Pulse Wave Velocity Measurement in the Carotid Artery Using an LED-LED Array Pulse Oximeter

被引:1
|
作者
Campbell, Jake D. [1 ]
Holder-Pearson, Lui [1 ]
Pretty, Christopher G. [1 ]
Bones, Phil [2 ]
Chase, J. Geoffrey [1 ]
机构
[1] Univ Canterbury, Mech Engn Dept, Christchurch, New Zealand
[2] Univ Canterbury, Elect Engn Dept, Christchurch, New Zealand
来源
IFAC PAPERSONLINE | 2020年 / 53卷 / 02期
基金
欧盟地平线“2020”;
关键词
Pulse wave velocity; non-invasive; photoplethysmography; LED-LED detection; carotid artery; STIFFNESS;
D O I
10.1016/j.ifacol.2020.12.406
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
Pulse wave velocity (PWV) is frequently used as an early indicator of risk of cardiovascular disease. Conventional methods of PWV measurement are invasive and measure the regional PWV, introducing errors from unknown measurement distance to masking local changes in compliance. This paper describes the development and testing of a non-invasive PWV sensor using photoplethysmograph signals. The sensor measures the pulse in the carotid artery with three sensor arrays spaced at 20 mm, 30 mm and 50 mm spacing. Each array of 20 LED-LED sensors are placed at 5 mm to get the largest amplitude pulse across the neck, and to allow for inaccurate sensor placement. LEDs are used as light emitters and the inherent capacitance of reverse biased LEDs measure the reflected light. The foot-foot and phase difference methods were used to calculate the PWV at each measurement distance. The foot-foot method was more reliable than the phase difference at all distances with a PWV of 5.26 ms(-1) in a single-subject trial. The sample rate of 570 Hz was deemed too slow as one sample difference resulted in a PWV change of 1.5 m s(-1). The developed sensor measured the local PWV within the expected physiological range around 6 m s(-1). All future measurements will be measured at 1 kHz and an increased LED output intensity. Copyright (C) 2020 The Authors.
引用
收藏
页码:16031 / 16036
页数:6
相关论文
共 50 条
  • [1] Pulse Wave Velocity in the Carotid Artery
    Sorensen, Gertrud Laura
    Jensen, Julie Brinck
    Udesen, Jesper
    Holfort, Iben Kraglund
    Jensen, Jorgen Arendt
    2008 IEEE ULTRASONICS SYMPOSIUM, VOLS 1-4 AND APPENDIX, 2008, : 1386 - +
  • [2] Measurement of local pulse wave velocity for carotid artery by using an ultrasound-based method
    Tang, Chieh-Ju
    Lee, Po-Yang
    Chuang, Yi-Hsiang
    Huang, Chih-Chung
    ULTRASONICS, 2020, 102
  • [3] THE EFFECT OF VARYING LED INTENSITY ON PULSE OXIMETER ACCURACY
    DEKOCK, JP
    REYNOLDS, KJ
    TARASSENKO, L
    MOYLE, JTB
    JOURNAL OF MEDICAL ENGINEERING & TECHNOLOGY, 1991, 15 (03) : 111 - 116
  • [4] Assessment of ultrasound pulsation loops for measurement of local pulse wave velocity of carotid artery
    ZHAO Weijia
    ZHANG Yufeng
    LI Zhiyao
    CHEN Jianhua
    GAO Lian
    ZHANG Junhua
    ChineseJournalofAcoustics, 2021, 40 (02) : 253 - 265
  • [5] Assessment of ultrasound pulsation loops for measurement of local pulse wave velocity of carotid artery
    Zhao, Weijia
    Zhang, Yufeng
    Li, Zhiyao
    Chen, Jianhua
    Gao, Lian
    Zhang, Junhua
    Shengxue Xuebao/Acta Acustica, 2019, 44 (06): : 1045 - 1052
  • [6] PULSE WAVE VELOCITY AND TRANSMISSION AT THE CAROTID ARTERY AND THE ASCENDING AORTA
    Smith, Sara M.
    Marin, Justine
    Adams, Amari
    West, Keith
    Hao, Zhili
    PROCEEDINGS OF ASME 2021 INTERNATIONAL MECHANICAL ENGINEERING CONGRESS AND EXPOSITION (IMECE2021), VOL 5, 2021,
  • [7] Measurement of regional pulse-wave velocity using spatial compound imaging of the common carotid artery in vivo
    Nagaoka, Ryo
    Masuno, Genta
    Kobayashi, Kazuto
    Yoshizawa, Shin
    Umemura, Shin-ichiro
    Saijo, Yoshifumi
    ULTRASONICS, 2015, 55 : 92 - 103
  • [8] Using a phonocardiography in a pulse wave velocity measurement
    Jelinek, M
    Dobes, J
    Pousek, L
    Hána, K
    PROCEEDINGS OF THE 3RD IEEE INTERNATIONAL SYMPOSIUM ON SIGNAL PROCESSING AND INFORMATION TECHNOLOGY, 2003, : 491 - 493
  • [9] Pulse wave velocity measurement along the ulnar artery in the wrist region using a high frequency ultrasonic array
    Benchemoul, Maxime
    Mateo, Tony
    Savery, David
    Gehin, Claudine
    Massot, Bertrand
    Ferin, Guillaume
    Vince, Philippe
    Flesch, Martin
    2021 43RD ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE & BIOLOGY SOCIETY (EMBC), 2021, : 4123 - 4127
  • [10] Measurement of carotid pulse wave velocity using ultrafast ultrasound imaging in hypertensive patients
    Xiaopeng Li
    Jue Jiang
    Hong Zhang
    Hua Wang
    Donggang Han
    Qi Zhou
    Ya Gao
    Shanshan Yu
    Yanhua Qi
    Journal of Medical Ultrasonics, 2017, 44 : 183 - 190