Objective: To determine the reliability of a gas analyzer while assessing oxygen consumption (VO2) during a 5-min walking test. Design: Forty healthy participants were connected to the KB 1 -C and walked for 5 mins on a 5-m walkway. TotalVO(2) or energy expenditure (EE) was obtained by averaging the samples for each minute for 5 mins of walking. Walk distance (D), gait speed (S), and gait energy cost (C) were also evaluated. Results: The ICC2,1 for EE, D, S, and C were 0.88, 0.92, 0.92, and 0.67, respectively. The coefficient of variation across trials for each variable was 7.4, 3.6, 3.2, and 9%, respectively. Conclusion: All but C generated high coefficients of reliability. All variables demonstrated acceptable test-retest variability. There was low variability between participants for C.