Thomson's multitaper approach combined with coherent plane-wave compounding to reduce speckle in ultrasound imaging

被引:15
|
作者
Toulemonde, Matthieu [1 ,2 ]
Basset, Olivier [1 ]
Tortoli, Piero [2 ]
Cachard, Christian [1 ]
机构
[1] Univ Lyon 1, INSA Lyon, CREATIS, CNRS,Inserm,UMR 5220,U630, Villeurbanne, France
[2] Univ Florence, Dept Informat Engn, Microelect Syst Design Lab, I-50121 Florence, Italy
关键词
Ultrasound imaging; Speckle; Thomson's multitaper; Coherent plane-wave compounding; FRAME RATE ULTRASONOGRAPHY; BREAST;
D O I
10.1016/j.ultras.2014.09.006
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
In ultrasound imaging, the speckle pattern limits the image quality. Spatial and frequency compounding are commonly used to reduce speckle noise or improve the contrast. Although recent implementations can preserve a frame rate that is compatible with real-time imaging (e. g., synthetic aperture compounding), most classic compounding techniques are based on the coherent combination of several radiofrequency images from the same investigated area, which reduces the frame rate. Furthermore, Thomson's multitaper approach aims to smooth the speckle by incoherently combining the obtained B-mode images after applying different apodization windows on the same original data. With only one acquisition, the frame rate remains high, but the spatial resolution is decreased. To improve the resolution and contrast while reducing the speckle noise, this paper proposes combining the coherent plane-wave compounding technique (CPWC) with Thomson's multitaper method. The resulting multitaper coherent plane-wave compounding (MCPWC) takes advantage of coherent and incoherent approaches. Simulations and experimental results demonstrate that in terms of the signal-to-noise ratio, contrast, and resolution, the image quality is increased using plane wave emissions at approximately ten steering angles with three Thomson's tapers. Outside the focal area, the lateral resolution is improved by a factor of 2, and the contrast is increased by approximately 2 dB compared with images obtained using a single focalization technique and Thomson's multitaper approach. (C) 2014 Elsevier B.V. All rights reserved.
引用
收藏
页码:390 / 398
页数:9
相关论文
共 50 条
  • [21] Retrospective Transmit Beamforming and Coherent Plane-Wave Compounding for Microvascular Doppler Imaging: A Comparison Study
    Golfetto, Cristiana
    Ekroll, Ingvild Kinn
    Torp, Hans
    Lovstakken, Lasse
    Avdal, Jorgen
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2021, 68 (04) : 1105 - 1116
  • [22] Coherent Plane Wave Compounding Combined With Tensor Completion Applied for Ultrafast Imaging
    Afrakhteh, Sajjad
    Behnam, Hamid
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2021, 68 (10) : 3094 - 3103
  • [23] Ultrasound Coherent Compound Plane-Wave Imaging of Ocular Anatomy and Blood Flow
    Urs, Raksha
    Ketterling, Jeffrey A.
    Gross, Daniel
    Silverman, Ronald H.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2016, 57 (12)
  • [24] Compound Coherent Plane-Wave Ultrasound (CCPU) Imaging of Vascular Malformations in the Orbit
    Campbell, Ashley
    Callahan, Alison
    Urs, Raksha
    Silverman, Ronald H.
    Kazim, Michael
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2017, 58 (08)
  • [25] Mean-Squared Error Beamforming for Coherent Plane-Wave Compounding
    Nguyen, Nghia Q.
    Prager, Richard W.
    2019 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2019, : 316 - 319
  • [26] SPECTRA OF ULTRASOUND DOPPLER RESPONSE USING PLANE-WAVE COMPOUNDING TECHNIQUE
    Barannik, Evgen A.
    Hrytsenko, Mykhailo O.
    EAST EUROPEAN JOURNAL OF PHYSICS, 2024, (01): : 476 - 484
  • [27] Plane-Wave Imaging Challenge in Medical Ultrasound
    Liebgott, H.
    Rodriguez-Molares, A.
    Cervenansky, F.
    Jensen, J. A.
    Bernard, O.
    2016 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2016,
  • [28] Experimental Implementation of a Pulse Compression Technique Using Coherent Plane-Wave Compounding
    Benane, Yanis Mehdi
    Bujoreanu, Denis
    Lavarello, Roberto J.
    Varray, Francois
    Escoffre, Jean-Michel
    Novell, Anthony
    Cachard, Christian
    Basset, Olivier
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2018, 65 (06) : 1025 - 1036
  • [29] Dynamic coherence factor based on the standard deviation for coherent plane-wave compounding
    Wang, Yadan
    Zheng, Chichao
    Peng, Hu
    COMPUTERS IN BIOLOGY AND MEDICINE, 2019, 108 : 249 - 262
  • [30] Adaptive Beamforming in Ultrasound Imaging with Plane Wave Compounding
    Zhao, Jinxin
    Zeng, Xing
    Yu, Jinhua
    Wang, Yuanyuan
    2014 INTERNATIONAL CONFERENCE ON AUDIO, LANGUAGE AND IMAGE PROCESSING (ICALIP), VOLS 1-2, 2014, : 53 - 57