Accuracy of Corneal Thickness by Swept-Source Optical Coherence Tomography and Scheimpflug Camera in Virgin and Treated Fuchs Endothelial Dystrophy

被引:11
|
作者
Arnalich-Montiel, Francisco [1 ]
Ortiz-Toquero, Sara [2 ]
Auladell, Clara [1 ]
Couceiro, Ana [1 ]
机构
[1] Ramon y Cajal Univ Hosp, Dept Ophthalmol, Carretera Colmenar Viejo Km 9,100, Madrid 28034, Spain
[2] Univ Valladolid, Sch Optometry, IOBA Eye Inst, Optometry Res Grp, Valladolid, Spain
关键词
Fuchs endothelial dystrophy; DSAEK; DMEK; optical coherence tomography; Scheimpflug tomography; corneal thickness; REPEATABILITY; PACHYMETRY; ANTERIOR; PENTACAM; KERATOCONUS;
D O I
10.1097/ICO.0000000000001530
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To assess intraobserver repeatability, intersession reproducibility, and agreement of swept-source Fourier-domain optical coherence tomography (SS-OCT) and the Scheimpflug camera in measuring corneal thickness in virgin and grafted eyes with Fuchs endothelial corneal dystrophy (FECD). Methods: Thirty-six control eyes, 35 FECD eyes, 30 FECD with corneal edema eyes, 25 Descemet stripping automated endothelial keratoplasty (DSAEK) eyes, and 29 Descemet membrane endothelial keratoplasty (DMEK) eyes were included. The apical center, pupillary center, and thinnest corneal thickness were determined in 3 consecutive images and repeated 2 weeks later. Repeatability and reproducibility coefficients, intraclass correlation coefficients, and 95% limits of agreement (LOA) between measurements were calculated. Agreement between devices was assessed using Bland- Altman analysis. Results: Corneal thickness measurements were highly reproducible and repeatable with both systems. SS-OCT showed better repeatability in all corneal locations in the normal, FECD, FECD with edema, DSAEK, and DMEK groups (coefficient of variation <= 0.60%, <= 0.36%, <= 0.43%, <= 1.09%, and <= 0.48%, respectively) than the Scheimpflug (coefficient of variation <= 1.15%, <= 0.92%, <= 1.10%, <= 1.25%, and <= 1.14%, respectively). Between-session 95% LOA for SS-OCT was less than 3% for all groups except for the FECD with edema group, being almost double using the Scheimpflug camera. Differences between instruments were statistically significant in all groups and locations (P < 0.01) except in the DSAEK group (P <= 0.51); however, SS-OCT underestimated all measurements. Conclusions: SS-OCT provides better reproducible and repeatable measurements of corneal thickness than those obtained with the Scheimpflug camera in patients with FECD or an endothelial transplant. Variations between examinations higher than the 95% LOA observed in our study should raise awareness of changes in the endothelial function.
引用
收藏
页码:727 / 733
页数:7
相关论文
共 50 条
  • [41] Widefield Swept-Source Optical Coherence Tomography Angiography
    Shahidzadeh, Anoush
    Kim, Alice Yon
    Aziz, Hassan A.
    Durbin, Mary K.
    Puliafito, Carmen A.
    Kashani, Amir H.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2016, 57 (12)
  • [42] Utility of the Swept Source Optical Coherence Tomography for Measurements of Central Corneal Thickness
    Kim, Im Gyu
    Lee, Chong Eun
    Lee, Jun Seok
    Seo, Sam
    Lee, Kyoo Won
    JOURNAL OF THE KOREAN OPHTHALMOLOGICAL SOCIETY, 2016, 57 (10): : 1542 - 1548
  • [43] Posterior corneal aberrations and visual acuity in Fuchs endothelial corneal dystrophy classified by Scheimpflug tomography
    Patel, Sanjay V.
    Treichel, Emily J.
    Hodge, David O.
    Spiegel, Matthew R.
    Baratz, Keith H.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2020, 61 (07)
  • [44] Reliability and repeatability of swept-source Fourier-domain optical coherence tomography and Scheimpflug imaging in keratoconus
    Szalai, Eszter
    Berta, Andras
    Hassan, Ziad
    Modis, Laszlo, Jr.
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2012, 38 (03): : 485 - 494
  • [45] Longitudinal Evaluation of Cornea With Swept-Source Optical Coherence Tomography and Scheimpflug Imaging Before and After Lasik
    Chan, Tommy C. Y.
    Biswas, Sayantan
    Yu, Marco
    Jhanji, Vishal
    MEDICINE, 2015, 94 (30)
  • [46] Choroidal Thickness in Relation to Bone Mineral Density with Swept-Source Optical Coherence Tomography
    Jiang, Lin
    Qian, Yiwen
    Li, Qingjian
    Jiang, Jing
    Zhang, Yu
    Che, Xin
    Wang, Zhiliang
    JOURNAL OF OPHTHALMOLOGY, 2021, 2021
  • [47] Effect of Hemodialysis on Peripapillary Choroidal Thickness Measured by Swept-Source Optical Coherence Tomography
    Lee, Won June
    Hong, Rimkyung
    Kang, Min Ho
    Cho, Heeyoon
    Han, Sang-Woong
    Yi, Joo-Hark
    Shin, Yong Un
    Seong, Mincheol
    JOURNAL OF GLAUCOMA, 2021, 30 (06) : 459 - 464
  • [48] Factors influencing choroidal thickness measurements using swept-source optical coherence tomography
    Caramoy, Albert
    Kirchhof, Bernd
    Cursiefen, Claus
    Heindl, Ludwig M.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2015, 56 (07)
  • [49] Choroidal thickness in normal Indian eyes using swept-source optical coherence tomography
    Bhayana, Amber A.
    Kumar, Vinod
    Tayade, Akshay
    Chandra, Mahesh
    Chandra, Parijat
    Kumar, Atul
    INDIAN JOURNAL OF OPHTHALMOLOGY, 2019, 67 (02) : 252 - 255
  • [50] Topographic Variations of Choroidal Thickness in Healthy Eyes on Swept-Source Optical Coherence Tomography
    Touhami, Sara
    Philippakis, Elise
    Mrejen, Sarah
    Couturier, Aude
    Casteran, Celine
    Levent, Priscille
    Tadayoni, Ramin
    Gaudric, Alain
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2020, 61 (03)