial aging. Continued monitoring will be necessary to elucidate the clinical efficacy of these devices.
引用
收藏
页码:418 / 423
页数:6
相关论文
共 50 条
  • [1] Enhanced wear performance of highly crosslinked UHMWPE for artificial joints
    Chiesa, R
    Tanzi, MC
    Alfonsi, S
    Paracchini, L
    Moscatelli, M
    Cigada, A
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 2000, 50 (03): : 381 - 387
  • [2] Unified wear model for highly crosslinked ultra-high molecular weight polyethylenes (UHMWPE)
    Muratoglu, OK
    Bragdon, CR
    O'Connor, DO
    Jasty, M
    Harris, WH
    Gul, R
    McGarry, F
    BIOMATERIALS, 1999, 20 (16) : 1463 - 1470
  • [3] Retrieved Highly Crosslinked UHMWPE Acetabular Liners Have Similar Wear Damage as Conventional UHMWPE
    Schroder, David T.
    Kelly, Natalie H.
    Wright, Timothy M.
    Parks, Michael L.
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2011, 469 (02) : 387 - 394
  • [4] Do Tissues From THA Revision of Highly Crosslinked UHMWPE Liners Contain Wear Debris and Associated Inflammation?
    Baxter, Ryan M.
    Freeman, Theresa A.
    Kurtz, Steven M.
    Steinbeck, Marla J.
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2011, 469 (08) : 2308 - 2317
  • [5] High crystallinity makes excellent wear resistance in crosslinked UHMWPE
    Wang, Zirui
    Li, Quan
    Chen, Xianchun
    Zhang, Qin
    Wang, Ke
    POLYMER, 2023, 280
  • [6] Natural polyphenol-stabilised highly crosslinked UHMWPE with high mechanical properties and low wear for joint implants
    Fu, Jun
    Shen, Jie
    Gao, Guorong
    Xu, Yuhao
    Hou, Ruixia
    Cong, Yang
    Cheng, Yajun
    JOURNAL OF MATERIALS CHEMISTRY B, 2013, 1 (37) : 4727 - 4735
  • [7] The wear of highly crosslinked UHMWPE in the presence of abrasive particles: hip and knee simulator studies
    Laurent, MP
    Blanchard, CR
    Yao, JQ
    Johnson, TS
    Gilbertson, LN
    Swarts, DF
    Crowninshield, RD
    CROSSLINKED AND THERMALLLY TREATED ULTRA-HIGH MOLECULAR WEIGHT POLYETHYLENE FOR JOINT REPLACEMENTS, 2004, 1445 : 86 - 103
  • [8] The influence of stress conditions on the wear of UHMWPE for total joint replacements
    Barbour, PSM
    Barton, DC
    Fisher, J
    JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE, 1997, 8 (10) : 603 - 611
  • [9] The influence of stress conditions on the wear of UHMWPE for total joint replacements
    P. S. M BARBOUR
    D. C BARTON
    J FISHER
    Journal of Materials Science: Materials in Medicine, 1997, 8 : 603 - 611
  • [10] Effects of simulated oxidation on the in vitro wear and mechanical properties of irradiated and melted highly crosslinked UHMWPE
    Oral, Ebru
    Neils, Andrew L.
    Doshi, Brinda N.
    Fu, Jun
    Muratoglu, Orhun K.
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2016, 104 (02) : 316 - 322